In-Vitro Evaluation of Two Types of Neonatal Oxygenators in Handling Gaseous Microemboli and Maintaining Optimal Hemodynamic Stability During Cardiopulmonary Bypass

OBJECTIVE: Usually only FDA-approved oxygenators are subject of studies by the international scientific community. The objective of this study is to evaluate two types of neonatal membrane oxygenators in terms of transmembrane pressure gradient, hemodynamic energy transmission and gaseous microembol...

Descripción completa

Detalles Bibliográficos
Autores principales: Marupudi, Neelima, Wang, Shigang, Canêo, Luiz Fernando, Jatene, Fabio Biscegli, Kunselman, Allen R., Undar, Akif
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Sociedade Brasileira de Cirurgia Cardiovascular 2016
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5144558/
https://www.ncbi.nlm.nih.gov/pubmed/27982342
http://dx.doi.org/10.5935/1678-9741.20160075
_version_ 1782473145937035264
author Marupudi, Neelima
Wang, Shigang
Canêo, Luiz Fernando
Jatene, Fabio Biscegli
Kunselman, Allen R.
Undar, Akif
author_facet Marupudi, Neelima
Wang, Shigang
Canêo, Luiz Fernando
Jatene, Fabio Biscegli
Kunselman, Allen R.
Undar, Akif
author_sort Marupudi, Neelima
collection PubMed
description OBJECTIVE: Usually only FDA-approved oxygenators are subject of studies by the international scientific community. The objective of this study is to evaluate two types of neonatal membrane oxygenators in terms of transmembrane pressure gradient, hemodynamic energy transmission and gaseous microemboli capture in simulated cardiopulmonary bypass systems. METHODS: We investigated the Braile Infant 1500 (Braile Biomédica, São José do Rio Preto, Brazil), an oxygenator commonly used in Brazilian operating rooms, and compared it to the Dideco Kids D100 (Sorin Group, Arvada, CO, USA), that is an FDA-approved and widely used model in the USA. Cardiopulmonary bypass circuits were primed with lactated Ringer's solution and packed red blood cells (Hematocrit 40%). Trials were conducted at flow rates of 500 ml/min and 700 ml/min at 35ºC and 25ºC. Real-time pressure and flow data were recorded using a custom-based data acquisition system. For gaseous microemboli testing, 5cc of air were manually injected into the venous line. Gaseous microemboli were recorded using the Emboli Detection and Classification Quantifier. RESULTS: Braile Infant 1500 had a lower pressure drop (P<0.01) and a higher total hemodynamic energy delivered to the pseudopatient (P<0.01). However, there was a higher raw number of gaseous microemboli seen prior to oxygenator at lower temperatures with the Braile oxygenator compared to the Kids D100 (P<0.01). CONCLUSION: Braile Infant 1500 oxygenator had a better hemodynamic performance compared to the Dideco Kids D100 oxygenator. Braile had more gaseous microemboli detected at the pre-oxygenator site under hypothermia, but delivered a smaller percentage of air emboli to the pseudopatient than the Dideco oxygenator.
format Online
Article
Text
id pubmed-5144558
institution National Center for Biotechnology Information
language English
publishDate 2016
publisher Sociedade Brasileira de Cirurgia Cardiovascular
record_format MEDLINE/PubMed
spelling pubmed-51445582016-12-09 In-Vitro Evaluation of Two Types of Neonatal Oxygenators in Handling Gaseous Microemboli and Maintaining Optimal Hemodynamic Stability During Cardiopulmonary Bypass Marupudi, Neelima Wang, Shigang Canêo, Luiz Fernando Jatene, Fabio Biscegli Kunselman, Allen R. Undar, Akif Braz J Cardiovasc Surg Original Articles OBJECTIVE: Usually only FDA-approved oxygenators are subject of studies by the international scientific community. The objective of this study is to evaluate two types of neonatal membrane oxygenators in terms of transmembrane pressure gradient, hemodynamic energy transmission and gaseous microemboli capture in simulated cardiopulmonary bypass systems. METHODS: We investigated the Braile Infant 1500 (Braile Biomédica, São José do Rio Preto, Brazil), an oxygenator commonly used in Brazilian operating rooms, and compared it to the Dideco Kids D100 (Sorin Group, Arvada, CO, USA), that is an FDA-approved and widely used model in the USA. Cardiopulmonary bypass circuits were primed with lactated Ringer's solution and packed red blood cells (Hematocrit 40%). Trials were conducted at flow rates of 500 ml/min and 700 ml/min at 35ºC and 25ºC. Real-time pressure and flow data were recorded using a custom-based data acquisition system. For gaseous microemboli testing, 5cc of air were manually injected into the venous line. Gaseous microemboli were recorded using the Emboli Detection and Classification Quantifier. RESULTS: Braile Infant 1500 had a lower pressure drop (P<0.01) and a higher total hemodynamic energy delivered to the pseudopatient (P<0.01). However, there was a higher raw number of gaseous microemboli seen prior to oxygenator at lower temperatures with the Braile oxygenator compared to the Kids D100 (P<0.01). CONCLUSION: Braile Infant 1500 oxygenator had a better hemodynamic performance compared to the Dideco Kids D100 oxygenator. Braile had more gaseous microemboli detected at the pre-oxygenator site under hypothermia, but delivered a smaller percentage of air emboli to the pseudopatient than the Dideco oxygenator. Sociedade Brasileira de Cirurgia Cardiovascular 2016 /pmc/articles/PMC5144558/ /pubmed/27982342 http://dx.doi.org/10.5935/1678-9741.20160075 Text en http://creativecommons.org/licenses/by/4.0/ This is an Open Access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Original Articles
Marupudi, Neelima
Wang, Shigang
Canêo, Luiz Fernando
Jatene, Fabio Biscegli
Kunselman, Allen R.
Undar, Akif
In-Vitro Evaluation of Two Types of Neonatal Oxygenators in Handling Gaseous Microemboli and Maintaining Optimal Hemodynamic Stability During Cardiopulmonary Bypass
title In-Vitro Evaluation of Two Types of Neonatal Oxygenators in Handling Gaseous Microemboli and Maintaining Optimal Hemodynamic Stability During Cardiopulmonary Bypass
title_full In-Vitro Evaluation of Two Types of Neonatal Oxygenators in Handling Gaseous Microemboli and Maintaining Optimal Hemodynamic Stability During Cardiopulmonary Bypass
title_fullStr In-Vitro Evaluation of Two Types of Neonatal Oxygenators in Handling Gaseous Microemboli and Maintaining Optimal Hemodynamic Stability During Cardiopulmonary Bypass
title_full_unstemmed In-Vitro Evaluation of Two Types of Neonatal Oxygenators in Handling Gaseous Microemboli and Maintaining Optimal Hemodynamic Stability During Cardiopulmonary Bypass
title_short In-Vitro Evaluation of Two Types of Neonatal Oxygenators in Handling Gaseous Microemboli and Maintaining Optimal Hemodynamic Stability During Cardiopulmonary Bypass
title_sort in-vitro evaluation of two types of neonatal oxygenators in handling gaseous microemboli and maintaining optimal hemodynamic stability during cardiopulmonary bypass
topic Original Articles
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5144558/
https://www.ncbi.nlm.nih.gov/pubmed/27982342
http://dx.doi.org/10.5935/1678-9741.20160075
work_keys_str_mv AT marupudineelima invitroevaluationoftwotypesofneonataloxygenatorsinhandlinggaseousmicroemboliandmaintainingoptimalhemodynamicstabilityduringcardiopulmonarybypass
AT wangshigang invitroevaluationoftwotypesofneonataloxygenatorsinhandlinggaseousmicroemboliandmaintainingoptimalhemodynamicstabilityduringcardiopulmonarybypass
AT caneoluizfernando invitroevaluationoftwotypesofneonataloxygenatorsinhandlinggaseousmicroemboliandmaintainingoptimalhemodynamicstabilityduringcardiopulmonarybypass
AT jatenefabiobiscegli invitroevaluationoftwotypesofneonataloxygenatorsinhandlinggaseousmicroemboliandmaintainingoptimalhemodynamicstabilityduringcardiopulmonarybypass
AT kunselmanallenr invitroevaluationoftwotypesofneonataloxygenatorsinhandlinggaseousmicroemboliandmaintainingoptimalhemodynamicstabilityduringcardiopulmonarybypass
AT undarakif invitroevaluationoftwotypesofneonataloxygenatorsinhandlinggaseousmicroemboliandmaintainingoptimalhemodynamicstabilityduringcardiopulmonarybypass