Long-term reliability of sweet-tip type screw-in leads

BACKGROUND: Active fixation leads have provided stable atrial and ventricular pacing; however, long-term follow-up data have not been satisfactory. The purpose of this study was to investigate the long-term reliability of active fixation leads and their electrical characteristic stability. METHODS:...

Descripción completa

Detalles Bibliográficos
Autores principales: Takano, Asuka, Sekita, Gaku, Watanabe, Minako, Mukaida, Hiroshi, Komatsu, Sayaka, Tabuchi, Haruna, Hayashi, Hidemori, Tokano, Takashi, Sumiyoshi, Masataka, Nakazato, Yuji, Daida, Hiroyuki
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Elsevier 2017
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5300837/
https://www.ncbi.nlm.nih.gov/pubmed/28217223
http://dx.doi.org/10.1016/j.joa.2016.04.008
_version_ 1782506253163954176
author Takano, Asuka
Sekita, Gaku
Watanabe, Minako
Mukaida, Hiroshi
Komatsu, Sayaka
Tabuchi, Haruna
Hayashi, Hidemori
Tokano, Takashi
Sumiyoshi, Masataka
Nakazato, Yuji
Daida, Hiroyuki
author_facet Takano, Asuka
Sekita, Gaku
Watanabe, Minako
Mukaida, Hiroshi
Komatsu, Sayaka
Tabuchi, Haruna
Hayashi, Hidemori
Tokano, Takashi
Sumiyoshi, Masataka
Nakazato, Yuji
Daida, Hiroyuki
author_sort Takano, Asuka
collection PubMed
description BACKGROUND: Active fixation leads have provided stable atrial and ventricular pacing; however, long-term follow-up data have not been satisfactory. The purpose of this study was to investigate the long-term reliability of active fixation leads and their electrical characteristic stability. METHODS: A total of 1196 pacing leads were implanted in 830 patients consecutively between 2002 and 2013. In this retrospective study, we were able to trace 1092 leads in 750 patients to investigate the prognosis of implanted leads. The measurement values (including pacing thresholds, sensing amplitudes, and lead impedances of both the atrial and ventricular leads) were obtained from medical records at the time of implantation and during follow up at the outpatient device clinic. All pacing leads were FINELINE II Sterox EZ Leads (Boston Scientific, MN, USA), which are sweet-tip type screw-in active fixation leads, except for the shock leads in patients with implantable cardioverter defibrillator. RESULTS: The mean follow-up period was 51.3±29.2 months (median, 48 months). A total of 1092 leads were implanted in either the atrium (682 leads) or the ventricle (410 leads). Venous access was achieved through cephalic vein cut down (CVC) method (914 leads) or the subclavian vein puncture (SVP) method (178 leads). The overall lead survival rate was 99.6% at both 5 and 10 years. Lead fracture was observed in 4 of 1092 leads (0.37%), all of which were implanted by the SVP method. No lead fracture occurred among patients wherein CVC method was applied (p<0.01). Device-related infection was observed in four patients (0.53%). CONCLUSIONS: The overall reliability and stability of sweet-tip type screw-in leads were satisfactory throughout the long-term follow-up period (median, 4 years). Because it was associated with less lead fractures, cut-down access from the cephalic vein may be recommended as the first-line approach when considering the importance of long-term durability of pacing leads.
format Online
Article
Text
id pubmed-5300837
institution National Center for Biotechnology Information
language English
publishDate 2017
publisher Elsevier
record_format MEDLINE/PubMed
spelling pubmed-53008372017-02-17 Long-term reliability of sweet-tip type screw-in leads Takano, Asuka Sekita, Gaku Watanabe, Minako Mukaida, Hiroshi Komatsu, Sayaka Tabuchi, Haruna Hayashi, Hidemori Tokano, Takashi Sumiyoshi, Masataka Nakazato, Yuji Daida, Hiroyuki J Arrhythm Original Article BACKGROUND: Active fixation leads have provided stable atrial and ventricular pacing; however, long-term follow-up data have not been satisfactory. The purpose of this study was to investigate the long-term reliability of active fixation leads and their electrical characteristic stability. METHODS: A total of 1196 pacing leads were implanted in 830 patients consecutively between 2002 and 2013. In this retrospective study, we were able to trace 1092 leads in 750 patients to investigate the prognosis of implanted leads. The measurement values (including pacing thresholds, sensing amplitudes, and lead impedances of both the atrial and ventricular leads) were obtained from medical records at the time of implantation and during follow up at the outpatient device clinic. All pacing leads were FINELINE II Sterox EZ Leads (Boston Scientific, MN, USA), which are sweet-tip type screw-in active fixation leads, except for the shock leads in patients with implantable cardioverter defibrillator. RESULTS: The mean follow-up period was 51.3±29.2 months (median, 48 months). A total of 1092 leads were implanted in either the atrium (682 leads) or the ventricle (410 leads). Venous access was achieved through cephalic vein cut down (CVC) method (914 leads) or the subclavian vein puncture (SVP) method (178 leads). The overall lead survival rate was 99.6% at both 5 and 10 years. Lead fracture was observed in 4 of 1092 leads (0.37%), all of which were implanted by the SVP method. No lead fracture occurred among patients wherein CVC method was applied (p<0.01). Device-related infection was observed in four patients (0.53%). CONCLUSIONS: The overall reliability and stability of sweet-tip type screw-in leads were satisfactory throughout the long-term follow-up period (median, 4 years). Because it was associated with less lead fractures, cut-down access from the cephalic vein may be recommended as the first-line approach when considering the importance of long-term durability of pacing leads. Elsevier 2017-02 2016-06-22 /pmc/articles/PMC5300837/ /pubmed/28217223 http://dx.doi.org/10.1016/j.joa.2016.04.008 Text en © 2016 The Authors http://creativecommons.org/licenses/by-nc-nd/4.0/ This is an open access article under the CC BY-NC-ND license (http://creativecommons.org/licenses/by-nc-nd/4.0/).
spellingShingle Original Article
Takano, Asuka
Sekita, Gaku
Watanabe, Minako
Mukaida, Hiroshi
Komatsu, Sayaka
Tabuchi, Haruna
Hayashi, Hidemori
Tokano, Takashi
Sumiyoshi, Masataka
Nakazato, Yuji
Daida, Hiroyuki
Long-term reliability of sweet-tip type screw-in leads
title Long-term reliability of sweet-tip type screw-in leads
title_full Long-term reliability of sweet-tip type screw-in leads
title_fullStr Long-term reliability of sweet-tip type screw-in leads
title_full_unstemmed Long-term reliability of sweet-tip type screw-in leads
title_short Long-term reliability of sweet-tip type screw-in leads
title_sort long-term reliability of sweet-tip type screw-in leads
topic Original Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5300837/
https://www.ncbi.nlm.nih.gov/pubmed/28217223
http://dx.doi.org/10.1016/j.joa.2016.04.008
work_keys_str_mv AT takanoasuka longtermreliabilityofsweettiptypescrewinleads
AT sekitagaku longtermreliabilityofsweettiptypescrewinleads
AT watanabeminako longtermreliabilityofsweettiptypescrewinleads
AT mukaidahiroshi longtermreliabilityofsweettiptypescrewinleads
AT komatsusayaka longtermreliabilityofsweettiptypescrewinleads
AT tabuchiharuna longtermreliabilityofsweettiptypescrewinleads
AT hayashihidemori longtermreliabilityofsweettiptypescrewinleads
AT tokanotakashi longtermreliabilityofsweettiptypescrewinleads
AT sumiyoshimasataka longtermreliabilityofsweettiptypescrewinleads
AT nakazatoyuji longtermreliabilityofsweettiptypescrewinleads
AT daidahiroyuki longtermreliabilityofsweettiptypescrewinleads