Cargando…

Predictors of clinical outcome in transfemoral TAVI: Circumferential iliofemoral calcifications and manufacturer-derived recommendations

OBJECTIVE: This study aimed to investigate the predictive value of circumferential iliofemoral calcifications and current manufacturer recommendations, which are not evidence-based, in transfemoral (TF) transcatheter aortic valve implantation (TAVI) METHODS: A patient cohort with a broad range of il...

Descripción completa

Detalles Bibliográficos
Autores principales: Reinthaler, Markus, Aggarwal, Sunil K., Palma, Rodney De, Landmesser, Ulf, Froehlich, Georg, Yap, John, Meier, Pascal, J. Mullen, Michael
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Kare Publishing 2015
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5336838/
https://www.ncbi.nlm.nih.gov/pubmed/25413227
http://dx.doi.org/10.5152/akd.2014.5311
_version_ 1782512272857366528
author Reinthaler, Markus
Aggarwal, Sunil K.
Palma, Rodney De
Landmesser, Ulf
Froehlich, Georg
Yap, John
Meier, Pascal
J. Mullen, Michael
author_facet Reinthaler, Markus
Aggarwal, Sunil K.
Palma, Rodney De
Landmesser, Ulf
Froehlich, Georg
Yap, John
Meier, Pascal
J. Mullen, Michael
author_sort Reinthaler, Markus
collection PubMed
description OBJECTIVE: This study aimed to investigate the predictive value of circumferential iliofemoral calcifications and current manufacturer recommendations, which are not evidence-based, in transfemoral (TF) transcatheter aortic valve implantation (TAVI) METHODS: A patient cohort with a broad range of iliofemoral anatomies undergoing TF TAVI (n=132) were retrospectively divided as “suitable” (n=76, 58%) and “unsuitable” (n=56, 42%) candidates according to current recommendations. Iliofemoral angiography and reconstructed mul- tislice CT (MSCT) images were used for access screening in the majority of patients. RESULTS: Vessel properties were significantly worse in the “unsuitable group.” The sheath-to-iliofemoral artery ratio (SIFAR) and calcium score were 1.35±0.2 and 1.7±0.8 in the unsuitable group, compared to 1.0±0.12 (p<0.0001) and 1.0±0.7 (p=0.0001) in the “suitable” patients. Major vascular complications (MVCs) occurred more frequently in the “unsuitable” group (10.7% vs. 2.6%, p=0.07) and were predicted by SIFAR [OR: 64, 95% CI: 1.4-2971, p=0.03] and circumferential iliofemoral calcifications [OR: 6, 95% CI: 1.2-26, p=0.025]. In the multivariate analysis, circumferential calcifications [HR: 3.6, 95% CI: 1-13.2, p=0.043] but not major vascular complications (MVCs) or manufacturer recommendations were associated with increased mortality. CONCLUSION: According to our results, manufacturer recommendations are safe but overly conservative. Circumferential iliofemoral calcifications may provide independent prognostic information in patients undergoing TAVI.
format Online
Article
Text
id pubmed-5336838
institution National Center for Biotechnology Information
language English
publishDate 2015
publisher Kare Publishing
record_format MEDLINE/PubMed
spelling pubmed-53368382017-06-28 Predictors of clinical outcome in transfemoral TAVI: Circumferential iliofemoral calcifications and manufacturer-derived recommendations Reinthaler, Markus Aggarwal, Sunil K. Palma, Rodney De Landmesser, Ulf Froehlich, Georg Yap, John Meier, Pascal J. Mullen, Michael Anatol J Cardiol Original Investigation OBJECTIVE: This study aimed to investigate the predictive value of circumferential iliofemoral calcifications and current manufacturer recommendations, which are not evidence-based, in transfemoral (TF) transcatheter aortic valve implantation (TAVI) METHODS: A patient cohort with a broad range of iliofemoral anatomies undergoing TF TAVI (n=132) were retrospectively divided as “suitable” (n=76, 58%) and “unsuitable” (n=56, 42%) candidates according to current recommendations. Iliofemoral angiography and reconstructed mul- tislice CT (MSCT) images were used for access screening in the majority of patients. RESULTS: Vessel properties were significantly worse in the “unsuitable group.” The sheath-to-iliofemoral artery ratio (SIFAR) and calcium score were 1.35±0.2 and 1.7±0.8 in the unsuitable group, compared to 1.0±0.12 (p<0.0001) and 1.0±0.7 (p=0.0001) in the “suitable” patients. Major vascular complications (MVCs) occurred more frequently in the “unsuitable” group (10.7% vs. 2.6%, p=0.07) and were predicted by SIFAR [OR: 64, 95% CI: 1.4-2971, p=0.03] and circumferential iliofemoral calcifications [OR: 6, 95% CI: 1.2-26, p=0.025]. In the multivariate analysis, circumferential calcifications [HR: 3.6, 95% CI: 1-13.2, p=0.043] but not major vascular complications (MVCs) or manufacturer recommendations were associated with increased mortality. CONCLUSION: According to our results, manufacturer recommendations are safe but overly conservative. Circumferential iliofemoral calcifications may provide independent prognostic information in patients undergoing TAVI. Kare Publishing 2015-04 2014-04-28 /pmc/articles/PMC5336838/ /pubmed/25413227 http://dx.doi.org/10.5152/akd.2014.5311 Text en Copyright © 2016 Turkish Society of Cardiology http://creativecommons.org/licenses/by-nc-sa/4.0 This work is licensed under a Creative Commons Attribution-NonCommercial 4.0 International License
spellingShingle Original Investigation
Reinthaler, Markus
Aggarwal, Sunil K.
Palma, Rodney De
Landmesser, Ulf
Froehlich, Georg
Yap, John
Meier, Pascal
J. Mullen, Michael
Predictors of clinical outcome in transfemoral TAVI: Circumferential iliofemoral calcifications and manufacturer-derived recommendations
title Predictors of clinical outcome in transfemoral TAVI: Circumferential iliofemoral calcifications and manufacturer-derived recommendations
title_full Predictors of clinical outcome in transfemoral TAVI: Circumferential iliofemoral calcifications and manufacturer-derived recommendations
title_fullStr Predictors of clinical outcome in transfemoral TAVI: Circumferential iliofemoral calcifications and manufacturer-derived recommendations
title_full_unstemmed Predictors of clinical outcome in transfemoral TAVI: Circumferential iliofemoral calcifications and manufacturer-derived recommendations
title_short Predictors of clinical outcome in transfemoral TAVI: Circumferential iliofemoral calcifications and manufacturer-derived recommendations
title_sort predictors of clinical outcome in transfemoral tavi: circumferential iliofemoral calcifications and manufacturer-derived recommendations
topic Original Investigation
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5336838/
https://www.ncbi.nlm.nih.gov/pubmed/25413227
http://dx.doi.org/10.5152/akd.2014.5311
work_keys_str_mv AT reinthalermarkus predictorsofclinicaloutcomeintransfemoraltavicircumferentialiliofemoralcalcificationsandmanufacturerderivedrecommendations
AT aggarwalsunilk predictorsofclinicaloutcomeintransfemoraltavicircumferentialiliofemoralcalcificationsandmanufacturerderivedrecommendations
AT palmarodneyde predictorsofclinicaloutcomeintransfemoraltavicircumferentialiliofemoralcalcificationsandmanufacturerderivedrecommendations
AT landmesserulf predictorsofclinicaloutcomeintransfemoraltavicircumferentialiliofemoralcalcificationsandmanufacturerderivedrecommendations
AT froehlichgeorg predictorsofclinicaloutcomeintransfemoraltavicircumferentialiliofemoralcalcificationsandmanufacturerderivedrecommendations
AT yapjohn predictorsofclinicaloutcomeintransfemoraltavicircumferentialiliofemoralcalcificationsandmanufacturerderivedrecommendations
AT meierpascal predictorsofclinicaloutcomeintransfemoraltavicircumferentialiliofemoralcalcificationsandmanufacturerderivedrecommendations
AT jmullenmichael predictorsofclinicaloutcomeintransfemoraltavicircumferentialiliofemoralcalcificationsandmanufacturerderivedrecommendations