Measuring shear-wave speed with point shear-wave elastography and MR elastography: a phantom study

OBJECTIVES: To compare shear-wave speed (SWS) measured by ultrasound-based point shear-wave elastography (pSWE) and MR elastography (MRE) on phantoms with a known shear modulus, and to assess method validity and variability. METHODS: 5 homogeneous phantoms of different stiffnesses were made. Shear m...

Descripción completa

Detalles Bibliográficos
Autores principales: Kishimoto, Riwa, Suga, Mikio, Koyama, Atsuhisa, Omatsu, Tokuhiko, Tachibana, Yasuhiko, Ebner, Daniel K, Obata, Takayuki
Formato: Online Artículo Texto
Lenguaje:English
Publicado: BMJ Publishing Group 2017
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5223661/
https://www.ncbi.nlm.nih.gov/pubmed/28057657
http://dx.doi.org/10.1136/bmjopen-2016-013925
Descripción
Sumario:OBJECTIVES: To compare shear-wave speed (SWS) measured by ultrasound-based point shear-wave elastography (pSWE) and MR elastography (MRE) on phantoms with a known shear modulus, and to assess method validity and variability. METHODS: 5 homogeneous phantoms of different stiffnesses were made. Shear modulus was measured by a rheometer, and this value was used as the standard. 10 SWS measurements were obtained at 4 different depths with 1.0–4.5 MHz convex (4C1) and 4.0–9.0 MHz linear (9L4) transducers using pSWE. MRE was carried out once per phantom, and SWSs at 5 different depths were obtained. These SWSs were then compared with those from a rheometer using linear regression analyses. RESULTS: SWSs obtained with both pSWE as well as MRE had a strong correlation with those obtained by a rheometer (R(2)>0.97). The relative difference in SWS between the procedures was from −25.2% to 25.6% for all phantoms, and from −8.1% to 6.9% when the softest and hardest phantoms were excluded. Depth dependency was noted in the 9L4 transducer of pSWE and MRE. CONCLUSIONS: SWSs from pSWE and MRE showed a good correlation with a rheometer-determined SWS. Although based on phantom studies, SWSs obtained with these methods are not always equivalent, the measurement can be thought of as reliable and these SWSs were reasonably close to each other for the middle range of stiffness within the measurable range.