Validation of the RisingSun RS‐651 Blood Pressure Monitor Based on Auscultation in Adults According to the ANSI/AAMI/ISO 81060‐2:2013 Standard
This study validated the RisingSun RS‐651 blood pressure (BP) monitor based on auscultation in adults according to the American National Standards Institute/Association for the Advancement of Medical Instrumentation/International Organization for Standardization (ANSI/AAMI/ISO) 81060‐2:2013 standard...
Autores principales: | , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
John Wiley and Sons Inc.
2016
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5216443/ https://www.ncbi.nlm.nih.gov/pubmed/27271472 http://dx.doi.org/10.1111/jch.12859 |
Sumario: | This study validated the RisingSun RS‐651 blood pressure (BP) monitor based on auscultation in adults according to the American National Standards Institute/Association for the Advancement of Medical Instrumentation/International Organization for Standardization (ANSI/AAMI/ISO) 81060‐2:2013 standard. The RS‐651 device was evaluated in a study of 97 participants. The same arm simultaneous method, as defined in the ANSI/AAMI/ISO standard, was used. The mean differences±standard deviation for criterion 1 were 0.8±2.3 mm Hg for systolic BP (SBP) and −0.1±2.9 mm Hg for diastolic BP (DBP). Analysis for criterion 2 resulted in values of 0.8±1.5 mm Hg for SBP and −0.1±2.1 mm Hg for DBP. All of the data fulfilled the ANSI/AAMI/ISO 81060‐2:2013 standard requirements to pass the validation. The RisingSun RS‐651 device can be recommended for both clinical and self/home use in adults according to the ANSI/AAMI/ISO 81060‐2:2013 standard. |
---|