Critical analysis of ineffective post implantation implantable cardioverter-defibrillator-testing
AIM: To test of the implantable-cardioverter-defibrillator is done at the time of implantation. We investigate if any testing should be performed. METHODS: All consecutive patients between January 2006 and December 2008 undergoing implantable cardioverter-defibrillator (ICD) implantation/replacement...
Autores principales: | , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Baishideng Publishing Group Inc
2017
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5329744/ https://www.ncbi.nlm.nih.gov/pubmed/28289531 http://dx.doi.org/10.4330/wjc.v9.i2.167 |
_version_ | 1782511116659720192 |
---|---|
author | Roos, Markus Geller, J Christoph Ohlow, Marc-Alexander |
author_facet | Roos, Markus Geller, J Christoph Ohlow, Marc-Alexander |
author_sort | Roos, Markus |
collection | PubMed |
description | AIM: To test of the implantable-cardioverter-defibrillator is done at the time of implantation. We investigate if any testing should be performed. METHODS: All consecutive patients between January 2006 and December 2008 undergoing implantable cardioverter-defibrillator (ICD) implantation/replacement (a total of 634 patients) were included in the retrospective study. RESULTS: Sixteen patients (2.5%) were not tested (9 with LA/LV-thrombus, 7 due to operator’s decision). Analyzed were 618 patients [76% men, 66.4 + 11 years, 24% secondary prevention (SP), 46% with left ventricular ejection fraction (LVEF) < 20%, 56% had coronary artery disease (CAD)] undergoing defibrillation safety testing (SMT) with an energy of 21 + 2.3 J. In 22/618 patients (3.6%) induced ventricular fibrillation (VF) could not be terminated with maximum energy of the ICD. Six of those (27%) had successful SMT after system modification or shock lead repositioning, 14 patients (64%) received a subcutaneous electrode array. Younger age (P = 0.0003), non-CAD (P = 0.007) and VF as index event for SP (P = 0.05) were associated with a higher incidence of ineffective SMT. LVEF < 20% and incomplete revascularisation in patients with CAD had no impact on SMT. CONCLUSION: Defibrillation testing is well-tolerated. An ineffective SMT occurred in 4% and two third of those needed implantation of a subcutaneous electrode array to pass a SMT > 10 J. |
format | Online Article Text |
id | pubmed-5329744 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2017 |
publisher | Baishideng Publishing Group Inc |
record_format | MEDLINE/PubMed |
spelling | pubmed-53297442017-03-13 Critical analysis of ineffective post implantation implantable cardioverter-defibrillator-testing Roos, Markus Geller, J Christoph Ohlow, Marc-Alexander World J Cardiol Observational Study AIM: To test of the implantable-cardioverter-defibrillator is done at the time of implantation. We investigate if any testing should be performed. METHODS: All consecutive patients between January 2006 and December 2008 undergoing implantable cardioverter-defibrillator (ICD) implantation/replacement (a total of 634 patients) were included in the retrospective study. RESULTS: Sixteen patients (2.5%) were not tested (9 with LA/LV-thrombus, 7 due to operator’s decision). Analyzed were 618 patients [76% men, 66.4 + 11 years, 24% secondary prevention (SP), 46% with left ventricular ejection fraction (LVEF) < 20%, 56% had coronary artery disease (CAD)] undergoing defibrillation safety testing (SMT) with an energy of 21 + 2.3 J. In 22/618 patients (3.6%) induced ventricular fibrillation (VF) could not be terminated with maximum energy of the ICD. Six of those (27%) had successful SMT after system modification or shock lead repositioning, 14 patients (64%) received a subcutaneous electrode array. Younger age (P = 0.0003), non-CAD (P = 0.007) and VF as index event for SP (P = 0.05) were associated with a higher incidence of ineffective SMT. LVEF < 20% and incomplete revascularisation in patients with CAD had no impact on SMT. CONCLUSION: Defibrillation testing is well-tolerated. An ineffective SMT occurred in 4% and two third of those needed implantation of a subcutaneous electrode array to pass a SMT > 10 J. Baishideng Publishing Group Inc 2017-02-26 2017-02-26 /pmc/articles/PMC5329744/ /pubmed/28289531 http://dx.doi.org/10.4330/wjc.v9.i2.167 Text en ©The Author(s) 2017. Published by Baishideng Publishing Group Inc. All rights reserved. http://creativecommons.org/licenses/by-nc/4.0/ Open-Access: This article is an open-access article which was selected by an in-house editor and fully peer-reviewed by external reviewers. It is distributed in accordance with the Creative Commons Attribution Non Commercial (CC BY-NC 4.0) license, which permits others to distribute, remix, adapt, build upon this work non-commercially, and license their derivative works on different terms, provided the original work is properly cited and the use is non-commercial. See: http://creativecommons.org/licenses/by-nc/4.0/ |
spellingShingle | Observational Study Roos, Markus Geller, J Christoph Ohlow, Marc-Alexander Critical analysis of ineffective post implantation implantable cardioverter-defibrillator-testing |
title | Critical analysis of ineffective post implantation implantable cardioverter-defibrillator-testing |
title_full | Critical analysis of ineffective post implantation implantable cardioverter-defibrillator-testing |
title_fullStr | Critical analysis of ineffective post implantation implantable cardioverter-defibrillator-testing |
title_full_unstemmed | Critical analysis of ineffective post implantation implantable cardioverter-defibrillator-testing |
title_short | Critical analysis of ineffective post implantation implantable cardioverter-defibrillator-testing |
title_sort | critical analysis of ineffective post implantation implantable cardioverter-defibrillator-testing |
topic | Observational Study |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5329744/ https://www.ncbi.nlm.nih.gov/pubmed/28289531 http://dx.doi.org/10.4330/wjc.v9.i2.167 |
work_keys_str_mv | AT roosmarkus criticalanalysisofineffectivepostimplantationimplantablecardioverterdefibrillatortesting AT gellerjchristoph criticalanalysisofineffectivepostimplantationimplantablecardioverterdefibrillatortesting AT ohlowmarcalexander criticalanalysisofineffectivepostimplantationimplantablecardioverterdefibrillatortesting |