Reproducibility of pulmonary blood flow measurements by phase-contrast MRI using different 1.5 T MR scanners at two institutions

BACKGROUND: Magnetic resonance imaging (MRI) can be beneficial for diagnosis of disease by offering quantitative information. However, reproducibility can be a major problem when there is a numerical threshold in multi-institution, multi-vendor situations. PURPOSE: To measure pulmonary blood flow wi...

Descripción completa

Detalles Bibliográficos
Autores principales: Iraha, Rin, Tsuchiya, Nanae, Yamashiro, Tsuneo, Iwasawa, Tae, Murayama, Sadayuki
Formato: Online Artículo Texto
Lenguaje:English
Publicado: SAGE Publications 2017
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5298552/
https://www.ncbi.nlm.nih.gov/pubmed/28210495
http://dx.doi.org/10.1177/2058460116684370
_version_ 1782505881648234496
author Iraha, Rin
Tsuchiya, Nanae
Yamashiro, Tsuneo
Iwasawa, Tae
Murayama, Sadayuki
author_facet Iraha, Rin
Tsuchiya, Nanae
Yamashiro, Tsuneo
Iwasawa, Tae
Murayama, Sadayuki
author_sort Iraha, Rin
collection PubMed
description BACKGROUND: Magnetic resonance imaging (MRI) can be beneficial for diagnosis of disease by offering quantitative information. However, reproducibility can be a major problem when there is a numerical threshold in multi-institution, multi-vendor situations. PURPOSE: To measure pulmonary blood flow with phase-contrast (PC) imaging using two different MR scanners (1.5 T) at different institutions in the same participants and to examine the reproducibility of the measurements. MATERIAL AND METHODS: Participants were 10 healthy volunteers (5 men; age range, 27–36 years). The measurements included the mean and maximal blood velocities, the mean blood flow volume, and the acceleration time and volume (AT and AV), derived from the time-flow curve of the PC-MRI. Simultaneously obtained maximal, minimal, and mean areas from regions of interest set in the pulmonary artery were also calculated. In order to calculate the reproducibility of the quantitative variables, intra-class correlation coefficients (ICCs) were employed. When an adequate ICC was obtained, Bland–Altman analysis was conducted to identify any systematic bias. RESULTS: The ICCs were almost perfect for the mean blood flow volume and the AV (r = 0.82 and 0.80), and were substantial in the mean and maximal areas, and the AT (r = 0.63, 0.74, and 0.64, respectively). However, there was a fixed bias in the area measurement between the two scanners. Also, the AV had a proportional bias. CONCLUSION: Our results reveal that various indices derived from PC-MRI on different MR scanners are promising as common indices for pulmonary flow assessment. Research and clinical use of PC-MRI for the pulmonary artery is expected to extend to multi-institution situations.
format Online
Article
Text
id pubmed-5298552
institution National Center for Biotechnology Information
language English
publishDate 2017
publisher SAGE Publications
record_format MEDLINE/PubMed
spelling pubmed-52985522017-02-16 Reproducibility of pulmonary blood flow measurements by phase-contrast MRI using different 1.5 T MR scanners at two institutions Iraha, Rin Tsuchiya, Nanae Yamashiro, Tsuneo Iwasawa, Tae Murayama, Sadayuki Acta Radiol Open Research BACKGROUND: Magnetic resonance imaging (MRI) can be beneficial for diagnosis of disease by offering quantitative information. However, reproducibility can be a major problem when there is a numerical threshold in multi-institution, multi-vendor situations. PURPOSE: To measure pulmonary blood flow with phase-contrast (PC) imaging using two different MR scanners (1.5 T) at different institutions in the same participants and to examine the reproducibility of the measurements. MATERIAL AND METHODS: Participants were 10 healthy volunteers (5 men; age range, 27–36 years). The measurements included the mean and maximal blood velocities, the mean blood flow volume, and the acceleration time and volume (AT and AV), derived from the time-flow curve of the PC-MRI. Simultaneously obtained maximal, minimal, and mean areas from regions of interest set in the pulmonary artery were also calculated. In order to calculate the reproducibility of the quantitative variables, intra-class correlation coefficients (ICCs) were employed. When an adequate ICC was obtained, Bland–Altman analysis was conducted to identify any systematic bias. RESULTS: The ICCs were almost perfect for the mean blood flow volume and the AV (r = 0.82 and 0.80), and were substantial in the mean and maximal areas, and the AT (r = 0.63, 0.74, and 0.64, respectively). However, there was a fixed bias in the area measurement between the two scanners. Also, the AV had a proportional bias. CONCLUSION: Our results reveal that various indices derived from PC-MRI on different MR scanners are promising as common indices for pulmonary flow assessment. Research and clinical use of PC-MRI for the pulmonary artery is expected to extend to multi-institution situations. SAGE Publications 2017-01-01 /pmc/articles/PMC5298552/ /pubmed/28210495 http://dx.doi.org/10.1177/2058460116684370 Text en © The Foundation Acta Radiologica 2017 http://creativecommons.org/licenses/by-nc/3.0/ This article is distributed under the terms of the Creative Commons Attribution-NonCommercial 3.0 License (http://www.creativecommons.org/licenses/by-nc/3.0/) which permits non-commercial use, reproduction and distribution of the work without further permission provided the original work is attributed as specified on the SAGE and Open Access pages (https://us.sagepub.com/en-us/nam/open-access-at-sage).
spellingShingle Research
Iraha, Rin
Tsuchiya, Nanae
Yamashiro, Tsuneo
Iwasawa, Tae
Murayama, Sadayuki
Reproducibility of pulmonary blood flow measurements by phase-contrast MRI using different 1.5 T MR scanners at two institutions
title Reproducibility of pulmonary blood flow measurements by phase-contrast MRI using different 1.5 T MR scanners at two institutions
title_full Reproducibility of pulmonary blood flow measurements by phase-contrast MRI using different 1.5 T MR scanners at two institutions
title_fullStr Reproducibility of pulmonary blood flow measurements by phase-contrast MRI using different 1.5 T MR scanners at two institutions
title_full_unstemmed Reproducibility of pulmonary blood flow measurements by phase-contrast MRI using different 1.5 T MR scanners at two institutions
title_short Reproducibility of pulmonary blood flow measurements by phase-contrast MRI using different 1.5 T MR scanners at two institutions
title_sort reproducibility of pulmonary blood flow measurements by phase-contrast mri using different 1.5 t mr scanners at two institutions
topic Research
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5298552/
https://www.ncbi.nlm.nih.gov/pubmed/28210495
http://dx.doi.org/10.1177/2058460116684370
work_keys_str_mv AT iraharin reproducibilityofpulmonarybloodflowmeasurementsbyphasecontrastmriusingdifferent15tmrscannersattwoinstitutions
AT tsuchiyananae reproducibilityofpulmonarybloodflowmeasurementsbyphasecontrastmriusingdifferent15tmrscannersattwoinstitutions
AT yamashirotsuneo reproducibilityofpulmonarybloodflowmeasurementsbyphasecontrastmriusingdifferent15tmrscannersattwoinstitutions
AT iwasawatae reproducibilityofpulmonarybloodflowmeasurementsbyphasecontrastmriusingdifferent15tmrscannersattwoinstitutions
AT murayamasadayuki reproducibilityofpulmonarybloodflowmeasurementsbyphasecontrastmriusingdifferent15tmrscannersattwoinstitutions