Cargando…

Flash Glucose-Sensing Technology as a Replacement for Blood Glucose Monitoring for the Management of Insulin-Treated Type 2 Diabetes: a Multicenter, Open-Label Randomized Controlled Trial

INTRODUCTION: Glycemic control in participants with insulin-treated diabetes remains challenging. We assessed safety and efficacy of new flash glucose-sensing technology to replace self-monitoring of blood glucose (SMBG). METHODS: This open-label randomized controlled study (ClinicalTrials.gov, NCT0...

Descripción completa

Detalles Bibliográficos
Autores principales: Haak, Thomas, Hanaire, Hélène, Ajjan, Ramzi, Hermanns, Norbert, Riveline, Jean-Pierre, Rayman, Gerry
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Springer Healthcare 2016
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5306122/
https://www.ncbi.nlm.nih.gov/pubmed/28000140
http://dx.doi.org/10.1007/s13300-016-0223-6
_version_ 1782507133930045440
author Haak, Thomas
Hanaire, Hélène
Ajjan, Ramzi
Hermanns, Norbert
Riveline, Jean-Pierre
Rayman, Gerry
author_facet Haak, Thomas
Hanaire, Hélène
Ajjan, Ramzi
Hermanns, Norbert
Riveline, Jean-Pierre
Rayman, Gerry
author_sort Haak, Thomas
collection PubMed
description INTRODUCTION: Glycemic control in participants with insulin-treated diabetes remains challenging. We assessed safety and efficacy of new flash glucose-sensing technology to replace self-monitoring of blood glucose (SMBG). METHODS: This open-label randomized controlled study (ClinicalTrials.gov, NCT02082184) enrolled adults with type 2 diabetes on intensive insulin therapy from 26 European diabetes centers. Following 2 weeks of blinded sensor wear, 2:1 (intervention/control) randomization (centrally, using biased-coin minimization dependant on study center and insulin administration) was to control (SMBG) or intervention (glucose-sensing technology). Participants and investigators were not masked to group allocation. Primary outcome was difference in HbA1c at 6 months in the full analysis set. Prespecified secondary outcomes included time in hypoglycemia, effect of age, and patient satisfaction. RESULTS: Participants (n = 224) were randomized (149 intervention, 75 controls). At 6 months, there was no difference in the change in HbA1c between intervention and controls: −3.1 ± 0.75 mmol/mol, [−0.29 ± 0.07% (mean ± SE)] and −3.4 ± 1.04 mmol/mol (−0.31 ± 0.09%) respectively; p = 0.8222. A difference was detected in participants aged <65 years [−5.7 ± 0.96 mmol/mol (−0.53 ± 0.09%) and −2.2 ± 1.31 mmol/mol (−0.20 ± 0.12%), respectively; p = 0.0301]. Time in hypoglycemia <3.9 mmol/L (70 mg/dL) reduced by 0.47 ± 0.13 h/day [mean ± SE (p = 0.0006)], and <3.1 mmol/L (55 mg/dL) reduced by 0.22 ± 0.07 h/day (p = 0.0014) for intervention participants compared with controls; reductions of 43% and 53%, respectively. SMBG frequency, similar at baseline, decreased in intervention participants from 3.8 ± 1.4 tests/day (mean ± SD) to 0.3 ± 0.7, remaining unchanged in controls. Treatment satisfaction was higher in intervention compared with controls (DTSQ 13.1 ± 0.50 (mean ± SE) and 9.0 ± 0.72, respectively; p < 0.0001). No serious adverse events or severe hypoglycemic events were reported related to sensor data use. Forty-two serious events [16 (10.7%) intervention participants, 12 (16.0%) controls] were not device-related. Six intervention participants reported nine adverse events for sensor-wear reactions (two severe, six moderate, one mild). CONCLUSION: Flash glucose-sensing technology use in type 2 diabetes with intensive insulin therapy results in no difference in HbA1c change and reduced hypoglycemia, thus offering a safe, effective replacement for SMBG. TRIAL REGISTRATION: ClinicalTrials.gov identifier: NCT02082184. FUNDING: Abbott Diabetes Care. ELECTRONIC SUPPLEMENTARY MATERIAL: The online version of this article (doi:10.1007/s13300-016-0223-6) contains supplementary material, which is available to authorized users.
format Online
Article
Text
id pubmed-5306122
institution National Center for Biotechnology Information
language English
publishDate 2016
publisher Springer Healthcare
record_format MEDLINE/PubMed
spelling pubmed-53061222017-02-27 Flash Glucose-Sensing Technology as a Replacement for Blood Glucose Monitoring for the Management of Insulin-Treated Type 2 Diabetes: a Multicenter, Open-Label Randomized Controlled Trial Haak, Thomas Hanaire, Hélène Ajjan, Ramzi Hermanns, Norbert Riveline, Jean-Pierre Rayman, Gerry Diabetes Ther Original Research INTRODUCTION: Glycemic control in participants with insulin-treated diabetes remains challenging. We assessed safety and efficacy of new flash glucose-sensing technology to replace self-monitoring of blood glucose (SMBG). METHODS: This open-label randomized controlled study (ClinicalTrials.gov, NCT02082184) enrolled adults with type 2 diabetes on intensive insulin therapy from 26 European diabetes centers. Following 2 weeks of blinded sensor wear, 2:1 (intervention/control) randomization (centrally, using biased-coin minimization dependant on study center and insulin administration) was to control (SMBG) or intervention (glucose-sensing technology). Participants and investigators were not masked to group allocation. Primary outcome was difference in HbA1c at 6 months in the full analysis set. Prespecified secondary outcomes included time in hypoglycemia, effect of age, and patient satisfaction. RESULTS: Participants (n = 224) were randomized (149 intervention, 75 controls). At 6 months, there was no difference in the change in HbA1c between intervention and controls: −3.1 ± 0.75 mmol/mol, [−0.29 ± 0.07% (mean ± SE)] and −3.4 ± 1.04 mmol/mol (−0.31 ± 0.09%) respectively; p = 0.8222. A difference was detected in participants aged <65 years [−5.7 ± 0.96 mmol/mol (−0.53 ± 0.09%) and −2.2 ± 1.31 mmol/mol (−0.20 ± 0.12%), respectively; p = 0.0301]. Time in hypoglycemia <3.9 mmol/L (70 mg/dL) reduced by 0.47 ± 0.13 h/day [mean ± SE (p = 0.0006)], and <3.1 mmol/L (55 mg/dL) reduced by 0.22 ± 0.07 h/day (p = 0.0014) for intervention participants compared with controls; reductions of 43% and 53%, respectively. SMBG frequency, similar at baseline, decreased in intervention participants from 3.8 ± 1.4 tests/day (mean ± SD) to 0.3 ± 0.7, remaining unchanged in controls. Treatment satisfaction was higher in intervention compared with controls (DTSQ 13.1 ± 0.50 (mean ± SE) and 9.0 ± 0.72, respectively; p < 0.0001). No serious adverse events or severe hypoglycemic events were reported related to sensor data use. Forty-two serious events [16 (10.7%) intervention participants, 12 (16.0%) controls] were not device-related. Six intervention participants reported nine adverse events for sensor-wear reactions (two severe, six moderate, one mild). CONCLUSION: Flash glucose-sensing technology use in type 2 diabetes with intensive insulin therapy results in no difference in HbA1c change and reduced hypoglycemia, thus offering a safe, effective replacement for SMBG. TRIAL REGISTRATION: ClinicalTrials.gov identifier: NCT02082184. FUNDING: Abbott Diabetes Care. ELECTRONIC SUPPLEMENTARY MATERIAL: The online version of this article (doi:10.1007/s13300-016-0223-6) contains supplementary material, which is available to authorized users. Springer Healthcare 2016-12-20 2017-02 /pmc/articles/PMC5306122/ /pubmed/28000140 http://dx.doi.org/10.1007/s13300-016-0223-6 Text en © The Author(s) 2016 https://creativecommons.org/licenses/by-nc/4.0/This article is distributed under the terms of the Creative Commons Attribution-NonCommercial 4.0 International License (http://creativecommons.org/licenses/by-nc/4.0/ (https://creativecommons.org/licenses/by-nc/4.0/) ), which permits any noncommercial use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made.
spellingShingle Original Research
Haak, Thomas
Hanaire, Hélène
Ajjan, Ramzi
Hermanns, Norbert
Riveline, Jean-Pierre
Rayman, Gerry
Flash Glucose-Sensing Technology as a Replacement for Blood Glucose Monitoring for the Management of Insulin-Treated Type 2 Diabetes: a Multicenter, Open-Label Randomized Controlled Trial
title Flash Glucose-Sensing Technology as a Replacement for Blood Glucose Monitoring for the Management of Insulin-Treated Type 2 Diabetes: a Multicenter, Open-Label Randomized Controlled Trial
title_full Flash Glucose-Sensing Technology as a Replacement for Blood Glucose Monitoring for the Management of Insulin-Treated Type 2 Diabetes: a Multicenter, Open-Label Randomized Controlled Trial
title_fullStr Flash Glucose-Sensing Technology as a Replacement for Blood Glucose Monitoring for the Management of Insulin-Treated Type 2 Diabetes: a Multicenter, Open-Label Randomized Controlled Trial
title_full_unstemmed Flash Glucose-Sensing Technology as a Replacement for Blood Glucose Monitoring for the Management of Insulin-Treated Type 2 Diabetes: a Multicenter, Open-Label Randomized Controlled Trial
title_short Flash Glucose-Sensing Technology as a Replacement for Blood Glucose Monitoring for the Management of Insulin-Treated Type 2 Diabetes: a Multicenter, Open-Label Randomized Controlled Trial
title_sort flash glucose-sensing technology as a replacement for blood glucose monitoring for the management of insulin-treated type 2 diabetes: a multicenter, open-label randomized controlled trial
topic Original Research
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5306122/
https://www.ncbi.nlm.nih.gov/pubmed/28000140
http://dx.doi.org/10.1007/s13300-016-0223-6
work_keys_str_mv AT haakthomas flashglucosesensingtechnologyasareplacementforbloodglucosemonitoringforthemanagementofinsulintreatedtype2diabetesamulticenteropenlabelrandomizedcontrolledtrial
AT hanairehelene flashglucosesensingtechnologyasareplacementforbloodglucosemonitoringforthemanagementofinsulintreatedtype2diabetesamulticenteropenlabelrandomizedcontrolledtrial
AT ajjanramzi flashglucosesensingtechnologyasareplacementforbloodglucosemonitoringforthemanagementofinsulintreatedtype2diabetesamulticenteropenlabelrandomizedcontrolledtrial
AT hermannsnorbert flashglucosesensingtechnologyasareplacementforbloodglucosemonitoringforthemanagementofinsulintreatedtype2diabetesamulticenteropenlabelrandomizedcontrolledtrial
AT rivelinejeanpierre flashglucosesensingtechnologyasareplacementforbloodglucosemonitoringforthemanagementofinsulintreatedtype2diabetesamulticenteropenlabelrandomizedcontrolledtrial
AT raymangerry flashglucosesensingtechnologyasareplacementforbloodglucosemonitoringforthemanagementofinsulintreatedtype2diabetesamulticenteropenlabelrandomizedcontrolledtrial