Validation of the RisingSun RS‐651 Blood Pressure Monitor Based on Auscultation in Adults According to the ANSI/AAMI/ISO 81060‐2:2013 Standard
This study validated the RisingSun RS‐651 blood pressure (BP) monitor based on auscultation in adults according to the American National Standards Institute/Association for the Advancement of Medical Instrumentation/International Organization for Standardization (ANSI/AAMI/ISO) 81060‐2:2013 standard...
Autores principales: | , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
John Wiley and Sons Inc.
2016
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5216443/ https://www.ncbi.nlm.nih.gov/pubmed/27271472 http://dx.doi.org/10.1111/jch.12859 |
_version_ | 1782491924794441728 |
---|---|
author | She, Jin Guan, Xizhou Liu, Yanyong Xiang, Haiyan |
author_facet | She, Jin Guan, Xizhou Liu, Yanyong Xiang, Haiyan |
author_sort | She, Jin |
collection | PubMed |
description | This study validated the RisingSun RS‐651 blood pressure (BP) monitor based on auscultation in adults according to the American National Standards Institute/Association for the Advancement of Medical Instrumentation/International Organization for Standardization (ANSI/AAMI/ISO) 81060‐2:2013 standard. The RS‐651 device was evaluated in a study of 97 participants. The same arm simultaneous method, as defined in the ANSI/AAMI/ISO standard, was used. The mean differences±standard deviation for criterion 1 were 0.8±2.3 mm Hg for systolic BP (SBP) and −0.1±2.9 mm Hg for diastolic BP (DBP). Analysis for criterion 2 resulted in values of 0.8±1.5 mm Hg for SBP and −0.1±2.1 mm Hg for DBP. All of the data fulfilled the ANSI/AAMI/ISO 81060‐2:2013 standard requirements to pass the validation. The RisingSun RS‐651 device can be recommended for both clinical and self/home use in adults according to the ANSI/AAMI/ISO 81060‐2:2013 standard. |
format | Online Article Text |
id | pubmed-5216443 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2016 |
publisher | John Wiley and Sons Inc. |
record_format | MEDLINE/PubMed |
spelling | pubmed-52164432017-01-18 Validation of the RisingSun RS‐651 Blood Pressure Monitor Based on Auscultation in Adults According to the ANSI/AAMI/ISO 81060‐2:2013 Standard She, Jin Guan, Xizhou Liu, Yanyong Xiang, Haiyan J Clin Hypertens (Greenwich) Blood Pressure Measurement This study validated the RisingSun RS‐651 blood pressure (BP) monitor based on auscultation in adults according to the American National Standards Institute/Association for the Advancement of Medical Instrumentation/International Organization for Standardization (ANSI/AAMI/ISO) 81060‐2:2013 standard. The RS‐651 device was evaluated in a study of 97 participants. The same arm simultaneous method, as defined in the ANSI/AAMI/ISO standard, was used. The mean differences±standard deviation for criterion 1 were 0.8±2.3 mm Hg for systolic BP (SBP) and −0.1±2.9 mm Hg for diastolic BP (DBP). Analysis for criterion 2 resulted in values of 0.8±1.5 mm Hg for SBP and −0.1±2.1 mm Hg for DBP. All of the data fulfilled the ANSI/AAMI/ISO 81060‐2:2013 standard requirements to pass the validation. The RisingSun RS‐651 device can be recommended for both clinical and self/home use in adults according to the ANSI/AAMI/ISO 81060‐2:2013 standard. John Wiley and Sons Inc. 2016-06-07 /pmc/articles/PMC5216443/ /pubmed/27271472 http://dx.doi.org/10.1111/jch.12859 Text en © 2016 The Authors. The Journal of Clinical Hypertension Published by Wiley Periodicals, Inc. https://creativecommons.org/licenses/by-nc-nd/4.0/This is an open access article under the terms of the http://creativecommons.org/licenses/by-nc-nd/4.0/ (https://creativecommons.org/licenses/by-nc-nd/4.0/) License, which permits use and distribution in any medium, provided the original work is properly cited, the use is non‐commercial and no modifications or adaptations are made. |
spellingShingle | Blood Pressure Measurement She, Jin Guan, Xizhou Liu, Yanyong Xiang, Haiyan Validation of the RisingSun RS‐651 Blood Pressure Monitor Based on Auscultation in Adults According to the ANSI/AAMI/ISO 81060‐2:2013 Standard |
title | Validation of the RisingSun RS‐651 Blood Pressure Monitor Based on Auscultation in Adults According to the ANSI/AAMI/ISO 81060‐2:2013 Standard |
title_full | Validation of the RisingSun RS‐651 Blood Pressure Monitor Based on Auscultation in Adults According to the ANSI/AAMI/ISO 81060‐2:2013 Standard |
title_fullStr | Validation of the RisingSun RS‐651 Blood Pressure Monitor Based on Auscultation in Adults According to the ANSI/AAMI/ISO 81060‐2:2013 Standard |
title_full_unstemmed | Validation of the RisingSun RS‐651 Blood Pressure Monitor Based on Auscultation in Adults According to the ANSI/AAMI/ISO 81060‐2:2013 Standard |
title_short | Validation of the RisingSun RS‐651 Blood Pressure Monitor Based on Auscultation in Adults According to the ANSI/AAMI/ISO 81060‐2:2013 Standard |
title_sort | validation of the risingsun rs‐651 blood pressure monitor based on auscultation in adults according to the ansi/aami/iso 81060‐2:2013 standard |
topic | Blood Pressure Measurement |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5216443/ https://www.ncbi.nlm.nih.gov/pubmed/27271472 http://dx.doi.org/10.1111/jch.12859 |
work_keys_str_mv | AT shejin validationoftherisingsunrs651bloodpressuremonitorbasedonauscultationinadultsaccordingtotheansiaamiiso8106022013standard AT guanxizhou validationoftherisingsunrs651bloodpressuremonitorbasedonauscultationinadultsaccordingtotheansiaamiiso8106022013standard AT liuyanyong validationoftherisingsunrs651bloodpressuremonitorbasedonauscultationinadultsaccordingtotheansiaamiiso8106022013standard AT xianghaiyan validationoftherisingsunrs651bloodpressuremonitorbasedonauscultationinadultsaccordingtotheansiaamiiso8106022013standard |