Cargando…

Free-Field Reciprocity Calibration Of Microphones

Standardized methods for the primary free-field calibration of laboratory standard microphones deal with Type L (ANSI S1.10-1967, R1977) “one-inch” diameter microphones. However, the use of “1/2-inch” diameter microphones for measurement of the sound pressure level in acoustic fields is increasing....

Descripción completa

Detalles Bibliográficos
Autores principales: Burnett, Edwin D., Nedzelnitsky, Victor
Formato: Online Artículo Texto
Lenguaje:English
Publicado: [Gaithersburg, MD] : U.S. Dept. of Commerce, National Institute of Standards and Technology 1987
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5286998/
http://dx.doi.org/10.6028/jres.092.013
_version_ 1782504088869535744
author Burnett, Edwin D.
Nedzelnitsky, Victor
author_facet Burnett, Edwin D.
Nedzelnitsky, Victor
author_sort Burnett, Edwin D.
collection PubMed
description Standardized methods for the primary free-field calibration of laboratory standard microphones deal with Type L (ANSI S1.10-1967, R1977) “one-inch” diameter microphones. However, the use of “1/2-inch” diameter microphones for measurement of the sound pressure level in acoustic fields is increasing. Consequently, the NBS has developed a fixed-cost measurement service for the free-field calibration of these microphones by the reciprocity method over the range 2.5 kHz to 20 kHz. For this service, the apparatus and procedures, including essential properties of the anechoic chamber in which the calibrations are performed, are described. Opportunities for improvements are noted. The frequency-dependent positions of the apparent acoustic centers of the microphones were obtained. The overall uncertainty estimate for free-field calibration, expressed as the sum of the magnitude of credible bounds on the systematic component (s) and the random component (2σ, where σ is the standard deviation) is 0.16 dB or better (s=0.06 dB, 2σ=0.10 dB) at frequencies 1.25 kHz ⩽f⩽ 5kHz, and 0.07 dB or better (s=0.02 dB, 2σ=0.05 dB) for 5 kHz ⩽f⩽ 20 kHz. Comparison for given microphones of the measured difference between free-field and pressure response levels with the difference calculated by diffraction theory (derived by Matsui) indicates agreement of 0.16 dB or better in the low-frequency range (1.25 kHz to about 4 kHz) where free-field reciprocity measurements encounter the greatest experimental difficulties. This agreement is consistent with the estimated uncertainties of free-field and pressure calibration by the reciprocity method.
format Online
Article
Text
id pubmed-5286998
institution National Center for Biotechnology Information
language English
publishDate 1987
publisher [Gaithersburg, MD] : U.S. Dept. of Commerce, National Institute of Standards and Technology
record_format MEDLINE/PubMed
spelling pubmed-52869982021-08-02 Free-Field Reciprocity Calibration Of Microphones Burnett, Edwin D. Nedzelnitsky, Victor J Res Natl Bur Stand (1977) Article Standardized methods for the primary free-field calibration of laboratory standard microphones deal with Type L (ANSI S1.10-1967, R1977) “one-inch” diameter microphones. However, the use of “1/2-inch” diameter microphones for measurement of the sound pressure level in acoustic fields is increasing. Consequently, the NBS has developed a fixed-cost measurement service for the free-field calibration of these microphones by the reciprocity method over the range 2.5 kHz to 20 kHz. For this service, the apparatus and procedures, including essential properties of the anechoic chamber in which the calibrations are performed, are described. Opportunities for improvements are noted. The frequency-dependent positions of the apparent acoustic centers of the microphones were obtained. The overall uncertainty estimate for free-field calibration, expressed as the sum of the magnitude of credible bounds on the systematic component (s) and the random component (2σ, where σ is the standard deviation) is 0.16 dB or better (s=0.06 dB, 2σ=0.10 dB) at frequencies 1.25 kHz ⩽f⩽ 5kHz, and 0.07 dB or better (s=0.02 dB, 2σ=0.05 dB) for 5 kHz ⩽f⩽ 20 kHz. Comparison for given microphones of the measured difference between free-field and pressure response levels with the difference calculated by diffraction theory (derived by Matsui) indicates agreement of 0.16 dB or better in the low-frequency range (1.25 kHz to about 4 kHz) where free-field reciprocity measurements encounter the greatest experimental difficulties. This agreement is consistent with the estimated uncertainties of free-field and pressure calibration by the reciprocity method. [Gaithersburg, MD] : U.S. Dept. of Commerce, National Institute of Standards and Technology 1987 1987-04-01 /pmc/articles/PMC5286998/ http://dx.doi.org/10.6028/jres.092.013 Text en https://creativecommons.org/publicdomain/zero/1.0/ The Journal of Research of the National Bureau of Standards is a publication of the U.S. Government. The papers are in the public domain and are not subject to copyright in the United States. Articles from J Res may contain photographs or illustrations copyrighted by other commercial organizations or individuals that may not be used without obtaining prior approval from the holder of the copyright.
spellingShingle Article
Burnett, Edwin D.
Nedzelnitsky, Victor
Free-Field Reciprocity Calibration Of Microphones
title Free-Field Reciprocity Calibration Of Microphones
title_full Free-Field Reciprocity Calibration Of Microphones
title_fullStr Free-Field Reciprocity Calibration Of Microphones
title_full_unstemmed Free-Field Reciprocity Calibration Of Microphones
title_short Free-Field Reciprocity Calibration Of Microphones
title_sort free-field reciprocity calibration of microphones
topic Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5286998/
http://dx.doi.org/10.6028/jres.092.013
work_keys_str_mv AT burnettedwind freefieldreciprocitycalibrationofmicrophones
AT nedzelnitskyvictor freefieldreciprocitycalibrationofmicrophones