A simple and sensitive high-performance liquid chromatography method for determination of ciprofloxacin in bioavailability studies of conventional and gastroretentive prolonged-release formulations
BACKGROUND: A very simple, sensitive, and accurate high-performance liquid chromatography (HPLC) method with ultraviolet detector was developed and applied to determine ciprofloxacin in human plasma following administration of a gastroretentive formulation developed in our laboratory. MATERIALS AND...
Autores principales: | , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Medknow Publications & Media Pvt Ltd
2016
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5137234/ https://www.ncbi.nlm.nih.gov/pubmed/27995102 http://dx.doi.org/10.4103/2277-9175.190995 |
_version_ | 1782471877443190784 |
---|---|
author | Emami, Jaber Rezazadeh, Mahboubeh |
author_facet | Emami, Jaber Rezazadeh, Mahboubeh |
author_sort | Emami, Jaber |
collection | PubMed |
description | BACKGROUND: A very simple, sensitive, and accurate high-performance liquid chromatography (HPLC) method with ultraviolet detector was developed and applied to determine ciprofloxacin in human plasma following administration of a gastroretentive formulation developed in our laboratory. MATERIALS AND METHODS: HPLC analysis was performed on a C(18) μ-Bondapack column (250 mm × 3.9 mm) using acetonitrile: potassium dihydrogen phosphate solution 0.1 M (20:80, v/v, pH 3) at a flow rate of 1.5 ml/min and eluate was monitored at 276 nm. After addition of phenacetin as internal standard, plasma samples were treated with 0.1 M phosphate buffer (pH: 7) and followed by extraction with dichloromethane. The method was validated for linearity, precision, accuracy, limit of quantitation (LOQ), robustness, stability, and applied in bioavailability studies of our developed gastroretentive formulation in healthy volunteers. RESULTS: The calibration curves were linear over the concentration range 0.025–4 μg/ml with the detection limit of 15 ng/ml. Accuracy % were within 93–115 and the coefficient of variance % ranged from 0.20 to 12.8. The very low LOQ (25 ng/ml) allowed avoiding fluorometric detection which is more expensive and is not available in all laboratories. Ciprofloxacin was stable in samples with no evidence of degradation during 3 freeze-thaw cycles and 3 months storage at –70°C. CONCLUSION: This validated HPLC method was successfully used for the determination of ciprofloxacin in human plasma following oral administration of controlled release formulation, conventional immediate-release tablets and when administered concomitantly with divalent and trivalent cations such as aluminum-, magnesium-, or calcium-containing products under which the bioavailability of ciprofloxacin is significantly reduced. |
format | Online Article Text |
id | pubmed-5137234 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2016 |
publisher | Medknow Publications & Media Pvt Ltd |
record_format | MEDLINE/PubMed |
spelling | pubmed-51372342016-12-19 A simple and sensitive high-performance liquid chromatography method for determination of ciprofloxacin in bioavailability studies of conventional and gastroretentive prolonged-release formulations Emami, Jaber Rezazadeh, Mahboubeh Adv Biomed Res Original Article BACKGROUND: A very simple, sensitive, and accurate high-performance liquid chromatography (HPLC) method with ultraviolet detector was developed and applied to determine ciprofloxacin in human plasma following administration of a gastroretentive formulation developed in our laboratory. MATERIALS AND METHODS: HPLC analysis was performed on a C(18) μ-Bondapack column (250 mm × 3.9 mm) using acetonitrile: potassium dihydrogen phosphate solution 0.1 M (20:80, v/v, pH 3) at a flow rate of 1.5 ml/min and eluate was monitored at 276 nm. After addition of phenacetin as internal standard, plasma samples were treated with 0.1 M phosphate buffer (pH: 7) and followed by extraction with dichloromethane. The method was validated for linearity, precision, accuracy, limit of quantitation (LOQ), robustness, stability, and applied in bioavailability studies of our developed gastroretentive formulation in healthy volunteers. RESULTS: The calibration curves were linear over the concentration range 0.025–4 μg/ml with the detection limit of 15 ng/ml. Accuracy % were within 93–115 and the coefficient of variance % ranged from 0.20 to 12.8. The very low LOQ (25 ng/ml) allowed avoiding fluorometric detection which is more expensive and is not available in all laboratories. Ciprofloxacin was stable in samples with no evidence of degradation during 3 freeze-thaw cycles and 3 months storage at –70°C. CONCLUSION: This validated HPLC method was successfully used for the determination of ciprofloxacin in human plasma following oral administration of controlled release formulation, conventional immediate-release tablets and when administered concomitantly with divalent and trivalent cations such as aluminum-, magnesium-, or calcium-containing products under which the bioavailability of ciprofloxacin is significantly reduced. Medknow Publications & Media Pvt Ltd 2016-10-26 /pmc/articles/PMC5137234/ /pubmed/27995102 http://dx.doi.org/10.4103/2277-9175.190995 Text en Copyright: © 2016 Advanced Biomedical Research http://creativecommons.org/licenses/by-nc-sa/3.0 This is an open access article distributed under the terms of the Creative Commons Attribution-NonCommercial-ShareAlike 3.0 License, which allows others to remix, tweak, and build upon the work non-commercially, as long as the author is credited and the new creations are licensed under the identical terms. |
spellingShingle | Original Article Emami, Jaber Rezazadeh, Mahboubeh A simple and sensitive high-performance liquid chromatography method for determination of ciprofloxacin in bioavailability studies of conventional and gastroretentive prolonged-release formulations |
title | A simple and sensitive high-performance liquid chromatography method for determination of ciprofloxacin in bioavailability studies of conventional and gastroretentive prolonged-release formulations |
title_full | A simple and sensitive high-performance liquid chromatography method for determination of ciprofloxacin in bioavailability studies of conventional and gastroretentive prolonged-release formulations |
title_fullStr | A simple and sensitive high-performance liquid chromatography method for determination of ciprofloxacin in bioavailability studies of conventional and gastroretentive prolonged-release formulations |
title_full_unstemmed | A simple and sensitive high-performance liquid chromatography method for determination of ciprofloxacin in bioavailability studies of conventional and gastroretentive prolonged-release formulations |
title_short | A simple and sensitive high-performance liquid chromatography method for determination of ciprofloxacin in bioavailability studies of conventional and gastroretentive prolonged-release formulations |
title_sort | simple and sensitive high-performance liquid chromatography method for determination of ciprofloxacin in bioavailability studies of conventional and gastroretentive prolonged-release formulations |
topic | Original Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5137234/ https://www.ncbi.nlm.nih.gov/pubmed/27995102 http://dx.doi.org/10.4103/2277-9175.190995 |
work_keys_str_mv | AT emamijaber asimpleandsensitivehighperformanceliquidchromatographymethodfordeterminationofciprofloxacininbioavailabilitystudiesofconventionalandgastroretentiveprolongedreleaseformulations AT rezazadehmahboubeh asimpleandsensitivehighperformanceliquidchromatographymethodfordeterminationofciprofloxacininbioavailabilitystudiesofconventionalandgastroretentiveprolongedreleaseformulations AT emamijaber simpleandsensitivehighperformanceliquidchromatographymethodfordeterminationofciprofloxacininbioavailabilitystudiesofconventionalandgastroretentiveprolongedreleaseformulations AT rezazadehmahboubeh simpleandsensitivehighperformanceliquidchromatographymethodfordeterminationofciprofloxacininbioavailabilitystudiesofconventionalandgastroretentiveprolongedreleaseformulations |