Stereoscopic X-ray imaging, cone beam CT, and couch positioning in stereotactic radiotherapy of intracranial tumors: preliminary results from a cross-modality pilot installation

BACKGROUND: To assess the accuracy and precision of a fully integrated pilot installation of stereoscopic X-ray imaging and kV-CBCT for automatic couch positioning in stereotactic radiotherapy of intracranial tumors. Positioning errors as detected by stereoscopic X-ray imaging are compared to those...

Descripción completa

Detalles Bibliográficos
Autores principales: Zollner, Barbara, Heinz, Christian, Pitzler, Sabrina, Manapov, Farkhad, Kantz, Steffi, Rottler, Maya Christine, Niyazi, Maximilian, Ganswindt, Ute, Belka, Claus, Ballhausen, Hendrik
Formato: Online Artículo Texto
Lenguaje:English
Publicado: BioMed Central 2016
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5142336/
https://www.ncbi.nlm.nih.gov/pubmed/27927235
http://dx.doi.org/10.1186/s13014-016-0735-2
_version_ 1782472753403658240
author Zollner, Barbara
Heinz, Christian
Pitzler, Sabrina
Manapov, Farkhad
Kantz, Steffi
Rottler, Maya Christine
Niyazi, Maximilian
Ganswindt, Ute
Belka, Claus
Ballhausen, Hendrik
author_facet Zollner, Barbara
Heinz, Christian
Pitzler, Sabrina
Manapov, Farkhad
Kantz, Steffi
Rottler, Maya Christine
Niyazi, Maximilian
Ganswindt, Ute
Belka, Claus
Ballhausen, Hendrik
author_sort Zollner, Barbara
collection PubMed
description BACKGROUND: To assess the accuracy and precision of a fully integrated pilot installation of stereoscopic X-ray imaging and kV-CBCT for automatic couch positioning in stereotactic radiotherapy of intracranial tumors. Positioning errors as detected by stereoscopic X-ray imaging are compared to those by kV-CBCT (i.e. the accuracy of the new method is verified by the established method), and repeated X-ray images are compared (i.e. the precision of new method is determined intra-modally). METHODS: Preliminary results are reported from a study with 32 patients with intracranial tumors. Patients were treated with stereotactic radiotherapy guided by stereoscopic X-ray imaging and kV-CBCT. Patient positioning was automatically corrected by a robotic couch. Cross-modal discrepancies in position detection were measured (N = 42). Intra-modal improvements after correction and re-verification by stereoscopic X-ray imaging were measured (N = 70). The accuracy and precision of stereoscopic X-ray imaging and the accuracy and precision of CBCT were confirmed in phantom measurements (N = 12 shifts of a ball bearing phantom, N = 24 shifts of a head phantom). RESULTS: After correction based on stereoscopic X-ray imaging 95% of residual mean errors were below 0.4, 0.4, 0.5, and 0.7 mm (lateral, longitudinal, vertical, radial, respectively). Stereoscopic X-ray imaging and CBCT were in close agreement with an average discrepancy of 0.1, 0.5, 0.3 and 0.8 mm, respectively. 95% of discrepancies were below 0.8, 1.2, 1.0, and 1.4 mm, respectively. After correction and re-verification by stereoscopic X-ray imaging, the remaining intra-modal residual error was consistent with zero (p = 0.31, p = 0.48, p = 0.81 in lateral, longitudinal, and vertical direction; p-values from two-tailed t-test). The inherent technical accuracy and precision of stereoscopic X-ray imaging and the accuracy and precision of CBCT were found to be of the order of 0.1 mm in controlled phantom settings. CONCLUSIONS: In a routine clinical setting, both stereoscopic X-ray imaging and CBCT were able to reduce positioning errors by an order of magnitude. The end-to-end precision of the system, measured from the discrepancy (mean) between ExacTrac and CBCT, in a clinical setting seems to be about 0.8 mm radially, including couch positioning. The precision (measured from repeatability of ExacTrac, intra-modal) was found to be about 0.7 mm radially in a clinical setting.
format Online
Article
Text
id pubmed-5142336
institution National Center for Biotechnology Information
language English
publishDate 2016
publisher BioMed Central
record_format MEDLINE/PubMed
spelling pubmed-51423362016-12-15 Stereoscopic X-ray imaging, cone beam CT, and couch positioning in stereotactic radiotherapy of intracranial tumors: preliminary results from a cross-modality pilot installation Zollner, Barbara Heinz, Christian Pitzler, Sabrina Manapov, Farkhad Kantz, Steffi Rottler, Maya Christine Niyazi, Maximilian Ganswindt, Ute Belka, Claus Ballhausen, Hendrik Radiat Oncol Research BACKGROUND: To assess the accuracy and precision of a fully integrated pilot installation of stereoscopic X-ray imaging and kV-CBCT for automatic couch positioning in stereotactic radiotherapy of intracranial tumors. Positioning errors as detected by stereoscopic X-ray imaging are compared to those by kV-CBCT (i.e. the accuracy of the new method is verified by the established method), and repeated X-ray images are compared (i.e. the precision of new method is determined intra-modally). METHODS: Preliminary results are reported from a study with 32 patients with intracranial tumors. Patients were treated with stereotactic radiotherapy guided by stereoscopic X-ray imaging and kV-CBCT. Patient positioning was automatically corrected by a robotic couch. Cross-modal discrepancies in position detection were measured (N = 42). Intra-modal improvements after correction and re-verification by stereoscopic X-ray imaging were measured (N = 70). The accuracy and precision of stereoscopic X-ray imaging and the accuracy and precision of CBCT were confirmed in phantom measurements (N = 12 shifts of a ball bearing phantom, N = 24 shifts of a head phantom). RESULTS: After correction based on stereoscopic X-ray imaging 95% of residual mean errors were below 0.4, 0.4, 0.5, and 0.7 mm (lateral, longitudinal, vertical, radial, respectively). Stereoscopic X-ray imaging and CBCT were in close agreement with an average discrepancy of 0.1, 0.5, 0.3 and 0.8 mm, respectively. 95% of discrepancies were below 0.8, 1.2, 1.0, and 1.4 mm, respectively. After correction and re-verification by stereoscopic X-ray imaging, the remaining intra-modal residual error was consistent with zero (p = 0.31, p = 0.48, p = 0.81 in lateral, longitudinal, and vertical direction; p-values from two-tailed t-test). The inherent technical accuracy and precision of stereoscopic X-ray imaging and the accuracy and precision of CBCT were found to be of the order of 0.1 mm in controlled phantom settings. CONCLUSIONS: In a routine clinical setting, both stereoscopic X-ray imaging and CBCT were able to reduce positioning errors by an order of magnitude. The end-to-end precision of the system, measured from the discrepancy (mean) between ExacTrac and CBCT, in a clinical setting seems to be about 0.8 mm radially, including couch positioning. The precision (measured from repeatability of ExacTrac, intra-modal) was found to be about 0.7 mm radially in a clinical setting. BioMed Central 2016-12-07 /pmc/articles/PMC5142336/ /pubmed/27927235 http://dx.doi.org/10.1186/s13014-016-0735-2 Text en © The Author(s). 2016 Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.
spellingShingle Research
Zollner, Barbara
Heinz, Christian
Pitzler, Sabrina
Manapov, Farkhad
Kantz, Steffi
Rottler, Maya Christine
Niyazi, Maximilian
Ganswindt, Ute
Belka, Claus
Ballhausen, Hendrik
Stereoscopic X-ray imaging, cone beam CT, and couch positioning in stereotactic radiotherapy of intracranial tumors: preliminary results from a cross-modality pilot installation
title Stereoscopic X-ray imaging, cone beam CT, and couch positioning in stereotactic radiotherapy of intracranial tumors: preliminary results from a cross-modality pilot installation
title_full Stereoscopic X-ray imaging, cone beam CT, and couch positioning in stereotactic radiotherapy of intracranial tumors: preliminary results from a cross-modality pilot installation
title_fullStr Stereoscopic X-ray imaging, cone beam CT, and couch positioning in stereotactic radiotherapy of intracranial tumors: preliminary results from a cross-modality pilot installation
title_full_unstemmed Stereoscopic X-ray imaging, cone beam CT, and couch positioning in stereotactic radiotherapy of intracranial tumors: preliminary results from a cross-modality pilot installation
title_short Stereoscopic X-ray imaging, cone beam CT, and couch positioning in stereotactic radiotherapy of intracranial tumors: preliminary results from a cross-modality pilot installation
title_sort stereoscopic x-ray imaging, cone beam ct, and couch positioning in stereotactic radiotherapy of intracranial tumors: preliminary results from a cross-modality pilot installation
topic Research
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5142336/
https://www.ncbi.nlm.nih.gov/pubmed/27927235
http://dx.doi.org/10.1186/s13014-016-0735-2
work_keys_str_mv AT zollnerbarbara stereoscopicxrayimagingconebeamctandcouchpositioninginstereotacticradiotherapyofintracranialtumorspreliminaryresultsfromacrossmodalitypilotinstallation
AT heinzchristian stereoscopicxrayimagingconebeamctandcouchpositioninginstereotacticradiotherapyofintracranialtumorspreliminaryresultsfromacrossmodalitypilotinstallation
AT pitzlersabrina stereoscopicxrayimagingconebeamctandcouchpositioninginstereotacticradiotherapyofintracranialtumorspreliminaryresultsfromacrossmodalitypilotinstallation
AT manapovfarkhad stereoscopicxrayimagingconebeamctandcouchpositioninginstereotacticradiotherapyofintracranialtumorspreliminaryresultsfromacrossmodalitypilotinstallation
AT kantzsteffi stereoscopicxrayimagingconebeamctandcouchpositioninginstereotacticradiotherapyofintracranialtumorspreliminaryresultsfromacrossmodalitypilotinstallation
AT rottlermayachristine stereoscopicxrayimagingconebeamctandcouchpositioninginstereotacticradiotherapyofintracranialtumorspreliminaryresultsfromacrossmodalitypilotinstallation
AT niyazimaximilian stereoscopicxrayimagingconebeamctandcouchpositioninginstereotacticradiotherapyofintracranialtumorspreliminaryresultsfromacrossmodalitypilotinstallation
AT ganswindtute stereoscopicxrayimagingconebeamctandcouchpositioninginstereotacticradiotherapyofintracranialtumorspreliminaryresultsfromacrossmodalitypilotinstallation
AT belkaclaus stereoscopicxrayimagingconebeamctandcouchpositioninginstereotacticradiotherapyofintracranialtumorspreliminaryresultsfromacrossmodalitypilotinstallation
AT ballhausenhendrik stereoscopicxrayimagingconebeamctandcouchpositioninginstereotacticradiotherapyofintracranialtumorspreliminaryresultsfromacrossmodalitypilotinstallation