Perioperative onset of acquired von Willebrand syndrome: Comparison between HVAD, HeartMate II and on-pump coronary bypass surgery

OBJECTIVES: Acquired von Willebrand syndrome (AvWS) is associated with postoperative bleeding complications in patients with continuous flow left ventricular assist devices (CF-LVADs). The aim of this study is to analyze the perioperative vWF profile comparing an axial pump (HMII) to a centrifugal p...

Descripción completa

Detalles Bibliográficos
Autores principales: Feldmann, Christina, Zayat, Rashad, Goetzenich, Andreas, Aljalloud, Ali, Woelke, Eva, Maas, Judith, Tewarie, Lachmandath, Schmitz-Rode, Thomas, Autschbach, Ruediger, Steinseifer, Ulrich, Moza, Ajay
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Public Library of Science 2017
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5325196/
https://www.ncbi.nlm.nih.gov/pubmed/28234916
http://dx.doi.org/10.1371/journal.pone.0171029
_version_ 1782510332868034560
author Feldmann, Christina
Zayat, Rashad
Goetzenich, Andreas
Aljalloud, Ali
Woelke, Eva
Maas, Judith
Tewarie, Lachmandath
Schmitz-Rode, Thomas
Autschbach, Ruediger
Steinseifer, Ulrich
Moza, Ajay
author_facet Feldmann, Christina
Zayat, Rashad
Goetzenich, Andreas
Aljalloud, Ali
Woelke, Eva
Maas, Judith
Tewarie, Lachmandath
Schmitz-Rode, Thomas
Autschbach, Ruediger
Steinseifer, Ulrich
Moza, Ajay
author_sort Feldmann, Christina
collection PubMed
description OBJECTIVES: Acquired von Willebrand syndrome (AvWS) is associated with postoperative bleeding complications in patients with continuous flow left ventricular assist devices (CF-LVADs). The aim of this study is to analyze the perioperative vWF profile comparing an axial pump (HMII) to a centrifugal pump (HVAD) regarding the correlation between perioperative occurrence of AvWS, early- and late-postoperative bleeding events. METHODS: From July 2013 until March 2015 blood samples of 33 patients (12 HMII/ 8 HVAD/ 13 controls) were prospectively collected at 12 different time points and analyzed for the vWF antigen (vWF:Ag), its activity (vWF:Ac) and the vWF:Ac/vWF:Ag-ratio (vWF:ratio). The follow up period for postoperative bleeding events was from July 2013 until July 2016. RESULTS: Postoperatively, there was no difference in the vWF-profile between HVAD and HMII groups. However, a subgroup of patients already had significantly lower vWF:ratios preoperatively. Postoperatively, both CF-LVAD groups presented significantly lower vWF:ratios compared to the control group. Bleeding events per patient-year did not differ between the two groups (HMII vs. HVAD: 0.67 vs. 0.85, p = 0.685). We detected a correlation between vWF:ratio <0.7at LVAD-start (r = -0.583, p = 0.006) or at the end of surgery (r = -0.461, p = 0.035) and the occurrence of pericardial tamponade. In the control group, the drop in both vWF:Ag and vWF:Ac recovered immediately postoperatively above preoperative values. CONCLUSION: A subgroup of patients with end-stage heart failure already suffers AvWS preoperatively. In both CF-LVAD groups, AvWS begins immediately after surgery. Intraoperative vWF:ratios <0.7 correlate with higher incidences of pericardial tamponade and re-operation. The presumably dilutive effect of the heart lung machine on vWF vanishes immediately at the end of surgery, possibly as part of an acute-phase response.
format Online
Article
Text
id pubmed-5325196
institution National Center for Biotechnology Information
language English
publishDate 2017
publisher Public Library of Science
record_format MEDLINE/PubMed
spelling pubmed-53251962017-03-09 Perioperative onset of acquired von Willebrand syndrome: Comparison between HVAD, HeartMate II and on-pump coronary bypass surgery Feldmann, Christina Zayat, Rashad Goetzenich, Andreas Aljalloud, Ali Woelke, Eva Maas, Judith Tewarie, Lachmandath Schmitz-Rode, Thomas Autschbach, Ruediger Steinseifer, Ulrich Moza, Ajay PLoS One Research Article OBJECTIVES: Acquired von Willebrand syndrome (AvWS) is associated with postoperative bleeding complications in patients with continuous flow left ventricular assist devices (CF-LVADs). The aim of this study is to analyze the perioperative vWF profile comparing an axial pump (HMII) to a centrifugal pump (HVAD) regarding the correlation between perioperative occurrence of AvWS, early- and late-postoperative bleeding events. METHODS: From July 2013 until March 2015 blood samples of 33 patients (12 HMII/ 8 HVAD/ 13 controls) were prospectively collected at 12 different time points and analyzed for the vWF antigen (vWF:Ag), its activity (vWF:Ac) and the vWF:Ac/vWF:Ag-ratio (vWF:ratio). The follow up period for postoperative bleeding events was from July 2013 until July 2016. RESULTS: Postoperatively, there was no difference in the vWF-profile between HVAD and HMII groups. However, a subgroup of patients already had significantly lower vWF:ratios preoperatively. Postoperatively, both CF-LVAD groups presented significantly lower vWF:ratios compared to the control group. Bleeding events per patient-year did not differ between the two groups (HMII vs. HVAD: 0.67 vs. 0.85, p = 0.685). We detected a correlation between vWF:ratio <0.7at LVAD-start (r = -0.583, p = 0.006) or at the end of surgery (r = -0.461, p = 0.035) and the occurrence of pericardial tamponade. In the control group, the drop in both vWF:Ag and vWF:Ac recovered immediately postoperatively above preoperative values. CONCLUSION: A subgroup of patients with end-stage heart failure already suffers AvWS preoperatively. In both CF-LVAD groups, AvWS begins immediately after surgery. Intraoperative vWF:ratios <0.7 correlate with higher incidences of pericardial tamponade and re-operation. The presumably dilutive effect of the heart lung machine on vWF vanishes immediately at the end of surgery, possibly as part of an acute-phase response. Public Library of Science 2017-02-24 /pmc/articles/PMC5325196/ /pubmed/28234916 http://dx.doi.org/10.1371/journal.pone.0171029 Text en © 2017 Feldmann et al http://creativecommons.org/licenses/by/4.0/ This is an open access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/4.0/) , which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.
spellingShingle Research Article
Feldmann, Christina
Zayat, Rashad
Goetzenich, Andreas
Aljalloud, Ali
Woelke, Eva
Maas, Judith
Tewarie, Lachmandath
Schmitz-Rode, Thomas
Autschbach, Ruediger
Steinseifer, Ulrich
Moza, Ajay
Perioperative onset of acquired von Willebrand syndrome: Comparison between HVAD, HeartMate II and on-pump coronary bypass surgery
title Perioperative onset of acquired von Willebrand syndrome: Comparison between HVAD, HeartMate II and on-pump coronary bypass surgery
title_full Perioperative onset of acquired von Willebrand syndrome: Comparison between HVAD, HeartMate II and on-pump coronary bypass surgery
title_fullStr Perioperative onset of acquired von Willebrand syndrome: Comparison between HVAD, HeartMate II and on-pump coronary bypass surgery
title_full_unstemmed Perioperative onset of acquired von Willebrand syndrome: Comparison between HVAD, HeartMate II and on-pump coronary bypass surgery
title_short Perioperative onset of acquired von Willebrand syndrome: Comparison between HVAD, HeartMate II and on-pump coronary bypass surgery
title_sort perioperative onset of acquired von willebrand syndrome: comparison between hvad, heartmate ii and on-pump coronary bypass surgery
topic Research Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5325196/
https://www.ncbi.nlm.nih.gov/pubmed/28234916
http://dx.doi.org/10.1371/journal.pone.0171029
work_keys_str_mv AT feldmannchristina perioperativeonsetofacquiredvonwillebrandsyndromecomparisonbetweenhvadheartmateiiandonpumpcoronarybypasssurgery
AT zayatrashad perioperativeonsetofacquiredvonwillebrandsyndromecomparisonbetweenhvadheartmateiiandonpumpcoronarybypasssurgery
AT goetzenichandreas perioperativeonsetofacquiredvonwillebrandsyndromecomparisonbetweenhvadheartmateiiandonpumpcoronarybypasssurgery
AT aljalloudali perioperativeonsetofacquiredvonwillebrandsyndromecomparisonbetweenhvadheartmateiiandonpumpcoronarybypasssurgery
AT woelkeeva perioperativeonsetofacquiredvonwillebrandsyndromecomparisonbetweenhvadheartmateiiandonpumpcoronarybypasssurgery
AT maasjudith perioperativeonsetofacquiredvonwillebrandsyndromecomparisonbetweenhvadheartmateiiandonpumpcoronarybypasssurgery
AT tewarielachmandath perioperativeonsetofacquiredvonwillebrandsyndromecomparisonbetweenhvadheartmateiiandonpumpcoronarybypasssurgery
AT schmitzrodethomas perioperativeonsetofacquiredvonwillebrandsyndromecomparisonbetweenhvadheartmateiiandonpumpcoronarybypasssurgery
AT autschbachruediger perioperativeonsetofacquiredvonwillebrandsyndromecomparisonbetweenhvadheartmateiiandonpumpcoronarybypasssurgery
AT steinseiferulrich perioperativeonsetofacquiredvonwillebrandsyndromecomparisonbetweenhvadheartmateiiandonpumpcoronarybypasssurgery
AT mozaajay perioperativeonsetofacquiredvonwillebrandsyndromecomparisonbetweenhvadheartmateiiandonpumpcoronarybypasssurgery