On the Actinometric Measurement of Absolute Luminescence Quantum Yields

The theory of the measurement of luminescent quantum yields using chemical actinometry is described. The sample’s emission intensity is measured by nearly completely surrounding the sample with an actinometer solution, and the excitation intensity is directly measured with the same type of actinomet...

Descripción completa

Detalles Bibliográficos
Autores principales: Demas, J. N., Blumenthal, B. H.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: [Gaithersburg, MD] : U.S. Dept. of Commerce, National Institute of Standards and Technology 1976
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5293347/
https://www.ncbi.nlm.nih.gov/pubmed/32196269
http://dx.doi.org/10.6028/jres.080A.041
Descripción
Sumario:The theory of the measurement of luminescent quantum yields using chemical actinometry is described. The sample’s emission intensity is measured by nearly completely surrounding the sample with an actinometer solution, and the excitation intensity is directly measured with the same type of actinometer. The ratio of the measured sample emission intensity corrected for the fraction escaping through the excitation ports to the measured excitation intensity is the absolute luminescence yield. Equations, a suitable cell design, and computer calculated correction factors for different cell dimensions and optical densities are given. The absolute yield of the actinometer is not needed, only its relative response with wavelength. New quantum-flat actinometers which should greatly simplify the measurements are described.