The effect of fresh gas flow rate and type of anesthesia machine on time to reach target sevoflurane concentration
BACKGROUND: Anesthesia machines have been developed by the application of new technology for rapid and easier control of anesthetic concentration. In this study, we used a test lung to investigate whether the time taken to reach the target sevoflurane concentration varies with the rate of fresh gas...
Autores principales: | , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
BioMed Central
2017
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5248460/ https://www.ncbi.nlm.nih.gov/pubmed/28103806 http://dx.doi.org/10.1186/s12871-016-0294-y |
_version_ | 1782497271886118912 |
---|---|
author | Shin, Hye Won Yu, Hae Na Bae, Go Eun Huh, Hyub Park, Ji Yong Kim, Ji Young |
author_facet | Shin, Hye Won Yu, Hae Na Bae, Go Eun Huh, Hyub Park, Ji Yong Kim, Ji Young |
author_sort | Shin, Hye Won |
collection | PubMed |
description | BACKGROUND: Anesthesia machines have been developed by the application of new technology for rapid and easier control of anesthetic concentration. In this study, we used a test lung to investigate whether the time taken to reach the target sevoflurane concentration varies with the rate of fresh gas flow (FGF) and type of anesthesia machine (AM). METHODS: We measured the times taken to reach the target sevoflurane concentration (2 minimum alveolar concentration = 4%) at variable rates of FGF (0.5, 1, or 3 L/min) and different types of AM (Primus(®), Perseus(®), and Zeus(®) [Zeus(®)-F; Zeus(®) fresh gas mode, Zeus(®)-A; Zeus(®) auto-mode]). Concomitant ventilation was supplied using 100% O(2.) The AMs were connected to a test lung. A sevoflurane vaporizer setting of 6% was used in Primus(®), Perseus(®), and Zeus(®)-F; a target end-tidal setting of 4% was used in Zeus(®)-A (from a vaporizer setting of 0%). The time taken to reach the target concentration was measured in every group. RESULTS: When the same AM was used (Primus(®), Perseus(®), or Zeus(®)-F), the times to target concentration shortened as the FGF rate increased (P < 0.05). Conversely, when the same FGF rate was used, but with different AMs, the time to target concentration was shortest in Perseus(®), followed by Primus(®), and finally by Zeus(®)-F (P < 0.05). With regards to both modes of Zeus(®), at FGF rates of 0.5 and 1 L/min, the time to target concentration was shorter in Zeus(®)-A than in Zeus(®)-F; however, the time was longer in Zeus(®)-A than in Zeus(®)-F at FGF rate of 3 L/min (P < 0.05). CONCLUSION: Shorter times taken to reach the target concentration were associated with high FGF rates, smaller internal volume of the AM, proximity of the fresh gas inlets to patients, absence of a decoupling system, and use of blower-driven ventilators in AM. |
format | Online Article Text |
id | pubmed-5248460 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2017 |
publisher | BioMed Central |
record_format | MEDLINE/PubMed |
spelling | pubmed-52484602017-01-25 The effect of fresh gas flow rate and type of anesthesia machine on time to reach target sevoflurane concentration Shin, Hye Won Yu, Hae Na Bae, Go Eun Huh, Hyub Park, Ji Yong Kim, Ji Young BMC Anesthesiol Research Article BACKGROUND: Anesthesia machines have been developed by the application of new technology for rapid and easier control of anesthetic concentration. In this study, we used a test lung to investigate whether the time taken to reach the target sevoflurane concentration varies with the rate of fresh gas flow (FGF) and type of anesthesia machine (AM). METHODS: We measured the times taken to reach the target sevoflurane concentration (2 minimum alveolar concentration = 4%) at variable rates of FGF (0.5, 1, or 3 L/min) and different types of AM (Primus(®), Perseus(®), and Zeus(®) [Zeus(®)-F; Zeus(®) fresh gas mode, Zeus(®)-A; Zeus(®) auto-mode]). Concomitant ventilation was supplied using 100% O(2.) The AMs were connected to a test lung. A sevoflurane vaporizer setting of 6% was used in Primus(®), Perseus(®), and Zeus(®)-F; a target end-tidal setting of 4% was used in Zeus(®)-A (from a vaporizer setting of 0%). The time taken to reach the target concentration was measured in every group. RESULTS: When the same AM was used (Primus(®), Perseus(®), or Zeus(®)-F), the times to target concentration shortened as the FGF rate increased (P < 0.05). Conversely, when the same FGF rate was used, but with different AMs, the time to target concentration was shortest in Perseus(®), followed by Primus(®), and finally by Zeus(®)-F (P < 0.05). With regards to both modes of Zeus(®), at FGF rates of 0.5 and 1 L/min, the time to target concentration was shorter in Zeus(®)-A than in Zeus(®)-F; however, the time was longer in Zeus(®)-A than in Zeus(®)-F at FGF rate of 3 L/min (P < 0.05). CONCLUSION: Shorter times taken to reach the target concentration were associated with high FGF rates, smaller internal volume of the AM, proximity of the fresh gas inlets to patients, absence of a decoupling system, and use of blower-driven ventilators in AM. BioMed Central 2017-01-19 /pmc/articles/PMC5248460/ /pubmed/28103806 http://dx.doi.org/10.1186/s12871-016-0294-y Text en © The Author(s). 2017 Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated. |
spellingShingle | Research Article Shin, Hye Won Yu, Hae Na Bae, Go Eun Huh, Hyub Park, Ji Yong Kim, Ji Young The effect of fresh gas flow rate and type of anesthesia machine on time to reach target sevoflurane concentration |
title | The effect of fresh gas flow rate and type of anesthesia machine on time to reach target sevoflurane concentration |
title_full | The effect of fresh gas flow rate and type of anesthesia machine on time to reach target sevoflurane concentration |
title_fullStr | The effect of fresh gas flow rate and type of anesthesia machine on time to reach target sevoflurane concentration |
title_full_unstemmed | The effect of fresh gas flow rate and type of anesthesia machine on time to reach target sevoflurane concentration |
title_short | The effect of fresh gas flow rate and type of anesthesia machine on time to reach target sevoflurane concentration |
title_sort | effect of fresh gas flow rate and type of anesthesia machine on time to reach target sevoflurane concentration |
topic | Research Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5248460/ https://www.ncbi.nlm.nih.gov/pubmed/28103806 http://dx.doi.org/10.1186/s12871-016-0294-y |
work_keys_str_mv | AT shinhyewon theeffectoffreshgasflowrateandtypeofanesthesiamachineontimetoreachtargetsevofluraneconcentration AT yuhaena theeffectoffreshgasflowrateandtypeofanesthesiamachineontimetoreachtargetsevofluraneconcentration AT baegoeun theeffectoffreshgasflowrateandtypeofanesthesiamachineontimetoreachtargetsevofluraneconcentration AT huhhyub theeffectoffreshgasflowrateandtypeofanesthesiamachineontimetoreachtargetsevofluraneconcentration AT parkjiyong theeffectoffreshgasflowrateandtypeofanesthesiamachineontimetoreachtargetsevofluraneconcentration AT kimjiyoung theeffectoffreshgasflowrateandtypeofanesthesiamachineontimetoreachtargetsevofluraneconcentration AT shinhyewon effectoffreshgasflowrateandtypeofanesthesiamachineontimetoreachtargetsevofluraneconcentration AT yuhaena effectoffreshgasflowrateandtypeofanesthesiamachineontimetoreachtargetsevofluraneconcentration AT baegoeun effectoffreshgasflowrateandtypeofanesthesiamachineontimetoreachtargetsevofluraneconcentration AT huhhyub effectoffreshgasflowrateandtypeofanesthesiamachineontimetoreachtargetsevofluraneconcentration AT parkjiyong effectoffreshgasflowrateandtypeofanesthesiamachineontimetoreachtargetsevofluraneconcentration AT kimjiyoung effectoffreshgasflowrateandtypeofanesthesiamachineontimetoreachtargetsevofluraneconcentration |