Tx/Rx Head Coil Induces Less RF Transmit-Related Heating than Body Coil in Conductive Metallic Objects Outside the Active Area of the Head Coil

The transmit–receive (Tx/Rx) birdcage head coil is often used for excitation instead of the body coil because of the presumably lower risk of heating in and around conductive implants. However, this common practice has not been systematically tested. To investigate whether the Tx/Rx birdcage head co...

Descripción completa

Detalles Bibliográficos
Autores principales: Nagy, Zoltan, Oliver-Taylor, Aaron, Kuehne, Andre, Goluch, Sigrun, Weiskopf, Nikolaus
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Frontiers Media S.A. 2017
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5266708/
https://www.ncbi.nlm.nih.gov/pubmed/28184184
http://dx.doi.org/10.3389/fnins.2017.00015
_version_ 1782500496098983936
author Nagy, Zoltan
Oliver-Taylor, Aaron
Kuehne, Andre
Goluch, Sigrun
Weiskopf, Nikolaus
author_facet Nagy, Zoltan
Oliver-Taylor, Aaron
Kuehne, Andre
Goluch, Sigrun
Weiskopf, Nikolaus
author_sort Nagy, Zoltan
collection PubMed
description The transmit–receive (Tx/Rx) birdcage head coil is often used for excitation instead of the body coil because of the presumably lower risk of heating in and around conductive implants. However, this common practice has not been systematically tested. To investigate whether the Tx/Rx birdcage head coil produces less heating than the body coil when scanning individuals with implants, we used a 3T clinical scanner and made temperature measurements around a straight 15 cm conductor using either the Tx/Rx body or the head coil for excitation. Additionally, the transmitted fields of a Tx/Rx head coil were measured both in air and in gel using a resonant and a non-resonant B field probes as well as a non-resonant E field probe. Simulations using a finite-difference time domain solver were compared with the experimental findings. When the body coil was used for excitation, we observed heating around the 15 cm wire at various anatomical locations (both within and outside of the active volume of the head coil). Outside its active area, no such heating was observed while using the Tx/Rx head coil for excitation. The E and B fields of the Tx/Rx birdcage head coil extended well-beyond the physical dimensions of the coil. In air, the fields were monotonically decreasing, while in gel they were more complex with local maxima at the end of the ASTM phantom. These experimental findings were line with the simulations. While caution must always be exercised when scanning individuals with metallic implants, these findings support the use of the Tx/Rx birdcage head coil in place of the body coil at 3T in order to reduce the risk of heating in and around conductive implants that are remote from the head coil.
format Online
Article
Text
id pubmed-5266708
institution National Center for Biotechnology Information
language English
publishDate 2017
publisher Frontiers Media S.A.
record_format MEDLINE/PubMed
spelling pubmed-52667082017-02-09 Tx/Rx Head Coil Induces Less RF Transmit-Related Heating than Body Coil in Conductive Metallic Objects Outside the Active Area of the Head Coil Nagy, Zoltan Oliver-Taylor, Aaron Kuehne, Andre Goluch, Sigrun Weiskopf, Nikolaus Front Neurosci Neuroscience The transmit–receive (Tx/Rx) birdcage head coil is often used for excitation instead of the body coil because of the presumably lower risk of heating in and around conductive implants. However, this common practice has not been systematically tested. To investigate whether the Tx/Rx birdcage head coil produces less heating than the body coil when scanning individuals with implants, we used a 3T clinical scanner and made temperature measurements around a straight 15 cm conductor using either the Tx/Rx body or the head coil for excitation. Additionally, the transmitted fields of a Tx/Rx head coil were measured both in air and in gel using a resonant and a non-resonant B field probes as well as a non-resonant E field probe. Simulations using a finite-difference time domain solver were compared with the experimental findings. When the body coil was used for excitation, we observed heating around the 15 cm wire at various anatomical locations (both within and outside of the active volume of the head coil). Outside its active area, no such heating was observed while using the Tx/Rx head coil for excitation. The E and B fields of the Tx/Rx birdcage head coil extended well-beyond the physical dimensions of the coil. In air, the fields were monotonically decreasing, while in gel they were more complex with local maxima at the end of the ASTM phantom. These experimental findings were line with the simulations. While caution must always be exercised when scanning individuals with metallic implants, these findings support the use of the Tx/Rx birdcage head coil in place of the body coil at 3T in order to reduce the risk of heating in and around conductive implants that are remote from the head coil. Frontiers Media S.A. 2017-01-26 /pmc/articles/PMC5266708/ /pubmed/28184184 http://dx.doi.org/10.3389/fnins.2017.00015 Text en Copyright © 2017 Nagy, Oliver-Taylor, Kuehne, Goluch and Weiskopf. http://creativecommons.org/licenses/by/4.0/ This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) or licensor are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.
spellingShingle Neuroscience
Nagy, Zoltan
Oliver-Taylor, Aaron
Kuehne, Andre
Goluch, Sigrun
Weiskopf, Nikolaus
Tx/Rx Head Coil Induces Less RF Transmit-Related Heating than Body Coil in Conductive Metallic Objects Outside the Active Area of the Head Coil
title Tx/Rx Head Coil Induces Less RF Transmit-Related Heating than Body Coil in Conductive Metallic Objects Outside the Active Area of the Head Coil
title_full Tx/Rx Head Coil Induces Less RF Transmit-Related Heating than Body Coil in Conductive Metallic Objects Outside the Active Area of the Head Coil
title_fullStr Tx/Rx Head Coil Induces Less RF Transmit-Related Heating than Body Coil in Conductive Metallic Objects Outside the Active Area of the Head Coil
title_full_unstemmed Tx/Rx Head Coil Induces Less RF Transmit-Related Heating than Body Coil in Conductive Metallic Objects Outside the Active Area of the Head Coil
title_short Tx/Rx Head Coil Induces Less RF Transmit-Related Heating than Body Coil in Conductive Metallic Objects Outside the Active Area of the Head Coil
title_sort tx/rx head coil induces less rf transmit-related heating than body coil in conductive metallic objects outside the active area of the head coil
topic Neuroscience
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5266708/
https://www.ncbi.nlm.nih.gov/pubmed/28184184
http://dx.doi.org/10.3389/fnins.2017.00015
work_keys_str_mv AT nagyzoltan txrxheadcoilinduceslessrftransmitrelatedheatingthanbodycoilinconductivemetallicobjectsoutsidetheactiveareaoftheheadcoil
AT olivertayloraaron txrxheadcoilinduceslessrftransmitrelatedheatingthanbodycoilinconductivemetallicobjectsoutsidetheactiveareaoftheheadcoil
AT kuehneandre txrxheadcoilinduceslessrftransmitrelatedheatingthanbodycoilinconductivemetallicobjectsoutsidetheactiveareaoftheheadcoil
AT goluchsigrun txrxheadcoilinduceslessrftransmitrelatedheatingthanbodycoilinconductivemetallicobjectsoutsidetheactiveareaoftheheadcoil
AT weiskopfnikolaus txrxheadcoilinduceslessrftransmitrelatedheatingthanbodycoilinconductivemetallicobjectsoutsidetheactiveareaoftheheadcoil