Establishment of local diagnostic reference levels for quality control in intraoral radiography
OBJECTIVE: To investigate the dosage and imaging conditions for patients undergoing intraoral radiography at Meikai University Hospital and establish assurance and quality control data. METHODS: Tube voltage, exposure time, and air kinetic energy released per unit mass (air kerma) of three intraoral...
Autores principales: | , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Springer Japan
2016
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5216092/ https://www.ncbi.nlm.nih.gov/pubmed/28111498 http://dx.doi.org/10.1007/s11282-016-0245-9 |
_version_ | 1782491861920776192 |
---|---|
author | Izawa, Maki Harata, Yasuo Shiba, Noriyoshi Koizumi, Nobuhide Ozawa, Tomonori Takahashi, Nobutoshi Okumura, Yasuhiko |
author_facet | Izawa, Maki Harata, Yasuo Shiba, Noriyoshi Koizumi, Nobuhide Ozawa, Tomonori Takahashi, Nobutoshi Okumura, Yasuhiko |
author_sort | Izawa, Maki |
collection | PubMed |
description | OBJECTIVE: To investigate the dosage and imaging conditions for patients undergoing intraoral radiography at Meikai University Hospital and establish assurance and quality control data. METHODS: Tube voltage, exposure time, and air kinetic energy released per unit mass (air kerma) of three intraoral radiography units were measured. To calculate the patient entrance dose (PED) for each radiograph using Insight film, we extracted data for 1063 patients from their exposure records. The PED was compared with the diagnostic reference level (DRL) from the European Commission and the UK. RESULTS: The tube voltage of the three units was maintained at 60 ± 2 kV. Differences in exposure time were less than 1.7 % for all units. The air kerma rates were well maintained within a 4.2 % error. Based on the patient data, there were no significant differences in the mean exposure times for males and females for all anatomical sites. The mean PED ranged from 1.09 ± 0.31 mGy for the mandibular incisors to 2.42 ± 0.33 mGy for the maxillary molars. The mean PED at the mandibular molars using InSight film was 1.59 ± 0.20 mGy, being less than the recommended value based on the DRL for intraoral radiography in the UK. CONCLUSIONS: We concluded that radiographic conditions at the hospital have been properly maintained. This basic quality control data may assist other dental radiation facilities to reduce patient dosage. |
format | Online Article Text |
id | pubmed-5216092 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2016 |
publisher | Springer Japan |
record_format | MEDLINE/PubMed |
spelling | pubmed-52160922017-01-18 Establishment of local diagnostic reference levels for quality control in intraoral radiography Izawa, Maki Harata, Yasuo Shiba, Noriyoshi Koizumi, Nobuhide Ozawa, Tomonori Takahashi, Nobutoshi Okumura, Yasuhiko Oral Radiol Original Article OBJECTIVE: To investigate the dosage and imaging conditions for patients undergoing intraoral radiography at Meikai University Hospital and establish assurance and quality control data. METHODS: Tube voltage, exposure time, and air kinetic energy released per unit mass (air kerma) of three intraoral radiography units were measured. To calculate the patient entrance dose (PED) for each radiograph using Insight film, we extracted data for 1063 patients from their exposure records. The PED was compared with the diagnostic reference level (DRL) from the European Commission and the UK. RESULTS: The tube voltage of the three units was maintained at 60 ± 2 kV. Differences in exposure time were less than 1.7 % for all units. The air kerma rates were well maintained within a 4.2 % error. Based on the patient data, there were no significant differences in the mean exposure times for males and females for all anatomical sites. The mean PED ranged from 1.09 ± 0.31 mGy for the mandibular incisors to 2.42 ± 0.33 mGy for the maxillary molars. The mean PED at the mandibular molars using InSight film was 1.59 ± 0.20 mGy, being less than the recommended value based on the DRL for intraoral radiography in the UK. CONCLUSIONS: We concluded that radiographic conditions at the hospital have been properly maintained. This basic quality control data may assist other dental radiation facilities to reduce patient dosage. Springer Japan 2016-05-03 2017 /pmc/articles/PMC5216092/ /pubmed/28111498 http://dx.doi.org/10.1007/s11282-016-0245-9 Text en © The Author(s) 2016 Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. |
spellingShingle | Original Article Izawa, Maki Harata, Yasuo Shiba, Noriyoshi Koizumi, Nobuhide Ozawa, Tomonori Takahashi, Nobutoshi Okumura, Yasuhiko Establishment of local diagnostic reference levels for quality control in intraoral radiography |
title | Establishment of local diagnostic reference levels for quality control in intraoral radiography |
title_full | Establishment of local diagnostic reference levels for quality control in intraoral radiography |
title_fullStr | Establishment of local diagnostic reference levels for quality control in intraoral radiography |
title_full_unstemmed | Establishment of local diagnostic reference levels for quality control in intraoral radiography |
title_short | Establishment of local diagnostic reference levels for quality control in intraoral radiography |
title_sort | establishment of local diagnostic reference levels for quality control in intraoral radiography |
topic | Original Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5216092/ https://www.ncbi.nlm.nih.gov/pubmed/28111498 http://dx.doi.org/10.1007/s11282-016-0245-9 |
work_keys_str_mv | AT izawamaki establishmentoflocaldiagnosticreferencelevelsforqualitycontrolinintraoralradiography AT haratayasuo establishmentoflocaldiagnosticreferencelevelsforqualitycontrolinintraoralradiography AT shibanoriyoshi establishmentoflocaldiagnosticreferencelevelsforqualitycontrolinintraoralradiography AT koizuminobuhide establishmentoflocaldiagnosticreferencelevelsforqualitycontrolinintraoralradiography AT ozawatomonori establishmentoflocaldiagnosticreferencelevelsforqualitycontrolinintraoralradiography AT takahashinobutoshi establishmentoflocaldiagnosticreferencelevelsforqualitycontrolinintraoralradiography AT okumurayasuhiko establishmentoflocaldiagnosticreferencelevelsforqualitycontrolinintraoralradiography |