Cargando…

Hybrid Intelligent System to Perform Fault Detection on BIS Sensor During Surgeries

This paper presents a new fault detection system in hypnotic sensors used for general anesthesia during surgery. Drug infusion during surgery is based on information received from patient monitoring devices; accordingly, faults in sensor devices can put patient safety at risk. Our research offers a...

Descripción completa

Detalles Bibliográficos
Autores principales: Casteleiro-Roca, José-Luis, Calvo-Rolle, José Luis, Méndez Pérez, Juan Albino, Roqueñí Gutiérrez, Nieves, de Cos Juez, Francisco Javier
Formato: Online Artículo Texto
Lenguaje:English
Publicado: MDPI 2017
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5298752/
https://www.ncbi.nlm.nih.gov/pubmed/28106793
http://dx.doi.org/10.3390/s17010179
Descripción
Sumario:This paper presents a new fault detection system in hypnotic sensors used for general anesthesia during surgery. Drug infusion during surgery is based on information received from patient monitoring devices; accordingly, faults in sensor devices can put patient safety at risk. Our research offers a solution to cope with these undesirable scenarios. We focus on the anesthesia process using intravenous propofol as the hypnotic drug and employing a Bispectral Index (BIS(TM)) monitor to estimate the patient’s unconsciousness level. The method developed identifies BIS episodes affected by disturbances during surgery with null clinical value. Thus, the clinician—or the automatic controller—will not take those measures into account to calculate the drug dose. Our method compares the measured BIS signal with expected behavior predicted by the propofol dose provider and the electromyogram (EMG) signal. For the prediction of the BIS signal, a model based on a hybrid intelligent system architecture has been created. The model uses clustering combined with regression techniques. To validate its accuracy, a dataset taken during surgeries with general anesthesia was used. The proposed fault detection method for BIS sensor measures has also been verified using data from real cases. The obtained results prove the method’s effectiveness.