Performance of FACSPresto Point-of-Care Instrument for CD4-T Cell Enumeration in Human Immunodeficiency Virus (HIV)-Infected Patients Attending Care and Treatment Clinics in Belgium and Tanzania
BACKGROUND: CD4 T-cell counts are widely used to assess treatment eligibility and to follow-up HIV-infected patients. The World Health Organization prequalification of in vitro diagnostics program conducted a performance evaluation of the FACSPresto (BD Biosciences), a new point-of-care instrument t...
Autores principales: | , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Public Library of Science
2017
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5271305/ https://www.ncbi.nlm.nih.gov/pubmed/28129324 http://dx.doi.org/10.1371/journal.pone.0170248 |
_version_ | 1782501325368459264 |
---|---|
author | Daneau, Géraldine Aboud, Said Prat, Irena Urassa, Willy Kestens, Luc |
author_facet | Daneau, Géraldine Aboud, Said Prat, Irena Urassa, Willy Kestens, Luc |
author_sort | Daneau, Géraldine |
collection | PubMed |
description | BACKGROUND: CD4 T-cell counts are widely used to assess treatment eligibility and to follow-up HIV-infected patients. The World Health Organization prequalification of in vitro diagnostics program conducted a performance evaluation of the FACSPresto (BD Biosciences), a new point-of-care instrument to measure absolute CD4-T cell (CD4) counts and percentages in venous and capillary blood samples from HIV-infected patients. METHODS: Patients were recruited in Belgium (200 patients) and in Tanzania (247 patients). Venous blood samples were analyzed in two nearby reference laboratories. In addition, nurses/technicians collected a capillary blood sample by finger prick directly into a FACSPresto CD4 cartridge. Assay precision was assessed on fresh blood and on external quality control samples. Trueness (bias) was assessed by comparing results from FACSPresto with the reference (single-platform FACSCalibur). Clinical misclassification was measured at 200, 350 and 500 cells/μL thresholds. RESULTS: Intra-assay precision was < 6%, and inter-assay < 8%. CD4 results from FACSPresto and reference method resulted in regression slopes of 0.99–1.11 using either venous or capillary blood. Correlation was better for venous than for capillary blood (minimum 0.97 vs 0.93 respectively). Capillary blood resulted in a larger bias than venous blood, with 24 and 83 cells/μL for absolute CD4 counts on capillary blood in Antwerp and Dar es Salaam respectively, vs 12 and 41 cells/μL on venous blood. Bias on CD4% was < 1% on both venous and capillary blood, and was proportionally better than for absolute CD4 counts. Clinical misclassification was in line with the average overestimation, showing a very good specificity, but sensitivity around 70–90%. The rejection rate was 11% on first reading, leading to 6% of all samples without final result after a second reading. CONCLUSIONS: The FACSPresto performed very well on venous blood samples, and well on capillary blood samples. |
format | Online Article Text |
id | pubmed-5271305 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2017 |
publisher | Public Library of Science |
record_format | MEDLINE/PubMed |
spelling | pubmed-52713052017-02-06 Performance of FACSPresto Point-of-Care Instrument for CD4-T Cell Enumeration in Human Immunodeficiency Virus (HIV)-Infected Patients Attending Care and Treatment Clinics in Belgium and Tanzania Daneau, Géraldine Aboud, Said Prat, Irena Urassa, Willy Kestens, Luc PLoS One Research Article BACKGROUND: CD4 T-cell counts are widely used to assess treatment eligibility and to follow-up HIV-infected patients. The World Health Organization prequalification of in vitro diagnostics program conducted a performance evaluation of the FACSPresto (BD Biosciences), a new point-of-care instrument to measure absolute CD4-T cell (CD4) counts and percentages in venous and capillary blood samples from HIV-infected patients. METHODS: Patients were recruited in Belgium (200 patients) and in Tanzania (247 patients). Venous blood samples were analyzed in two nearby reference laboratories. In addition, nurses/technicians collected a capillary blood sample by finger prick directly into a FACSPresto CD4 cartridge. Assay precision was assessed on fresh blood and on external quality control samples. Trueness (bias) was assessed by comparing results from FACSPresto with the reference (single-platform FACSCalibur). Clinical misclassification was measured at 200, 350 and 500 cells/μL thresholds. RESULTS: Intra-assay precision was < 6%, and inter-assay < 8%. CD4 results from FACSPresto and reference method resulted in regression slopes of 0.99–1.11 using either venous or capillary blood. Correlation was better for venous than for capillary blood (minimum 0.97 vs 0.93 respectively). Capillary blood resulted in a larger bias than venous blood, with 24 and 83 cells/μL for absolute CD4 counts on capillary blood in Antwerp and Dar es Salaam respectively, vs 12 and 41 cells/μL on venous blood. Bias on CD4% was < 1% on both venous and capillary blood, and was proportionally better than for absolute CD4 counts. Clinical misclassification was in line with the average overestimation, showing a very good specificity, but sensitivity around 70–90%. The rejection rate was 11% on first reading, leading to 6% of all samples without final result after a second reading. CONCLUSIONS: The FACSPresto performed very well on venous blood samples, and well on capillary blood samples. Public Library of Science 2017-01-27 /pmc/articles/PMC5271305/ /pubmed/28129324 http://dx.doi.org/10.1371/journal.pone.0170248 Text en © 2017 Daneau et al http://creativecommons.org/licenses/by/4.0/ This is an open access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/4.0/) , which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited. |
spellingShingle | Research Article Daneau, Géraldine Aboud, Said Prat, Irena Urassa, Willy Kestens, Luc Performance of FACSPresto Point-of-Care Instrument for CD4-T Cell Enumeration in Human Immunodeficiency Virus (HIV)-Infected Patients Attending Care and Treatment Clinics in Belgium and Tanzania |
title | Performance of FACSPresto Point-of-Care Instrument for CD4-T Cell Enumeration in Human Immunodeficiency Virus (HIV)-Infected Patients Attending Care and Treatment Clinics in Belgium and Tanzania |
title_full | Performance of FACSPresto Point-of-Care Instrument for CD4-T Cell Enumeration in Human Immunodeficiency Virus (HIV)-Infected Patients Attending Care and Treatment Clinics in Belgium and Tanzania |
title_fullStr | Performance of FACSPresto Point-of-Care Instrument for CD4-T Cell Enumeration in Human Immunodeficiency Virus (HIV)-Infected Patients Attending Care and Treatment Clinics in Belgium and Tanzania |
title_full_unstemmed | Performance of FACSPresto Point-of-Care Instrument for CD4-T Cell Enumeration in Human Immunodeficiency Virus (HIV)-Infected Patients Attending Care and Treatment Clinics in Belgium and Tanzania |
title_short | Performance of FACSPresto Point-of-Care Instrument for CD4-T Cell Enumeration in Human Immunodeficiency Virus (HIV)-Infected Patients Attending Care and Treatment Clinics in Belgium and Tanzania |
title_sort | performance of facspresto point-of-care instrument for cd4-t cell enumeration in human immunodeficiency virus (hiv)-infected patients attending care and treatment clinics in belgium and tanzania |
topic | Research Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5271305/ https://www.ncbi.nlm.nih.gov/pubmed/28129324 http://dx.doi.org/10.1371/journal.pone.0170248 |
work_keys_str_mv | AT daneaugeraldine performanceoffacsprestopointofcareinstrumentforcd4tcellenumerationinhumanimmunodeficiencyvirushivinfectedpatientsattendingcareandtreatmentclinicsinbelgiumandtanzania AT aboudsaid performanceoffacsprestopointofcareinstrumentforcd4tcellenumerationinhumanimmunodeficiencyvirushivinfectedpatientsattendingcareandtreatmentclinicsinbelgiumandtanzania AT pratirena performanceoffacsprestopointofcareinstrumentforcd4tcellenumerationinhumanimmunodeficiencyvirushivinfectedpatientsattendingcareandtreatmentclinicsinbelgiumandtanzania AT urassawilly performanceoffacsprestopointofcareinstrumentforcd4tcellenumerationinhumanimmunodeficiencyvirushivinfectedpatientsattendingcareandtreatmentclinicsinbelgiumandtanzania AT kestensluc performanceoffacsprestopointofcareinstrumentforcd4tcellenumerationinhumanimmunodeficiencyvirushivinfectedpatientsattendingcareandtreatmentclinicsinbelgiumandtanzania |