Cargando…
Quasi-Static Calibration Method of a High-g Accelerometer
To solve the problem of resonance during quasi-static calibration of high-g accelerometers, we deduce the relationship between the minimum excitation pulse width and the resonant frequency of the calibrated accelerometer according to the second-order mathematical model of the accelerometer, and impr...
Autores principales: | , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
MDPI
2017
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5336105/ https://www.ncbi.nlm.nih.gov/pubmed/28230743 http://dx.doi.org/10.3390/s17020409 |
_version_ | 1782512162869084160 |
---|---|
author | Wang, Yan Fan, Jinbiao Zu, Jing Xu, Peng |
author_facet | Wang, Yan Fan, Jinbiao Zu, Jing Xu, Peng |
author_sort | Wang, Yan |
collection | PubMed |
description | To solve the problem of resonance during quasi-static calibration of high-g accelerometers, we deduce the relationship between the minimum excitation pulse width and the resonant frequency of the calibrated accelerometer according to the second-order mathematical model of the accelerometer, and improve the quasi-static calibration theory. We establish a quasi-static calibration testing system, which uses a gas gun to generate high-g acceleration signals, and apply a laser interferometer to reproduce the impact acceleration. These signals are used to drive the calibrated accelerometer. By comparing the excitation acceleration signal and the output responses of the calibrated accelerometer to the excitation signals, the impact sensitivity of the calibrated accelerometer is obtained. As indicated by the calibration test results, this calibration system produces excitation acceleration signals with a pulse width of less than 1000 μs, and realize the quasi-static calibration of high-g accelerometers with a resonant frequency above 20 kHz when the calibration error was 3%. |
format | Online Article Text |
id | pubmed-5336105 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2017 |
publisher | MDPI |
record_format | MEDLINE/PubMed |
spelling | pubmed-53361052017-03-16 Quasi-Static Calibration Method of a High-g Accelerometer Wang, Yan Fan, Jinbiao Zu, Jing Xu, Peng Sensors (Basel) Article To solve the problem of resonance during quasi-static calibration of high-g accelerometers, we deduce the relationship between the minimum excitation pulse width and the resonant frequency of the calibrated accelerometer according to the second-order mathematical model of the accelerometer, and improve the quasi-static calibration theory. We establish a quasi-static calibration testing system, which uses a gas gun to generate high-g acceleration signals, and apply a laser interferometer to reproduce the impact acceleration. These signals are used to drive the calibrated accelerometer. By comparing the excitation acceleration signal and the output responses of the calibrated accelerometer to the excitation signals, the impact sensitivity of the calibrated accelerometer is obtained. As indicated by the calibration test results, this calibration system produces excitation acceleration signals with a pulse width of less than 1000 μs, and realize the quasi-static calibration of high-g accelerometers with a resonant frequency above 20 kHz when the calibration error was 3%. MDPI 2017-02-20 /pmc/articles/PMC5336105/ /pubmed/28230743 http://dx.doi.org/10.3390/s17020409 Text en © 2017 by the authors. Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (http://creativecommons.org/licenses/by/4.0/). |
spellingShingle | Article Wang, Yan Fan, Jinbiao Zu, Jing Xu, Peng Quasi-Static Calibration Method of a High-g Accelerometer |
title | Quasi-Static Calibration Method of a High-g Accelerometer |
title_full | Quasi-Static Calibration Method of a High-g Accelerometer |
title_fullStr | Quasi-Static Calibration Method of a High-g Accelerometer |
title_full_unstemmed | Quasi-Static Calibration Method of a High-g Accelerometer |
title_short | Quasi-Static Calibration Method of a High-g Accelerometer |
title_sort | quasi-static calibration method of a high-g accelerometer |
topic | Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5336105/ https://www.ncbi.nlm.nih.gov/pubmed/28230743 http://dx.doi.org/10.3390/s17020409 |
work_keys_str_mv | AT wangyan quasistaticcalibrationmethodofahighgaccelerometer AT fanjinbiao quasistaticcalibrationmethodofahighgaccelerometer AT zujing quasistaticcalibrationmethodofahighgaccelerometer AT xupeng quasistaticcalibrationmethodofahighgaccelerometer |