Cargando…
Micro X-Ray Computed Tomography Mass Loss Assessment of Different UHMWPE: A Hip Joint Simulator Study on Standard vs. Cross-Linked Polyethylene
More than 60.000 hip arthroplasty are performed every year in Italy. Although Ultra-High-Molecular-Weight-Polyethylene remains the most used material as acetabular cup, wear of this material induces over time in vivo a foreign-body response and consequently osteolysis, pain, and the need of implant...
Autores principales: | , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Public Library of Science
2017
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5249200/ https://www.ncbi.nlm.nih.gov/pubmed/28107468 http://dx.doi.org/10.1371/journal.pone.0170263 |
_version_ | 1782497413840240640 |
---|---|
author | Affatato, Saverio Zanini, Filippo Carmignato, Simone |
author_facet | Affatato, Saverio Zanini, Filippo Carmignato, Simone |
author_sort | Affatato, Saverio |
collection | PubMed |
description | More than 60.000 hip arthroplasty are performed every year in Italy. Although Ultra-High-Molecular-Weight-Polyethylene remains the most used material as acetabular cup, wear of this material induces over time in vivo a foreign-body response and consequently osteolysis, pain, and the need of implant revision. Furthermore, oxidative wear of the polyethylene provoke several and severe failures. To solve these problems, highly cross-linked polyethylene and Vitamin-E-stabilized polyethylene were introduced in the last years. In in vitro experiments, various efforts have been made to compare the wear behavior of standard PE and vitamin-E infused liners. In this study we compared the in vitro wear behavior of two different configurations of cross-linked polyethylene (with and without the add of Vitamin E) vs. the standard polyethylene acetabular cups. The aim of the present study was to validate a micro X-ray computed tomography technique to assess the wear of different commercially available, polyethylene’s acetabular cups after wear simulation; in particular, the gravimetric method was used to provide reference wear values. The agreement between the two methods is documented in this paper. |
format | Online Article Text |
id | pubmed-5249200 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2017 |
publisher | Public Library of Science |
record_format | MEDLINE/PubMed |
spelling | pubmed-52492002017-02-06 Micro X-Ray Computed Tomography Mass Loss Assessment of Different UHMWPE: A Hip Joint Simulator Study on Standard vs. Cross-Linked Polyethylene Affatato, Saverio Zanini, Filippo Carmignato, Simone PLoS One Research Article More than 60.000 hip arthroplasty are performed every year in Italy. Although Ultra-High-Molecular-Weight-Polyethylene remains the most used material as acetabular cup, wear of this material induces over time in vivo a foreign-body response and consequently osteolysis, pain, and the need of implant revision. Furthermore, oxidative wear of the polyethylene provoke several and severe failures. To solve these problems, highly cross-linked polyethylene and Vitamin-E-stabilized polyethylene were introduced in the last years. In in vitro experiments, various efforts have been made to compare the wear behavior of standard PE and vitamin-E infused liners. In this study we compared the in vitro wear behavior of two different configurations of cross-linked polyethylene (with and without the add of Vitamin E) vs. the standard polyethylene acetabular cups. The aim of the present study was to validate a micro X-ray computed tomography technique to assess the wear of different commercially available, polyethylene’s acetabular cups after wear simulation; in particular, the gravimetric method was used to provide reference wear values. The agreement between the two methods is documented in this paper. Public Library of Science 2017-01-20 /pmc/articles/PMC5249200/ /pubmed/28107468 http://dx.doi.org/10.1371/journal.pone.0170263 Text en © 2017 Affatato et al http://creativecommons.org/licenses/by/4.0/ This is an open access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/4.0/) , which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited. |
spellingShingle | Research Article Affatato, Saverio Zanini, Filippo Carmignato, Simone Micro X-Ray Computed Tomography Mass Loss Assessment of Different UHMWPE: A Hip Joint Simulator Study on Standard vs. Cross-Linked Polyethylene |
title | Micro X-Ray Computed Tomography Mass Loss Assessment of Different UHMWPE: A Hip Joint Simulator Study on Standard vs. Cross-Linked Polyethylene |
title_full | Micro X-Ray Computed Tomography Mass Loss Assessment of Different UHMWPE: A Hip Joint Simulator Study on Standard vs. Cross-Linked Polyethylene |
title_fullStr | Micro X-Ray Computed Tomography Mass Loss Assessment of Different UHMWPE: A Hip Joint Simulator Study on Standard vs. Cross-Linked Polyethylene |
title_full_unstemmed | Micro X-Ray Computed Tomography Mass Loss Assessment of Different UHMWPE: A Hip Joint Simulator Study on Standard vs. Cross-Linked Polyethylene |
title_short | Micro X-Ray Computed Tomography Mass Loss Assessment of Different UHMWPE: A Hip Joint Simulator Study on Standard vs. Cross-Linked Polyethylene |
title_sort | micro x-ray computed tomography mass loss assessment of different uhmwpe: a hip joint simulator study on standard vs. cross-linked polyethylene |
topic | Research Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5249200/ https://www.ncbi.nlm.nih.gov/pubmed/28107468 http://dx.doi.org/10.1371/journal.pone.0170263 |
work_keys_str_mv | AT affatatosaverio microxraycomputedtomographymasslossassessmentofdifferentuhmwpeahipjointsimulatorstudyonstandardvscrosslinkedpolyethylene AT zaninifilippo microxraycomputedtomographymasslossassessmentofdifferentuhmwpeahipjointsimulatorstudyonstandardvscrosslinkedpolyethylene AT carmignatosimone microxraycomputedtomographymasslossassessmentofdifferentuhmwpeahipjointsimulatorstudyonstandardvscrosslinkedpolyethylene |