Assessment of a 2D electronic portal imaging devices-based dosimetry algorithm for pretreatment and in-vivo midplane dose verification

BACKGROUND: The use of electronic portal imaging devices (EPIDs) is a method for the dosimetric verification of radiotherapy plans, both pretreatment and in vivo. The aim of this study is to test a 2D EPID-based dosimetry algorithm for dose verification of some plans inside a homogenous and anthropo...

Descripción completa

Detalles Bibliográficos
Autores principales: Jomehzadeh, Ali, Shokrani, Parvaneh, Mohammadi, Mohammad, Amouheidari, Alireza
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Medknow Publications & Media Pvt Ltd 2016
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5157032/
https://www.ncbi.nlm.nih.gov/pubmed/28028511
http://dx.doi.org/10.4103/2277-9175.194799
_version_ 1782481377207255040
author Jomehzadeh, Ali
Shokrani, Parvaneh
Mohammadi, Mohammad
Amouheidari, Alireza
author_facet Jomehzadeh, Ali
Shokrani, Parvaneh
Mohammadi, Mohammad
Amouheidari, Alireza
author_sort Jomehzadeh, Ali
collection PubMed
description BACKGROUND: The use of electronic portal imaging devices (EPIDs) is a method for the dosimetric verification of radiotherapy plans, both pretreatment and in vivo. The aim of this study is to test a 2D EPID-based dosimetry algorithm for dose verification of some plans inside a homogenous and anthropomorphic phantom and in vivo as well. MATERIALS AND METHODS: Dose distributions were reconstructed from EPID images using a 2D EPID dosimetry algorithm inside a homogenous slab phantom for a simple 10 × 10 cm(2) box technique, 3D conformal (prostate, head-and-neck, and lung), and intensity-modulated radiation therapy (IMRT) prostate plans inside an anthropomorphic (Alderson) phantom and in the patients (one fraction in vivo) for 3D conformal plans (prostate, head-and-neck and lung). RESULTS: The planned and EPID dose difference at the isocenter, on an average, was 1.7% for pretreatment verification and less than 3% for all in vivo plans, except for head-and-neck, which was 3.6%. The mean γ values for a seven-field prostate IMRT plan delivered to the Alderson phantom varied from 0.28 to 0.65. For 3D conformal plans applied for the Alderson phantom, all γ1% values were within the tolerance level for all plans and in both anteroposterior and posteroanterior (AP-PA) beams. CONCLUSION: The 2D EPID-based dosimetry algorithm provides an accurate method to verify the dose of a simple 10 × 10 cm(2) field, in two dimensions, inside a homogenous slab phantom and an IMRT prostate plan, as well as in 3D conformal plans (prostate, head-and-neck, and lung plans) applied using an anthropomorphic phantom and in vivo. However, further investigation to improve the 2D EPID dosimetry algorithm for a head-and-neck case, is necessary.
format Online
Article
Text
id pubmed-5157032
institution National Center for Biotechnology Information
language English
publishDate 2016
publisher Medknow Publications & Media Pvt Ltd
record_format MEDLINE/PubMed
spelling pubmed-51570322016-12-27 Assessment of a 2D electronic portal imaging devices-based dosimetry algorithm for pretreatment and in-vivo midplane dose verification Jomehzadeh, Ali Shokrani, Parvaneh Mohammadi, Mohammad Amouheidari, Alireza Adv Biomed Res Original Article BACKGROUND: The use of electronic portal imaging devices (EPIDs) is a method for the dosimetric verification of radiotherapy plans, both pretreatment and in vivo. The aim of this study is to test a 2D EPID-based dosimetry algorithm for dose verification of some plans inside a homogenous and anthropomorphic phantom and in vivo as well. MATERIALS AND METHODS: Dose distributions were reconstructed from EPID images using a 2D EPID dosimetry algorithm inside a homogenous slab phantom for a simple 10 × 10 cm(2) box technique, 3D conformal (prostate, head-and-neck, and lung), and intensity-modulated radiation therapy (IMRT) prostate plans inside an anthropomorphic (Alderson) phantom and in the patients (one fraction in vivo) for 3D conformal plans (prostate, head-and-neck and lung). RESULTS: The planned and EPID dose difference at the isocenter, on an average, was 1.7% for pretreatment verification and less than 3% for all in vivo plans, except for head-and-neck, which was 3.6%. The mean γ values for a seven-field prostate IMRT plan delivered to the Alderson phantom varied from 0.28 to 0.65. For 3D conformal plans applied for the Alderson phantom, all γ1% values were within the tolerance level for all plans and in both anteroposterior and posteroanterior (AP-PA) beams. CONCLUSION: The 2D EPID-based dosimetry algorithm provides an accurate method to verify the dose of a simple 10 × 10 cm(2) field, in two dimensions, inside a homogenous slab phantom and an IMRT prostate plan, as well as in 3D conformal plans (prostate, head-and-neck, and lung plans) applied using an anthropomorphic phantom and in vivo. However, further investigation to improve the 2D EPID dosimetry algorithm for a head-and-neck case, is necessary. Medknow Publications & Media Pvt Ltd 2016-11-28 /pmc/articles/PMC5157032/ /pubmed/28028511 http://dx.doi.org/10.4103/2277-9175.194799 Text en Copyright: © 2016 Jomehzadeh. http://creativecommons.org/licenses/by-nc-sa/3.0 This is an open-access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.
spellingShingle Original Article
Jomehzadeh, Ali
Shokrani, Parvaneh
Mohammadi, Mohammad
Amouheidari, Alireza
Assessment of a 2D electronic portal imaging devices-based dosimetry algorithm for pretreatment and in-vivo midplane dose verification
title Assessment of a 2D electronic portal imaging devices-based dosimetry algorithm for pretreatment and in-vivo midplane dose verification
title_full Assessment of a 2D electronic portal imaging devices-based dosimetry algorithm for pretreatment and in-vivo midplane dose verification
title_fullStr Assessment of a 2D electronic portal imaging devices-based dosimetry algorithm for pretreatment and in-vivo midplane dose verification
title_full_unstemmed Assessment of a 2D electronic portal imaging devices-based dosimetry algorithm for pretreatment and in-vivo midplane dose verification
title_short Assessment of a 2D electronic portal imaging devices-based dosimetry algorithm for pretreatment and in-vivo midplane dose verification
title_sort assessment of a 2d electronic portal imaging devices-based dosimetry algorithm for pretreatment and in-vivo midplane dose verification
topic Original Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5157032/
https://www.ncbi.nlm.nih.gov/pubmed/28028511
http://dx.doi.org/10.4103/2277-9175.194799
work_keys_str_mv AT jomehzadehali assessmentofa2delectronicportalimagingdevicesbaseddosimetryalgorithmforpretreatmentandinvivomidplanedoseverification
AT shokraniparvaneh assessmentofa2delectronicportalimagingdevicesbaseddosimetryalgorithmforpretreatmentandinvivomidplanedoseverification
AT mohammadimohammad assessmentofa2delectronicportalimagingdevicesbaseddosimetryalgorithmforpretreatmentandinvivomidplanedoseverification
AT amouheidarialireza assessmentofa2delectronicportalimagingdevicesbaseddosimetryalgorithmforpretreatmentandinvivomidplanedoseverification