Cargando…

Feasibility Study on Applying Radiophotoluminescent Glass Dosimeters for CyberKnife SRS Dose Verification

CyberKnife is one of multiple modalities for stereotactic radiosurgery (SRS). Due to the nature of CyberKnife and the characteristics of SRS, dose evaluation of the CyberKnife procedure is critical. A radiophotoluminescent glass dosimeter was used to verify the dose accuracy for the CyberKnife proce...

Descripción completa

Detalles Bibliográficos
Autores principales: Hsu, Shih-Ming, Hung, Chao-Hsiung, Liao, Yi-Jen, Fu, Hsiao-Mei, Tsai, Jo-Ting, Huang, Yung-Hui, Huang, David Y. C.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Public Library of Science 2017
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5207685/
https://www.ncbi.nlm.nih.gov/pubmed/28046056
http://dx.doi.org/10.1371/journal.pone.0169252
_version_ 1782490413019430912
author Hsu, Shih-Ming
Hung, Chao-Hsiung
Liao, Yi-Jen
Fu, Hsiao-Mei
Tsai, Jo-Ting
Huang, Yung-Hui
Huang, David Y. C.
author_facet Hsu, Shih-Ming
Hung, Chao-Hsiung
Liao, Yi-Jen
Fu, Hsiao-Mei
Tsai, Jo-Ting
Huang, Yung-Hui
Huang, David Y. C.
author_sort Hsu, Shih-Ming
collection PubMed
description CyberKnife is one of multiple modalities for stereotactic radiosurgery (SRS). Due to the nature of CyberKnife and the characteristics of SRS, dose evaluation of the CyberKnife procedure is critical. A radiophotoluminescent glass dosimeter was used to verify the dose accuracy for the CyberKnife procedure and validate a viable dose verification system for CyberKnife treatment. A radiophotoluminescent glass dosimeter, thermoluminescent dosimeter, and Kodak EDR2 film were used to measure the lateral dose profile and percent depth dose of CyberKnife. A Monte Carlo simulation for dose verification was performed using BEAMnrc to verify the measured results. This study also used a radiophotoluminescent glass dosimeter coupled with an anthropomorphic phantom to evaluate the accuracy of the dose given by CyberKnife. Measurements from the radiophotoluminescent glass dosimeter were compared with the results of a thermoluminescent dosimeter and EDR2 film, and the differences found were less than 5%. The radiophotoluminescent glass dosimeter has some advantages in terms of dose measurements over CyberKnife, such as repeatability, stability, and small effective size. These advantages make radiophotoluminescent glass dosimeters a potential candidate dosimeter for the CyberKnife procedure. This study concludes that radiophotoluminescent glass dosimeters are a promising and reliable dosimeter for CyberKnife dose verification with clinically acceptable accuracy within 5%.
format Online
Article
Text
id pubmed-5207685
institution National Center for Biotechnology Information
language English
publishDate 2017
publisher Public Library of Science
record_format MEDLINE/PubMed
spelling pubmed-52076852017-01-19 Feasibility Study on Applying Radiophotoluminescent Glass Dosimeters for CyberKnife SRS Dose Verification Hsu, Shih-Ming Hung, Chao-Hsiung Liao, Yi-Jen Fu, Hsiao-Mei Tsai, Jo-Ting Huang, Yung-Hui Huang, David Y. C. PLoS One Research Article CyberKnife is one of multiple modalities for stereotactic radiosurgery (SRS). Due to the nature of CyberKnife and the characteristics of SRS, dose evaluation of the CyberKnife procedure is critical. A radiophotoluminescent glass dosimeter was used to verify the dose accuracy for the CyberKnife procedure and validate a viable dose verification system for CyberKnife treatment. A radiophotoluminescent glass dosimeter, thermoluminescent dosimeter, and Kodak EDR2 film were used to measure the lateral dose profile and percent depth dose of CyberKnife. A Monte Carlo simulation for dose verification was performed using BEAMnrc to verify the measured results. This study also used a radiophotoluminescent glass dosimeter coupled with an anthropomorphic phantom to evaluate the accuracy of the dose given by CyberKnife. Measurements from the radiophotoluminescent glass dosimeter were compared with the results of a thermoluminescent dosimeter and EDR2 film, and the differences found were less than 5%. The radiophotoluminescent glass dosimeter has some advantages in terms of dose measurements over CyberKnife, such as repeatability, stability, and small effective size. These advantages make radiophotoluminescent glass dosimeters a potential candidate dosimeter for the CyberKnife procedure. This study concludes that radiophotoluminescent glass dosimeters are a promising and reliable dosimeter for CyberKnife dose verification with clinically acceptable accuracy within 5%. Public Library of Science 2017-01-03 /pmc/articles/PMC5207685/ /pubmed/28046056 http://dx.doi.org/10.1371/journal.pone.0169252 Text en © 2017 Hsu et al http://creativecommons.org/licenses/by/4.0/ This is an open access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/4.0/) , which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.
spellingShingle Research Article
Hsu, Shih-Ming
Hung, Chao-Hsiung
Liao, Yi-Jen
Fu, Hsiao-Mei
Tsai, Jo-Ting
Huang, Yung-Hui
Huang, David Y. C.
Feasibility Study on Applying Radiophotoluminescent Glass Dosimeters for CyberKnife SRS Dose Verification
title Feasibility Study on Applying Radiophotoluminescent Glass Dosimeters for CyberKnife SRS Dose Verification
title_full Feasibility Study on Applying Radiophotoluminescent Glass Dosimeters for CyberKnife SRS Dose Verification
title_fullStr Feasibility Study on Applying Radiophotoluminescent Glass Dosimeters for CyberKnife SRS Dose Verification
title_full_unstemmed Feasibility Study on Applying Radiophotoluminescent Glass Dosimeters for CyberKnife SRS Dose Verification
title_short Feasibility Study on Applying Radiophotoluminescent Glass Dosimeters for CyberKnife SRS Dose Verification
title_sort feasibility study on applying radiophotoluminescent glass dosimeters for cyberknife srs dose verification
topic Research Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5207685/
https://www.ncbi.nlm.nih.gov/pubmed/28046056
http://dx.doi.org/10.1371/journal.pone.0169252
work_keys_str_mv AT hsushihming feasibilitystudyonapplyingradiophotoluminescentglassdosimetersforcyberknifesrsdoseverification
AT hungchaohsiung feasibilitystudyonapplyingradiophotoluminescentglassdosimetersforcyberknifesrsdoseverification
AT liaoyijen feasibilitystudyonapplyingradiophotoluminescentglassdosimetersforcyberknifesrsdoseverification
AT fuhsiaomei feasibilitystudyonapplyingradiophotoluminescentglassdosimetersforcyberknifesrsdoseverification
AT tsaijoting feasibilitystudyonapplyingradiophotoluminescentglassdosimetersforcyberknifesrsdoseverification
AT huangyunghui feasibilitystudyonapplyingradiophotoluminescentglassdosimetersforcyberknifesrsdoseverification
AT huangdavidyc feasibilitystudyonapplyingradiophotoluminescentglassdosimetersforcyberknifesrsdoseverification