The development and characterization of a novel yet simple 3D printed tool to facilitate phantom imaging of photoacoustic contrast agents
We report a new approach to preparing phantoms using 3D printing. This device supports plastic tubing containing the contrast agent and is immersed in a solution with absorption or scattering properties that mimic tissue. Up to 12 tubing samples could be placed in the device with sample-to-sample sp...
Autores principales: | , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Elsevier
2017
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5314822/ https://www.ncbi.nlm.nih.gov/pubmed/28239554 http://dx.doi.org/10.1016/j.pacs.2017.02.001 |
Sumario: | We report a new approach to preparing phantoms using 3D printing. This device supports plastic tubing containing the contrast agent and is immersed in a solution with absorption or scattering properties that mimic tissue. Up to 12 tubing samples could be placed in the device with sample-to-sample spacing as low as 0.3 mm and at a constant distance from the transducer (±0.16 mm), which is critical in validating photoacoustic contrast agents. We also studied different types of tubing and found that tubing with a larger outside diameter has more inherent signal. Both 40% India Ink and lipids in the immersion media modulated the signal. Finally, we created a depth phantom and found that signal decayed following a linear relationship (R(2) = 0.997) with respect to distance from the focal point. We include computer-assisted drafting code the community can use to print this phantom or customized versions of this phantom. |
---|