Which cuff should I use? Indirect blood pressure measurement for the diagnosis of hypertension in patients with obesity: a diagnostic accuracy review

OBJECTIVE: To determine the diagnostic accuracy of different methods of blood pressure (BP) measurement compared with reference standards for the diagnosis of hypertension in patients with obesity with a large arm circumference. DESIGN: Systematic review with meta-analysis with hierarchical summary...

Descripción completa

Detalles Bibliográficos
Autores principales: Irving, Greg, Holden, John, Stevens, Richard, McManus, Richard J
Formato: Online Artículo Texto
Lenguaje:English
Publicado: BMJ Publishing Group 2016
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5129068/
https://www.ncbi.nlm.nih.gov/pubmed/27810973
http://dx.doi.org/10.1136/bmjopen-2016-012429
_version_ 1782470526763008000
author Irving, Greg
Holden, John
Stevens, Richard
McManus, Richard J
author_facet Irving, Greg
Holden, John
Stevens, Richard
McManus, Richard J
author_sort Irving, Greg
collection PubMed
description OBJECTIVE: To determine the diagnostic accuracy of different methods of blood pressure (BP) measurement compared with reference standards for the diagnosis of hypertension in patients with obesity with a large arm circumference. DESIGN: Systematic review with meta-analysis with hierarchical summary receiver operating characteristic models. Bland-Altman analyses where individual patient data were available. Methodological quality appraised using Quality Assessment of Diagnostic Accuracy Studies 2 (QUADAS2) criteria. DATA SOURCES: MEDLINE, EMBASE, Cochrane, DARE, Medion and Trip databases were searched. ELIGIBILITY CRITERIA: Cross-sectional, randomised and cohort studies of diagnostic test accuracy that compared any non-invasive BP tests (upper arm, forearm, wrist, finger) with an appropriate reference standard (invasive BP, correctly fitting upper arm cuff, ambulatory BP monitoring) in primary care were included. RESULTS: 4037 potentially relevant papers were identified. 20 studies involving 26 different comparisons met the inclusion criteria. Individual patient data were available from 4 studies. No studies satisfied all QUADAS2 criteria. Compared with the reference test of invasive BP, a correctly fitting upper arm BP cuff had a sensitivity of 0.87 (0.79 to 0.93) and a specificity of 0.85 (0.64 to 0.95); insufficient evidence was available for other comparisons to invasive BP. Compared with the reference test of a correctly fitting upper arm cuff, BP measurement at the wrist had a sensitivity of 0.92 (0.64 to 0.99) and a specificity of 0.92 (0.85 to 0.87). Measurement with an incorrectly fitting standard cuff had a sensitivity of 0.73 (0.67 to 0.78) and a specificity of 0.76 (0.69 to 0.82). Measurement at the forearm had a sensitivity of 0.84 (0.71 to 0.92) and a specificity 0.75 of (0.66 to 0.83). Bland-Altman analysis of individual patient data from 3 studies comparing wrist and upper arm BP showed a mean difference of 0.46 mm Hg for systolic BP measurement and 2.2 mm Hg for diastolic BP measurement. CONCLUSIONS: BP measurement with a correctly fitting upper arm cuff is sufficiently sensitive and specific to diagnose hypertension in patients with obesity with a large upper arm circumference. If a correctly fitting upper arm cuff cannot be applied, an incorrectly fitting standard size cuff should not be used and BP measurement at the wrist should be considered.
format Online
Article
Text
id pubmed-5129068
institution National Center for Biotechnology Information
language English
publishDate 2016
publisher BMJ Publishing Group
record_format MEDLINE/PubMed
spelling pubmed-51290682016-12-02 Which cuff should I use? Indirect blood pressure measurement for the diagnosis of hypertension in patients with obesity: a diagnostic accuracy review Irving, Greg Holden, John Stevens, Richard McManus, Richard J BMJ Open General practice / Family practice OBJECTIVE: To determine the diagnostic accuracy of different methods of blood pressure (BP) measurement compared with reference standards for the diagnosis of hypertension in patients with obesity with a large arm circumference. DESIGN: Systematic review with meta-analysis with hierarchical summary receiver operating characteristic models. Bland-Altman analyses where individual patient data were available. Methodological quality appraised using Quality Assessment of Diagnostic Accuracy Studies 2 (QUADAS2) criteria. DATA SOURCES: MEDLINE, EMBASE, Cochrane, DARE, Medion and Trip databases were searched. ELIGIBILITY CRITERIA: Cross-sectional, randomised and cohort studies of diagnostic test accuracy that compared any non-invasive BP tests (upper arm, forearm, wrist, finger) with an appropriate reference standard (invasive BP, correctly fitting upper arm cuff, ambulatory BP monitoring) in primary care were included. RESULTS: 4037 potentially relevant papers were identified. 20 studies involving 26 different comparisons met the inclusion criteria. Individual patient data were available from 4 studies. No studies satisfied all QUADAS2 criteria. Compared with the reference test of invasive BP, a correctly fitting upper arm BP cuff had a sensitivity of 0.87 (0.79 to 0.93) and a specificity of 0.85 (0.64 to 0.95); insufficient evidence was available for other comparisons to invasive BP. Compared with the reference test of a correctly fitting upper arm cuff, BP measurement at the wrist had a sensitivity of 0.92 (0.64 to 0.99) and a specificity of 0.92 (0.85 to 0.87). Measurement with an incorrectly fitting standard cuff had a sensitivity of 0.73 (0.67 to 0.78) and a specificity of 0.76 (0.69 to 0.82). Measurement at the forearm had a sensitivity of 0.84 (0.71 to 0.92) and a specificity 0.75 of (0.66 to 0.83). Bland-Altman analysis of individual patient data from 3 studies comparing wrist and upper arm BP showed a mean difference of 0.46 mm Hg for systolic BP measurement and 2.2 mm Hg for diastolic BP measurement. CONCLUSIONS: BP measurement with a correctly fitting upper arm cuff is sufficiently sensitive and specific to diagnose hypertension in patients with obesity with a large upper arm circumference. If a correctly fitting upper arm cuff cannot be applied, an incorrectly fitting standard size cuff should not be used and BP measurement at the wrist should be considered. BMJ Publishing Group 2016-11-03 /pmc/articles/PMC5129068/ /pubmed/27810973 http://dx.doi.org/10.1136/bmjopen-2016-012429 Text en Published by the BMJ Publishing Group Limited. For permission to use (where not already granted under a licence) please go to http://www.bmj.com/company/products-services/rights-and-licensing/ This is an Open Access article distributed in accordance with the Creative Commons Attribution Non Commercial (CC BY-NC 4.0) license, which permits others to distribute, remix, adapt, build upon this work non-commercially, and license their derivative works on different terms, provided the original work is properly cited and the use is non-commercial. See: http://creativecommons.org/licenses/by-nc/4.0/
spellingShingle General practice / Family practice
Irving, Greg
Holden, John
Stevens, Richard
McManus, Richard J
Which cuff should I use? Indirect blood pressure measurement for the diagnosis of hypertension in patients with obesity: a diagnostic accuracy review
title Which cuff should I use? Indirect blood pressure measurement for the diagnosis of hypertension in patients with obesity: a diagnostic accuracy review
title_full Which cuff should I use? Indirect blood pressure measurement for the diagnosis of hypertension in patients with obesity: a diagnostic accuracy review
title_fullStr Which cuff should I use? Indirect blood pressure measurement for the diagnosis of hypertension in patients with obesity: a diagnostic accuracy review
title_full_unstemmed Which cuff should I use? Indirect blood pressure measurement for the diagnosis of hypertension in patients with obesity: a diagnostic accuracy review
title_short Which cuff should I use? Indirect blood pressure measurement for the diagnosis of hypertension in patients with obesity: a diagnostic accuracy review
title_sort which cuff should i use? indirect blood pressure measurement for the diagnosis of hypertension in patients with obesity: a diagnostic accuracy review
topic General practice / Family practice
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5129068/
https://www.ncbi.nlm.nih.gov/pubmed/27810973
http://dx.doi.org/10.1136/bmjopen-2016-012429
work_keys_str_mv AT irvinggreg whichcuffshouldiuseindirectbloodpressuremeasurementforthediagnosisofhypertensioninpatientswithobesityadiagnosticaccuracyreview
AT holdenjohn whichcuffshouldiuseindirectbloodpressuremeasurementforthediagnosisofhypertensioninpatientswithobesityadiagnosticaccuracyreview
AT stevensrichard whichcuffshouldiuseindirectbloodpressuremeasurementforthediagnosisofhypertensioninpatientswithobesityadiagnosticaccuracyreview
AT mcmanusrichardj whichcuffshouldiuseindirectbloodpressuremeasurementforthediagnosisofhypertensioninpatientswithobesityadiagnosticaccuracyreview