Performance of an Automated Fluorescence Antinuclear Antibody Image Analyzer
BACKGROUND: The gold standard for antinuclear antibody (ANA) screening is the indirect immunofluorescence (IIF) assay with human epithelial cells (HEp-2). However, a number of substantial disadvantages of manual IIF assays have highlighted the need for the automation and standardization of fluoresce...
Autores principales: | , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
The Korean Society for Laboratory Medicine
2017
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5339096/ https://www.ncbi.nlm.nih.gov/pubmed/28224770 http://dx.doi.org/10.3343/alm.2017.37.3.240 |
_version_ | 1782512610383495168 |
---|---|
author | Yoo, In Young Oh, Jong Won Cha, Hoon-Suk Koh, Eun-Mi Kang, Eun-Suk |
author_facet | Yoo, In Young Oh, Jong Won Cha, Hoon-Suk Koh, Eun-Mi Kang, Eun-Suk |
author_sort | Yoo, In Young |
collection | PubMed |
description | BACKGROUND: The gold standard for antinuclear antibody (ANA) screening is the indirect immunofluorescence (IIF) assay with human epithelial cells (HEp-2). However, a number of substantial disadvantages of manual IIF assays have highlighted the need for the automation and standardization of fluorescent ANA (FANA) testing. We evaluated the performance of EUROPattern Suite (Euroimmun AG, Germany), an automated FANA image analyzer, with regard to ANA detection and pattern recognition compared with conventional manual interpretation using the fluorescence microscopic IIF assay. METHODS: A total of 104 samples including 70 ANA-positive sera and 34 ANA-negative sera collected from September to October 2015 were included. The sensitivity, specificity, and pattern recognition function were evaluated to determine the performance of EUROPattern Suite compared with the manual IIF assay results. RESULTS: The sensitivity and specificity of EUROPattern Suite for ANA detection were 94.3% and 94.1%, respectively. The concordance rate between the two methods was 94.2%. For pattern recognition, 45.7% of the samples were assigned identical ANA patterns including simple and mixed. When major pattern matching was considered, 83.7% (41/49) and 95.2% (20/21) of the samples with simple and mixed patterns, respectively, showed concordant results between the two methods. CONCLUSIONS: EUROPattern Suite, an automated FANA image analyzer, provides a viable option for distinguishing between positive and negative results, although the ability to assign specific patterns is insufficient to replace manual microscopic interpretation. This automated system may increase efficiency in laboratories, in which a large number of samples need to be processed. |
format | Online Article Text |
id | pubmed-5339096 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2017 |
publisher | The Korean Society for Laboratory Medicine |
record_format | MEDLINE/PubMed |
spelling | pubmed-53390962017-05-01 Performance of an Automated Fluorescence Antinuclear Antibody Image Analyzer Yoo, In Young Oh, Jong Won Cha, Hoon-Suk Koh, Eun-Mi Kang, Eun-Suk Ann Lab Med Original Article BACKGROUND: The gold standard for antinuclear antibody (ANA) screening is the indirect immunofluorescence (IIF) assay with human epithelial cells (HEp-2). However, a number of substantial disadvantages of manual IIF assays have highlighted the need for the automation and standardization of fluorescent ANA (FANA) testing. We evaluated the performance of EUROPattern Suite (Euroimmun AG, Germany), an automated FANA image analyzer, with regard to ANA detection and pattern recognition compared with conventional manual interpretation using the fluorescence microscopic IIF assay. METHODS: A total of 104 samples including 70 ANA-positive sera and 34 ANA-negative sera collected from September to October 2015 were included. The sensitivity, specificity, and pattern recognition function were evaluated to determine the performance of EUROPattern Suite compared with the manual IIF assay results. RESULTS: The sensitivity and specificity of EUROPattern Suite for ANA detection were 94.3% and 94.1%, respectively. The concordance rate between the two methods was 94.2%. For pattern recognition, 45.7% of the samples were assigned identical ANA patterns including simple and mixed. When major pattern matching was considered, 83.7% (41/49) and 95.2% (20/21) of the samples with simple and mixed patterns, respectively, showed concordant results between the two methods. CONCLUSIONS: EUROPattern Suite, an automated FANA image analyzer, provides a viable option for distinguishing between positive and negative results, although the ability to assign specific patterns is insufficient to replace manual microscopic interpretation. This automated system may increase efficiency in laboratories, in which a large number of samples need to be processed. The Korean Society for Laboratory Medicine 2017-05 2017-02-17 /pmc/articles/PMC5339096/ /pubmed/28224770 http://dx.doi.org/10.3343/alm.2017.37.3.240 Text en © The Korean Society for Laboratory Medicine http://creativecommons.org/licenses/by-nc/4.0/ This is an Open Access article distributed under the terms of the Creative Commons Attribution Non-Commercial License (http://creativecommons.org/licenses/by-nc/4.0/) which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited. |
spellingShingle | Original Article Yoo, In Young Oh, Jong Won Cha, Hoon-Suk Koh, Eun-Mi Kang, Eun-Suk Performance of an Automated Fluorescence Antinuclear Antibody Image Analyzer |
title | Performance of an Automated Fluorescence Antinuclear Antibody Image Analyzer |
title_full | Performance of an Automated Fluorescence Antinuclear Antibody Image Analyzer |
title_fullStr | Performance of an Automated Fluorescence Antinuclear Antibody Image Analyzer |
title_full_unstemmed | Performance of an Automated Fluorescence Antinuclear Antibody Image Analyzer |
title_short | Performance of an Automated Fluorescence Antinuclear Antibody Image Analyzer |
title_sort | performance of an automated fluorescence antinuclear antibody image analyzer |
topic | Original Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5339096/ https://www.ncbi.nlm.nih.gov/pubmed/28224770 http://dx.doi.org/10.3343/alm.2017.37.3.240 |
work_keys_str_mv | AT yooinyoung performanceofanautomatedfluorescenceantinuclearantibodyimageanalyzer AT ohjongwon performanceofanautomatedfluorescenceantinuclearantibodyimageanalyzer AT chahoonsuk performanceofanautomatedfluorescenceantinuclearantibodyimageanalyzer AT koheunmi performanceofanautomatedfluorescenceantinuclearantibodyimageanalyzer AT kangeunsuk performanceofanautomatedfluorescenceantinuclearantibodyimageanalyzer |