Validity of bedside blood glucose measurement in critically ill patients with intensive insulin therapy
BACKGROUND AND AIMS: There have been variable results on the practice of tight glycemic control, and studies have demonstrated that point-of-care (POC) glucometers have variable accuracy. Glucometers must be accurate, and many variables can affect blood glucose levels. The purpose of this study was...
Autores principales: | , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Medknow Publications & Media Pvt Ltd
2016
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5144527/ https://www.ncbi.nlm.nih.gov/pubmed/27994380 http://dx.doi.org/10.4103/0972-5229.194009 |
_version_ | 1782473142880436224 |
---|---|
author | Mahmoodpoor, Ata Hamishehkar, Hadi Shadvar, Kamran Sanaie, Sarvin Iranpour, Afshin Fattahi, Vahid |
author_facet | Mahmoodpoor, Ata Hamishehkar, Hadi Shadvar, Kamran Sanaie, Sarvin Iranpour, Afshin Fattahi, Vahid |
author_sort | Mahmoodpoor, Ata |
collection | PubMed |
description | BACKGROUND AND AIMS: There have been variable results on the practice of tight glycemic control, and studies have demonstrated that point-of-care (POC) glucometers have variable accuracy. Glucometers must be accurate, and many variables can affect blood glucose levels. The purpose of this study was to determine the difference between blood glucose concentrations obtained from POC glucometers and laboratory results in critically ill patients with intensive insulin therapy. MATERIALS AND METHODS: This was a descriptive study which enrolled 300 critically ill patients. Four samples of arterial blood were collected and analyzed at the bedside with the POC glucometer and also in the central laboratory to obtain the blood glucose level. To define the effect of various factors on this relation, we noted the levels of hemoglobin (Hb), PaO(2), body temperature, bilirubin, history of drug usage, and sepsis. RESULTS: There were not any significant differences between blood sugar levels using laboratory and glucometer methods of measurements. There was a good and significant correlation between glucose levels between two methods (r = 0.81, P < 0.001). Among evaluated factors (body temperature, bilirubin level, blood pressure, Hb level, PaO(2), sepsis, and drugs) which added one by one in model, just drugs decreased the correlation more than others (r = 0.78). CONCLUSIONS: The results of POC glucometer differ from laboratory glucose concentrations, especially in critically ill patients with unstable hemodynamic status while receiving several drugs. This may raise the concern about using POC devices for tight glycemic control in critically ill patients. These results should be interpreted with caution because of the large variation of accuracy among different glucometer devices. |
format | Online Article Text |
id | pubmed-5144527 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2016 |
publisher | Medknow Publications & Media Pvt Ltd |
record_format | MEDLINE/PubMed |
spelling | pubmed-51445272016-12-19 Validity of bedside blood glucose measurement in critically ill patients with intensive insulin therapy Mahmoodpoor, Ata Hamishehkar, Hadi Shadvar, Kamran Sanaie, Sarvin Iranpour, Afshin Fattahi, Vahid Indian J Crit Care Med Research Article BACKGROUND AND AIMS: There have been variable results on the practice of tight glycemic control, and studies have demonstrated that point-of-care (POC) glucometers have variable accuracy. Glucometers must be accurate, and many variables can affect blood glucose levels. The purpose of this study was to determine the difference between blood glucose concentrations obtained from POC glucometers and laboratory results in critically ill patients with intensive insulin therapy. MATERIALS AND METHODS: This was a descriptive study which enrolled 300 critically ill patients. Four samples of arterial blood were collected and analyzed at the bedside with the POC glucometer and also in the central laboratory to obtain the blood glucose level. To define the effect of various factors on this relation, we noted the levels of hemoglobin (Hb), PaO(2), body temperature, bilirubin, history of drug usage, and sepsis. RESULTS: There were not any significant differences between blood sugar levels using laboratory and glucometer methods of measurements. There was a good and significant correlation between glucose levels between two methods (r = 0.81, P < 0.001). Among evaluated factors (body temperature, bilirubin level, blood pressure, Hb level, PaO(2), sepsis, and drugs) which added one by one in model, just drugs decreased the correlation more than others (r = 0.78). CONCLUSIONS: The results of POC glucometer differ from laboratory glucose concentrations, especially in critically ill patients with unstable hemodynamic status while receiving several drugs. This may raise the concern about using POC devices for tight glycemic control in critically ill patients. These results should be interpreted with caution because of the large variation of accuracy among different glucometer devices. Medknow Publications & Media Pvt Ltd 2016-11 /pmc/articles/PMC5144527/ /pubmed/27994380 http://dx.doi.org/10.4103/0972-5229.194009 Text en Copyright: © Indian Journal of Critical Care Medicine http://creativecommons.org/licenses/by-nc-sa/3.0 This is an open access article distributed under the terms of the Creative Commons Attribution-NonCommercial-ShareAlike 3.0 License, which allows others to remix, tweak, and build upon the work non-commercially, as long as the author is credited and the new creations are licensed under the identical terms. |
spellingShingle | Research Article Mahmoodpoor, Ata Hamishehkar, Hadi Shadvar, Kamran Sanaie, Sarvin Iranpour, Afshin Fattahi, Vahid Validity of bedside blood glucose measurement in critically ill patients with intensive insulin therapy |
title | Validity of bedside blood glucose measurement in critically ill patients with intensive insulin therapy |
title_full | Validity of bedside blood glucose measurement in critically ill patients with intensive insulin therapy |
title_fullStr | Validity of bedside blood glucose measurement in critically ill patients with intensive insulin therapy |
title_full_unstemmed | Validity of bedside blood glucose measurement in critically ill patients with intensive insulin therapy |
title_short | Validity of bedside blood glucose measurement in critically ill patients with intensive insulin therapy |
title_sort | validity of bedside blood glucose measurement in critically ill patients with intensive insulin therapy |
topic | Research Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5144527/ https://www.ncbi.nlm.nih.gov/pubmed/27994380 http://dx.doi.org/10.4103/0972-5229.194009 |
work_keys_str_mv | AT mahmoodpoorata validityofbedsidebloodglucosemeasurementincriticallyillpatientswithintensiveinsulintherapy AT hamishehkarhadi validityofbedsidebloodglucosemeasurementincriticallyillpatientswithintensiveinsulintherapy AT shadvarkamran validityofbedsidebloodglucosemeasurementincriticallyillpatientswithintensiveinsulintherapy AT sanaiesarvin validityofbedsidebloodglucosemeasurementincriticallyillpatientswithintensiveinsulintherapy AT iranpourafshin validityofbedsidebloodglucosemeasurementincriticallyillpatientswithintensiveinsulintherapy AT fattahivahid validityofbedsidebloodglucosemeasurementincriticallyillpatientswithintensiveinsulintherapy |