Note on the Choice of a Sensitivity Weight In Precision Weighing

Good weighing practice usually dictates that, when using double-substitution weighing to determine the mass difference between two weights, the nominal value of the sensitivity weight used to calibrate the optical scale of the mass comparator be at least four times greater than the difference of the...

Descripción completa

Detalles Bibliográficos
Autor principal: Davis, R. S.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: [Gaithersburg, MD] : U.S. Dept. of Commerce, National Institute of Standards and Technology 1987
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5286969/
http://dx.doi.org/10.6028/jres.092.020
Descripción
Sumario:Good weighing practice usually dictates that, when using double-substitution weighing to determine the mass difference between two weights, the nominal value of the sensitivity weight used to calibrate the optical scale of the mass comparator be at least four times greater than the difference of the two weights being compared. However, there are times when other considerations must override this rule. We examine the theoretical basis for the rule and the penalty for violating it. Finally, we propose a modi-fied weighing scheme which imposes a much less stringent rule for the size of the sensitivity weight. The new scheme requires an additional balance reading, but does not increase the overall measurement time significantly.