Humidity Fixed Points of Binary Saturated Aqueous Solutions
An evaluated compilation of equilibrium relative humidities in air versus temperature from pure phase to approximately 10(5) pascal (1 atm) in pressure is presented for 28 binary saturated aqueous solutions. The relative humidities of the solutions range from about 3 to 98 percent. Using a data base...
Autor principal: | |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
[Gaithersburg, MD] : U.S. Dept. of Commerce, National Institute of Standards and Technology
1977
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5295834/ http://dx.doi.org/10.6028/jres.081A.011 |
Sumario: | An evaluated compilation of equilibrium relative humidities in air versus temperature from pure phase to approximately 10(5) pascal (1 atm) in pressure is presented for 28 binary saturated aqueous solutions. The relative humidities of the solutions range from about 3 to 98 percent. Using a data base from 21 separate investigations comprising 1106 individual measurements, fits were made by the method of least squares to regular polynomial equations with two through four coefficients. Equations and tables are presented along with the estimated uncertainties in the correlated results. |
---|