Initial Experience and Feasibility of the New Low-Profile Stingray Catheter as Part of the Antegrade Dissection and Re-Entry Revascularization Strategy for Coronary Chronic Total Occlusions

Case series Patient: Male, 77 • Male, 57 • Male, 73 Final Diagnosis: Coronary chronic total occlusion Symptoms: Angina pectoris Medication: — Clinical Procedure: Percutaneous coronary intervention of coronary chronic total occlusions Specialty: Cardiology OBJECTIVE: Unusual setting of medical care B...

Descripción completa

Detalles Bibliográficos
Autores principales: Maeremans, Joren, Palmers, Pieter-Jan, Dens, Joseph
Formato: Online Artículo Texto
Lenguaje:English
Publicado: International Scientific Literature, Inc. 2017
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5297402/
https://www.ncbi.nlm.nih.gov/pubmed/28138123
http://dx.doi.org/10.12659/AJCR.902178
Descripción
Sumario:Case series Patient: Male, 77 • Male, 57 • Male, 73 Final Diagnosis: Coronary chronic total occlusion Symptoms: Angina pectoris Medication: — Clinical Procedure: Percutaneous coronary intervention of coronary chronic total occlusions Specialty: Cardiology OBJECTIVE: Unusual setting of medical care BACKGROUND: During antegrade dissection re-entry (ADR) of chronic total occlusions (CTO), the first-generation Stingray catheter requires the use of large-bore guides (sheathless 7.5 Fr or 8 Fr), which increases the risk for access site-related complications and compromises radial approaches. Smaller guiding sizes necessitate long guidewires (e.g., 300 cm) or guidewire extensions for catheter advancement or removal. However, friction between guides and the Stingray catheter can result in unstable guidewire position or unintentional removal. Furthermore, failure to deliver the catheter at the distal re-entry zone is a common problem. To overcome issues of deliverability and reduce the need for pre-dilatations, with its inherent risk of creating subintimal hematomas, the Stingray low-profile (LP) balloon catheter was developed. CASE REPORT: We describe 3 cases of successful application of the novel Stingray LP catheter during ADR. In all cases, 7 Fr guiding catheters were successfully used in combination with the device. The lower profile facilitated a good exchange and delivery of the device, without the need for balloon pre-dilatations in 2 cases. This resulted in a limited subintimal plane, enabling a smooth puncture into the true lumen. One case presented with extreme levels of calcification and tortuosity, resulting in a high degree of friction, despite the lower catheter profile. No in-hospital coronary or access site-related complications occurred. CONCLUSIONS: This case report illustrates the feasibility of the Stingray LP catheter for the treatment of CTOs via the ADR technique. The lower profile of the catheter potentially increases the deliverability, safety, and exchangeability of the device.