On-line estimations of delivered radiation doses in three-dimensional conformal radiotherapy treatments of carcinoma uterine cervix patients in linear accelerator

Transmission of radiation fluence through patient's body has a correlation to the planned target dose. A method to estimate the delivered dose to target volumes was standardized using a beam level 0.6 cc ionization chamber (IC) positioned at electronic portal imaging device (EPID) plane from th...

Descripción completa

Detalles Bibliográficos
Autores principales: Putha, Suman Kumar, Saxena, P. U., Banerjee, S., Srinivas, Challapalli, Vadhiraja, B. M., Ravichandran, Ramamoorthy, Joan, Mary, Pai, K. Dinesh
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Medknow Publications & Media Pvt Ltd 2016
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5228045/
https://www.ncbi.nlm.nih.gov/pubmed/28144114
http://dx.doi.org/10.4103/0971-6203.195186
_version_ 1782493911178018816
author Putha, Suman Kumar
Saxena, P. U.
Banerjee, S.
Srinivas, Challapalli
Vadhiraja, B. M.
Ravichandran, Ramamoorthy
Joan, Mary
Pai, K. Dinesh
author_facet Putha, Suman Kumar
Saxena, P. U.
Banerjee, S.
Srinivas, Challapalli
Vadhiraja, B. M.
Ravichandran, Ramamoorthy
Joan, Mary
Pai, K. Dinesh
author_sort Putha, Suman Kumar
collection PubMed
description Transmission of radiation fluence through patient's body has a correlation to the planned target dose. A method to estimate the delivered dose to target volumes was standardized using a beam level 0.6 cc ionization chamber (IC) positioned at electronic portal imaging device (EPID) plane from the measured transit signal (S(t)) in patients with cancer of uterine cervix treated with three-dimensional conformal radiotherapy (3DCRT). The IC with buildup cap was mounted on linear accelerator EPID frame with fixed source to chamber distance of 146.3 cm, using a locally fabricated mount. S(t)s were obtained for different water phantom thicknesses and radiation field sizes which were then used to generate a calibration table against calculated midplane doses at isocenter (D(iso,TPS)), derived from the treatment planning system. A code was developed using MATLAB software which was used to estimate the in vivo dose at isocenter (D(iso,Transit)) from the measured S(t)s. A locally fabricated pelvic phantom validated the estimations of D(iso,Transit) before implementing this method on actual patients. On-line dose estimations were made (3 times during treatment for each patient) in 24 patients. The D(iso,Transit) agreement with D(iso,TPS) in phantom was within 1.7% and the mean percentage deviation with standard deviation is −1.37% ±2.03% (n = 72) observed in patients. Estimated in vivo dose at isocenter with this method provides a good agreement with planned ones which can be implemented as part of quality assurance in pelvic sites treated with simple techniques, for example, 3DCRT where there is a need for documentation of planned dose delivery.
format Online
Article
Text
id pubmed-5228045
institution National Center for Biotechnology Information
language English
publishDate 2016
publisher Medknow Publications & Media Pvt Ltd
record_format MEDLINE/PubMed
spelling pubmed-52280452017-01-31 On-line estimations of delivered radiation doses in three-dimensional conformal radiotherapy treatments of carcinoma uterine cervix patients in linear accelerator Putha, Suman Kumar Saxena, P. U. Banerjee, S. Srinivas, Challapalli Vadhiraja, B. M. Ravichandran, Ramamoorthy Joan, Mary Pai, K. Dinesh J Med Phys Original Article Transmission of radiation fluence through patient's body has a correlation to the planned target dose. A method to estimate the delivered dose to target volumes was standardized using a beam level 0.6 cc ionization chamber (IC) positioned at electronic portal imaging device (EPID) plane from the measured transit signal (S(t)) in patients with cancer of uterine cervix treated with three-dimensional conformal radiotherapy (3DCRT). The IC with buildup cap was mounted on linear accelerator EPID frame with fixed source to chamber distance of 146.3 cm, using a locally fabricated mount. S(t)s were obtained for different water phantom thicknesses and radiation field sizes which were then used to generate a calibration table against calculated midplane doses at isocenter (D(iso,TPS)), derived from the treatment planning system. A code was developed using MATLAB software which was used to estimate the in vivo dose at isocenter (D(iso,Transit)) from the measured S(t)s. A locally fabricated pelvic phantom validated the estimations of D(iso,Transit) before implementing this method on actual patients. On-line dose estimations were made (3 times during treatment for each patient) in 24 patients. The D(iso,Transit) agreement with D(iso,TPS) in phantom was within 1.7% and the mean percentage deviation with standard deviation is −1.37% ±2.03% (n = 72) observed in patients. Estimated in vivo dose at isocenter with this method provides a good agreement with planned ones which can be implemented as part of quality assurance in pelvic sites treated with simple techniques, for example, 3DCRT where there is a need for documentation of planned dose delivery. Medknow Publications & Media Pvt Ltd 2016 /pmc/articles/PMC5228045/ /pubmed/28144114 http://dx.doi.org/10.4103/0971-6203.195186 Text en Copyright: © 2016 Journal of Medical Physics http://creativecommons.org/licenses/by-nc-sa/3.0 This is an open access article distributed under the terms of the Creative Commons Attribution-NonCommercial-ShareAlike 3.0 License, which allows others to remix, tweak, and build upon the work non-commercially, as long as the author is credited and the new creations are licensed under the identical terms.
spellingShingle Original Article
Putha, Suman Kumar
Saxena, P. U.
Banerjee, S.
Srinivas, Challapalli
Vadhiraja, B. M.
Ravichandran, Ramamoorthy
Joan, Mary
Pai, K. Dinesh
On-line estimations of delivered radiation doses in three-dimensional conformal radiotherapy treatments of carcinoma uterine cervix patients in linear accelerator
title On-line estimations of delivered radiation doses in three-dimensional conformal radiotherapy treatments of carcinoma uterine cervix patients in linear accelerator
title_full On-line estimations of delivered radiation doses in three-dimensional conformal radiotherapy treatments of carcinoma uterine cervix patients in linear accelerator
title_fullStr On-line estimations of delivered radiation doses in three-dimensional conformal radiotherapy treatments of carcinoma uterine cervix patients in linear accelerator
title_full_unstemmed On-line estimations of delivered radiation doses in three-dimensional conformal radiotherapy treatments of carcinoma uterine cervix patients in linear accelerator
title_short On-line estimations of delivered radiation doses in three-dimensional conformal radiotherapy treatments of carcinoma uterine cervix patients in linear accelerator
title_sort on-line estimations of delivered radiation doses in three-dimensional conformal radiotherapy treatments of carcinoma uterine cervix patients in linear accelerator
topic Original Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5228045/
https://www.ncbi.nlm.nih.gov/pubmed/28144114
http://dx.doi.org/10.4103/0971-6203.195186
work_keys_str_mv AT puthasumankumar onlineestimationsofdeliveredradiationdosesinthreedimensionalconformalradiotherapytreatmentsofcarcinomauterinecervixpatientsinlinearaccelerator
AT saxenapu onlineestimationsofdeliveredradiationdosesinthreedimensionalconformalradiotherapytreatmentsofcarcinomauterinecervixpatientsinlinearaccelerator
AT banerjees onlineestimationsofdeliveredradiationdosesinthreedimensionalconformalradiotherapytreatmentsofcarcinomauterinecervixpatientsinlinearaccelerator
AT srinivaschallapalli onlineestimationsofdeliveredradiationdosesinthreedimensionalconformalradiotherapytreatmentsofcarcinomauterinecervixpatientsinlinearaccelerator
AT vadhirajabm onlineestimationsofdeliveredradiationdosesinthreedimensionalconformalradiotherapytreatmentsofcarcinomauterinecervixpatientsinlinearaccelerator
AT ravichandranramamoorthy onlineestimationsofdeliveredradiationdosesinthreedimensionalconformalradiotherapytreatmentsofcarcinomauterinecervixpatientsinlinearaccelerator
AT joanmary onlineestimationsofdeliveredradiationdosesinthreedimensionalconformalradiotherapytreatmentsofcarcinomauterinecervixpatientsinlinearaccelerator
AT paikdinesh onlineestimationsofdeliveredradiationdosesinthreedimensionalconformalradiotherapytreatmentsofcarcinomauterinecervixpatientsinlinearaccelerator