Cargando…
Simultaneous ramp right heart catheterization and echocardiography in a ReliantHeart left ventricular assist device
Many clinicians caring for patients with continuous flow left ventricular assist devices (CF-LVAD) use ramp right heart catheterization (RHC) studies to optimize pump speed and also to troubleshoot CF-LVAD malfunction. An investigational device, the ReliantHeart Heart Assist 5 (Houston, TX), provide...
Autores principales: | , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Baishideng Publishing Group Inc
2017
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5253195/ https://www.ncbi.nlm.nih.gov/pubmed/28163837 http://dx.doi.org/10.4330/wjc.v9.i1.55 |
_version_ | 1782498133056421888 |
---|---|
author | Banerjee, Dipanjan Dutt, Debleena Duclos, Sebastien Sallam, Karim Wheeler, Matthew Ha, Richard |
author_facet | Banerjee, Dipanjan Dutt, Debleena Duclos, Sebastien Sallam, Karim Wheeler, Matthew Ha, Richard |
author_sort | Banerjee, Dipanjan |
collection | PubMed |
description | Many clinicians caring for patients with continuous flow left ventricular assist devices (CF-LVAD) use ramp right heart catheterization (RHC) studies to optimize pump speed and also to troubleshoot CF-LVAD malfunction. An investigational device, the ReliantHeart Heart Assist 5 (Houston, TX), provides the added benefit of an ultrasonic flow probe on the outflow graft that directly measures flow through the CF-LVAD. We performed a simultaneous ramp RHC and echocardiogram on a patient who received the above CF-LVAD to optimize pump parameters and investigate elevated flow through the CF-LVAD as measured by the flow probe. We found that the patient’s hemodynamics were optimized at their baseline pump speed, and that the measured cardiac output via the Fick principle was lower than that measured by the flow probe. Right heart catheterization may be useful to investigate discrepancies between flow measured by a CF-LVAD and a patient’s clinical presentation, particularly in investigational devices where little clinical experience exists. More data is needed to elucidate the correlation between the flow measured by an ultrasonic probe and cardiac output as measured by RHC. |
format | Online Article Text |
id | pubmed-5253195 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2017 |
publisher | Baishideng Publishing Group Inc |
record_format | MEDLINE/PubMed |
spelling | pubmed-52531952017-02-04 Simultaneous ramp right heart catheterization and echocardiography in a ReliantHeart left ventricular assist device Banerjee, Dipanjan Dutt, Debleena Duclos, Sebastien Sallam, Karim Wheeler, Matthew Ha, Richard World J Cardiol Case Report Many clinicians caring for patients with continuous flow left ventricular assist devices (CF-LVAD) use ramp right heart catheterization (RHC) studies to optimize pump speed and also to troubleshoot CF-LVAD malfunction. An investigational device, the ReliantHeart Heart Assist 5 (Houston, TX), provides the added benefit of an ultrasonic flow probe on the outflow graft that directly measures flow through the CF-LVAD. We performed a simultaneous ramp RHC and echocardiogram on a patient who received the above CF-LVAD to optimize pump parameters and investigate elevated flow through the CF-LVAD as measured by the flow probe. We found that the patient’s hemodynamics were optimized at their baseline pump speed, and that the measured cardiac output via the Fick principle was lower than that measured by the flow probe. Right heart catheterization may be useful to investigate discrepancies between flow measured by a CF-LVAD and a patient’s clinical presentation, particularly in investigational devices where little clinical experience exists. More data is needed to elucidate the correlation between the flow measured by an ultrasonic probe and cardiac output as measured by RHC. Baishideng Publishing Group Inc 2017-01-26 2017-01-26 /pmc/articles/PMC5253195/ /pubmed/28163837 http://dx.doi.org/10.4330/wjc.v9.i1.55 Text en ©The Author(s) 2016. Published by Baishideng Publishing Group Inc. All rights reserved. http://creativecommons.org/licenses/by-nc/4.0/ Open-Access: This article is an open-access article which was selected by an in-house editor and fully peer-reviewed by external reviewers. It is distributed in accordance with the Creative Commons Attribution Non Commercial (CC BY-NC 4.0) license, which permits others to distribute, remix, adapt, build upon this work non-commercially, and license their derivative works on different terms, provided the original work is properly cited and the use is non-commercial. See: http://creativecommons.org/licenses/by-nc/4.0/ |
spellingShingle | Case Report Banerjee, Dipanjan Dutt, Debleena Duclos, Sebastien Sallam, Karim Wheeler, Matthew Ha, Richard Simultaneous ramp right heart catheterization and echocardiography in a ReliantHeart left ventricular assist device |
title | Simultaneous ramp right heart catheterization and echocardiography in a ReliantHeart left ventricular assist device |
title_full | Simultaneous ramp right heart catheterization and echocardiography in a ReliantHeart left ventricular assist device |
title_fullStr | Simultaneous ramp right heart catheterization and echocardiography in a ReliantHeart left ventricular assist device |
title_full_unstemmed | Simultaneous ramp right heart catheterization and echocardiography in a ReliantHeart left ventricular assist device |
title_short | Simultaneous ramp right heart catheterization and echocardiography in a ReliantHeart left ventricular assist device |
title_sort | simultaneous ramp right heart catheterization and echocardiography in a reliantheart left ventricular assist device |
topic | Case Report |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5253195/ https://www.ncbi.nlm.nih.gov/pubmed/28163837 http://dx.doi.org/10.4330/wjc.v9.i1.55 |
work_keys_str_mv | AT banerjeedipanjan simultaneousramprightheartcatheterizationandechocardiographyinareliantheartleftventricularassistdevice AT duttdebleena simultaneousramprightheartcatheterizationandechocardiographyinareliantheartleftventricularassistdevice AT duclossebastien simultaneousramprightheartcatheterizationandechocardiographyinareliantheartleftventricularassistdevice AT sallamkarim simultaneousramprightheartcatheterizationandechocardiographyinareliantheartleftventricularassistdevice AT wheelermatthew simultaneousramprightheartcatheterizationandechocardiographyinareliantheartleftventricularassistdevice AT harichard simultaneousramprightheartcatheterizationandechocardiographyinareliantheartleftventricularassistdevice |