Cargando…
3D VMAT Verification Based on Monte Carlo Log File Simulation with Experimental Feedback from Film Dosimetry
A model based on a specific phantom, called QuAArC, has been designed for the evaluation of planning and verification systems of complex radiotherapy treatments, such as volumetric modulated arc therapy (VMAT). This model uses the high accuracy provided by the Monte Carlo (MC) simulation of log file...
Autores principales: | , , , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Public Library of Science
2016
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5117721/ https://www.ncbi.nlm.nih.gov/pubmed/27870878 http://dx.doi.org/10.1371/journal.pone.0166767 |
_version_ | 1782468853337423872 |
---|---|
author | Barbeiro, A. R. Ureba, A. Baeza, J. A. Linares, R. Perucha, M. Jiménez-Ortega, E. Velázquez, S. Mateos, J. C. Leal, A. |
author_facet | Barbeiro, A. R. Ureba, A. Baeza, J. A. Linares, R. Perucha, M. Jiménez-Ortega, E. Velázquez, S. Mateos, J. C. Leal, A. |
author_sort | Barbeiro, A. R. |
collection | PubMed |
description | A model based on a specific phantom, called QuAArC, has been designed for the evaluation of planning and verification systems of complex radiotherapy treatments, such as volumetric modulated arc therapy (VMAT). This model uses the high accuracy provided by the Monte Carlo (MC) simulation of log files and allows the experimental feedback from the high spatial resolution of films hosted in QuAArC. This cylindrical phantom was specifically designed to host films rolled at different radial distances able to take into account the entrance fluence and the 3D dose distribution. Ionization chamber measurements are also included in the feedback process for absolute dose considerations. In this way, automated MC simulation of treatment log files is implemented to calculate the actual delivery geometries, while the monitor units are experimentally adjusted to reconstruct the dose-volume histogram (DVH) on the patient CT. Prostate and head and neck clinical cases, previously planned with Monaco and Pinnacle treatment planning systems and verified with two different commercial systems (Delta4 and COMPASS), were selected in order to test operational feasibility of the proposed model. The proper operation of the feedback procedure was proved through the achieved high agreement between reconstructed dose distributions and the film measurements (global gamma passing rates > 90% for the 2%/2 mm criteria). The necessary discretization level of the log file for dose calculation and the potential mismatching between calculated control points and detection grid in the verification process were discussed. Besides the effect of dose calculation accuracy of the analytic algorithm implemented in treatment planning systems for a dynamic technique, it was discussed the importance of the detection density level and its location in VMAT specific phantom to obtain a more reliable DVH in the patient CT. The proposed model also showed enough robustness and efficiency to be considered as a pre-treatment VMAT verification system. |
format | Online Article Text |
id | pubmed-5117721 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2016 |
publisher | Public Library of Science |
record_format | MEDLINE/PubMed |
spelling | pubmed-51177212016-12-15 3D VMAT Verification Based on Monte Carlo Log File Simulation with Experimental Feedback from Film Dosimetry Barbeiro, A. R. Ureba, A. Baeza, J. A. Linares, R. Perucha, M. Jiménez-Ortega, E. Velázquez, S. Mateos, J. C. Leal, A. PLoS One Research Article A model based on a specific phantom, called QuAArC, has been designed for the evaluation of planning and verification systems of complex radiotherapy treatments, such as volumetric modulated arc therapy (VMAT). This model uses the high accuracy provided by the Monte Carlo (MC) simulation of log files and allows the experimental feedback from the high spatial resolution of films hosted in QuAArC. This cylindrical phantom was specifically designed to host films rolled at different radial distances able to take into account the entrance fluence and the 3D dose distribution. Ionization chamber measurements are also included in the feedback process for absolute dose considerations. In this way, automated MC simulation of treatment log files is implemented to calculate the actual delivery geometries, while the monitor units are experimentally adjusted to reconstruct the dose-volume histogram (DVH) on the patient CT. Prostate and head and neck clinical cases, previously planned with Monaco and Pinnacle treatment planning systems and verified with two different commercial systems (Delta4 and COMPASS), were selected in order to test operational feasibility of the proposed model. The proper operation of the feedback procedure was proved through the achieved high agreement between reconstructed dose distributions and the film measurements (global gamma passing rates > 90% for the 2%/2 mm criteria). The necessary discretization level of the log file for dose calculation and the potential mismatching between calculated control points and detection grid in the verification process were discussed. Besides the effect of dose calculation accuracy of the analytic algorithm implemented in treatment planning systems for a dynamic technique, it was discussed the importance of the detection density level and its location in VMAT specific phantom to obtain a more reliable DVH in the patient CT. The proposed model also showed enough robustness and efficiency to be considered as a pre-treatment VMAT verification system. Public Library of Science 2016-11-21 /pmc/articles/PMC5117721/ /pubmed/27870878 http://dx.doi.org/10.1371/journal.pone.0166767 Text en © 2016 Barbeiro et al http://creativecommons.org/licenses/by/4.0/ This is an open access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/4.0/) , which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited. |
spellingShingle | Research Article Barbeiro, A. R. Ureba, A. Baeza, J. A. Linares, R. Perucha, M. Jiménez-Ortega, E. Velázquez, S. Mateos, J. C. Leal, A. 3D VMAT Verification Based on Monte Carlo Log File Simulation with Experimental Feedback from Film Dosimetry |
title | 3D VMAT Verification Based on Monte Carlo Log File Simulation with Experimental Feedback from Film Dosimetry |
title_full | 3D VMAT Verification Based on Monte Carlo Log File Simulation with Experimental Feedback from Film Dosimetry |
title_fullStr | 3D VMAT Verification Based on Monte Carlo Log File Simulation with Experimental Feedback from Film Dosimetry |
title_full_unstemmed | 3D VMAT Verification Based on Monte Carlo Log File Simulation with Experimental Feedback from Film Dosimetry |
title_short | 3D VMAT Verification Based on Monte Carlo Log File Simulation with Experimental Feedback from Film Dosimetry |
title_sort | 3d vmat verification based on monte carlo log file simulation with experimental feedback from film dosimetry |
topic | Research Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5117721/ https://www.ncbi.nlm.nih.gov/pubmed/27870878 http://dx.doi.org/10.1371/journal.pone.0166767 |
work_keys_str_mv | AT barbeiroar 3dvmatverificationbasedonmontecarlologfilesimulationwithexperimentalfeedbackfromfilmdosimetry AT urebaa 3dvmatverificationbasedonmontecarlologfilesimulationwithexperimentalfeedbackfromfilmdosimetry AT baezaja 3dvmatverificationbasedonmontecarlologfilesimulationwithexperimentalfeedbackfromfilmdosimetry AT linaresr 3dvmatverificationbasedonmontecarlologfilesimulationwithexperimentalfeedbackfromfilmdosimetry AT perucham 3dvmatverificationbasedonmontecarlologfilesimulationwithexperimentalfeedbackfromfilmdosimetry AT jimenezortegae 3dvmatverificationbasedonmontecarlologfilesimulationwithexperimentalfeedbackfromfilmdosimetry AT velazquezs 3dvmatverificationbasedonmontecarlologfilesimulationwithexperimentalfeedbackfromfilmdosimetry AT mateosjc 3dvmatverificationbasedonmontecarlologfilesimulationwithexperimentalfeedbackfromfilmdosimetry AT leala 3dvmatverificationbasedonmontecarlologfilesimulationwithexperimentalfeedbackfromfilmdosimetry |