Critical analysis of ineffective post implantation implantable cardioverter-defibrillator-testing

AIM: To test of the implantable-cardioverter-defibrillator is done at the time of implantation. We investigate if any testing should be performed. METHODS: All consecutive patients between January 2006 and December 2008 undergoing implantable cardioverter-defibrillator (ICD) implantation/replacement...

Descripción completa

Detalles Bibliográficos
Autores principales: Roos, Markus, Geller, J Christoph, Ohlow, Marc-Alexander
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Baishideng Publishing Group Inc 2017
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5329744/
https://www.ncbi.nlm.nih.gov/pubmed/28289531
http://dx.doi.org/10.4330/wjc.v9.i2.167
_version_ 1782511116659720192
author Roos, Markus
Geller, J Christoph
Ohlow, Marc-Alexander
author_facet Roos, Markus
Geller, J Christoph
Ohlow, Marc-Alexander
author_sort Roos, Markus
collection PubMed
description AIM: To test of the implantable-cardioverter-defibrillator is done at the time of implantation. We investigate if any testing should be performed. METHODS: All consecutive patients between January 2006 and December 2008 undergoing implantable cardioverter-defibrillator (ICD) implantation/replacement (a total of 634 patients) were included in the retrospective study. RESULTS: Sixteen patients (2.5%) were not tested (9 with LA/LV-thrombus, 7 due to operator’s decision). Analyzed were 618 patients [76% men, 66.4 + 11 years, 24% secondary prevention (SP), 46% with left ventricular ejection fraction (LVEF) < 20%, 56% had coronary artery disease (CAD)] undergoing defibrillation safety testing (SMT) with an energy of 21 + 2.3 J. In 22/618 patients (3.6%) induced ventricular fibrillation (VF) could not be terminated with maximum energy of the ICD. Six of those (27%) had successful SMT after system modification or shock lead repositioning, 14 patients (64%) received a subcutaneous electrode array. Younger age (P = 0.0003), non-CAD (P = 0.007) and VF as index event for SP (P = 0.05) were associated with a higher incidence of ineffective SMT. LVEF < 20% and incomplete revascularisation in patients with CAD had no impact on SMT. CONCLUSION: Defibrillation testing is well-tolerated. An ineffective SMT occurred in 4% and two third of those needed implantation of a subcutaneous electrode array to pass a SMT > 10 J.
format Online
Article
Text
id pubmed-5329744
institution National Center for Biotechnology Information
language English
publishDate 2017
publisher Baishideng Publishing Group Inc
record_format MEDLINE/PubMed
spelling pubmed-53297442017-03-13 Critical analysis of ineffective post implantation implantable cardioverter-defibrillator-testing Roos, Markus Geller, J Christoph Ohlow, Marc-Alexander World J Cardiol Observational Study AIM: To test of the implantable-cardioverter-defibrillator is done at the time of implantation. We investigate if any testing should be performed. METHODS: All consecutive patients between January 2006 and December 2008 undergoing implantable cardioverter-defibrillator (ICD) implantation/replacement (a total of 634 patients) were included in the retrospective study. RESULTS: Sixteen patients (2.5%) were not tested (9 with LA/LV-thrombus, 7 due to operator’s decision). Analyzed were 618 patients [76% men, 66.4 + 11 years, 24% secondary prevention (SP), 46% with left ventricular ejection fraction (LVEF) < 20%, 56% had coronary artery disease (CAD)] undergoing defibrillation safety testing (SMT) with an energy of 21 + 2.3 J. In 22/618 patients (3.6%) induced ventricular fibrillation (VF) could not be terminated with maximum energy of the ICD. Six of those (27%) had successful SMT after system modification or shock lead repositioning, 14 patients (64%) received a subcutaneous electrode array. Younger age (P = 0.0003), non-CAD (P = 0.007) and VF as index event for SP (P = 0.05) were associated with a higher incidence of ineffective SMT. LVEF < 20% and incomplete revascularisation in patients with CAD had no impact on SMT. CONCLUSION: Defibrillation testing is well-tolerated. An ineffective SMT occurred in 4% and two third of those needed implantation of a subcutaneous electrode array to pass a SMT > 10 J. Baishideng Publishing Group Inc 2017-02-26 2017-02-26 /pmc/articles/PMC5329744/ /pubmed/28289531 http://dx.doi.org/10.4330/wjc.v9.i2.167 Text en ©The Author(s) 2017. Published by Baishideng Publishing Group Inc. All rights reserved. http://creativecommons.org/licenses/by-nc/4.0/ Open-Access: This article is an open-access article which was selected by an in-house editor and fully peer-reviewed by external reviewers. It is distributed in accordance with the Creative Commons Attribution Non Commercial (CC BY-NC 4.0) license, which permits others to distribute, remix, adapt, build upon this work non-commercially, and license their derivative works on different terms, provided the original work is properly cited and the use is non-commercial. See: http://creativecommons.org/licenses/by-nc/4.0/
spellingShingle Observational Study
Roos, Markus
Geller, J Christoph
Ohlow, Marc-Alexander
Critical analysis of ineffective post implantation implantable cardioverter-defibrillator-testing
title Critical analysis of ineffective post implantation implantable cardioverter-defibrillator-testing
title_full Critical analysis of ineffective post implantation implantable cardioverter-defibrillator-testing
title_fullStr Critical analysis of ineffective post implantation implantable cardioverter-defibrillator-testing
title_full_unstemmed Critical analysis of ineffective post implantation implantable cardioverter-defibrillator-testing
title_short Critical analysis of ineffective post implantation implantable cardioverter-defibrillator-testing
title_sort critical analysis of ineffective post implantation implantable cardioverter-defibrillator-testing
topic Observational Study
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5329744/
https://www.ncbi.nlm.nih.gov/pubmed/28289531
http://dx.doi.org/10.4330/wjc.v9.i2.167
work_keys_str_mv AT roosmarkus criticalanalysisofineffectivepostimplantationimplantablecardioverterdefibrillatortesting
AT gellerjchristoph criticalanalysisofineffectivepostimplantationimplantablecardioverterdefibrillatortesting
AT ohlowmarcalexander criticalanalysisofineffectivepostimplantationimplantablecardioverterdefibrillatortesting