Precise Coulometric Titration of Acids and Bases

Apparatus and techniques are described for the titration of acids and bases by constant-current coulometry. The precision and accuracy of the method are indicated by titrations of benzoic acid, potassium acid phthalate, adipic acid, hydrochloric acid, and sodium carbonate. Standard deviations of ±0....

Descripción completa

Detalles Bibliográficos
Autores principales: Taylor, John K., Smith, Stanley W.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: [Gaithersburg, MD] : U.S. Dept. of Commerce, National Institute of Standards and Technology 1959
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5287197/
https://www.ncbi.nlm.nih.gov/pubmed/31216146
http://dx.doi.org/10.6028/jres.063A.008
Descripción
Sumario:Apparatus and techniques are described for the titration of acids and bases by constant-current coulometry. The precision and accuracy of the method are indicated by titrations of benzoic acid, potassium acid phthalate, adipic acid, hydrochloric acid, and sodium carbonate. Standard deviations of ±0.003 percent have been obtained which indicate that the reliability of the method is equal to or exceeds that of classical analytical procedures.