Accuracy of Heart Rate Measurement Using Smartphones During Treadmill Exercise in Male Patients With Ischemic Heart Disease

OBJECTIVE: To evaluate the accuracy of a smartphone application measuring heart rates (HRs), during an exercise and discussed clinical potential of the smartphone application for cardiac rehabilitation exercise programs. METHODS: Patients with heart disease (14 with myocardial infarction, 2 with ang...

Descripción completa

Detalles Bibliográficos
Autores principales: Lee, Eun Sun, Lee, Jin Seok, Joo, Min Cheol, Kim, Ji Hee, Noh, Se Eung
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Korean Academy of Rehabilitation Medicine 2017
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5344814/
https://www.ncbi.nlm.nih.gov/pubmed/28289645
http://dx.doi.org/10.5535/arm.2017.41.1.129
_version_ 1782513596843950080
author Lee, Eun Sun
Lee, Jin Seok
Joo, Min Cheol
Kim, Ji Hee
Noh, Se Eung
author_facet Lee, Eun Sun
Lee, Jin Seok
Joo, Min Cheol
Kim, Ji Hee
Noh, Se Eung
author_sort Lee, Eun Sun
collection PubMed
description OBJECTIVE: To evaluate the accuracy of a smartphone application measuring heart rates (HRs), during an exercise and discussed clinical potential of the smartphone application for cardiac rehabilitation exercise programs. METHODS: Patients with heart disease (14 with myocardial infarction, 2 with angina pectoris) were recruited. Exercise protocol was comprised of a resting stage, Bruce stage II, Bruce stage III, and a recovery stage. To measure HR, subjects held smartphone in their hands and put the tip of their index finger on the built-in camera for 1 minute at each exercise stage such as resting stage, Bruce stage II, Bruce stage III, and recovery stage. The smartphones recorded photoplethysmography signal and HR was calculated every heart beat. HR data obtained from the smartphone during the exercise protocol was compared with the HR data obtained from a Holter electrocardiography monitor (control). RESULTS: In each exercise protocol stage (resting stage, Bruce stage II, Bruce stage III, and the recovery stage), the HR averages obtained from a Holter monitor were 76.40±12.73, 113.09±14.52, 115.64±15.15, and 81.53±13.08 bpm, respectively. The simultaneously measured HR averages obtained from a smartphone were 76.41±12.82, 112.38±15.06, 115.83±15.36, and 81.53±13 bpm, respectively. The intraclass correlation coefficient (95% confidence interval) was 1.00 (1.00–1.00), 0.99 (0.98–0.99), 0.94 (0.83–0.98), and 1.00 (0.99–1.00) in resting stage, Bruce stage II, Bruce stage III, and recovery stage, respectively. There was no statistically significant difference between the HRs measured by either device at each stage (p>0.05). CONCLUSION: The accuracy of measured HR from a smartphone was almost overlapped with the measurement from the Holter monitor in resting stage and recovery stage. However, we observed that the measurement error increased as the exercise intensity increased.
format Online
Article
Text
id pubmed-5344814
institution National Center for Biotechnology Information
language English
publishDate 2017
publisher Korean Academy of Rehabilitation Medicine
record_format MEDLINE/PubMed
spelling pubmed-53448142017-03-13 Accuracy of Heart Rate Measurement Using Smartphones During Treadmill Exercise in Male Patients With Ischemic Heart Disease Lee, Eun Sun Lee, Jin Seok Joo, Min Cheol Kim, Ji Hee Noh, Se Eung Ann Rehabil Med Original Article OBJECTIVE: To evaluate the accuracy of a smartphone application measuring heart rates (HRs), during an exercise and discussed clinical potential of the smartphone application for cardiac rehabilitation exercise programs. METHODS: Patients with heart disease (14 with myocardial infarction, 2 with angina pectoris) were recruited. Exercise protocol was comprised of a resting stage, Bruce stage II, Bruce stage III, and a recovery stage. To measure HR, subjects held smartphone in their hands and put the tip of their index finger on the built-in camera for 1 minute at each exercise stage such as resting stage, Bruce stage II, Bruce stage III, and recovery stage. The smartphones recorded photoplethysmography signal and HR was calculated every heart beat. HR data obtained from the smartphone during the exercise protocol was compared with the HR data obtained from a Holter electrocardiography monitor (control). RESULTS: In each exercise protocol stage (resting stage, Bruce stage II, Bruce stage III, and the recovery stage), the HR averages obtained from a Holter monitor were 76.40±12.73, 113.09±14.52, 115.64±15.15, and 81.53±13.08 bpm, respectively. The simultaneously measured HR averages obtained from a smartphone were 76.41±12.82, 112.38±15.06, 115.83±15.36, and 81.53±13 bpm, respectively. The intraclass correlation coefficient (95% confidence interval) was 1.00 (1.00–1.00), 0.99 (0.98–0.99), 0.94 (0.83–0.98), and 1.00 (0.99–1.00) in resting stage, Bruce stage II, Bruce stage III, and recovery stage, respectively. There was no statistically significant difference between the HRs measured by either device at each stage (p>0.05). CONCLUSION: The accuracy of measured HR from a smartphone was almost overlapped with the measurement from the Holter monitor in resting stage and recovery stage. However, we observed that the measurement error increased as the exercise intensity increased. Korean Academy of Rehabilitation Medicine 2017-02 2017-02-28 /pmc/articles/PMC5344814/ /pubmed/28289645 http://dx.doi.org/10.5535/arm.2017.41.1.129 Text en Copyright © 2017 by Korean Academy of Rehabilitation Medicine http://creativecommons.org/licenses/by-nc/4.0 This is an open-access article distributed under the terms of the Creative Commons Attribution Non-Commercial License (http://creativecommons.org/licenses/by-nc/4.0) which permits unrestricted noncommercial use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Original Article
Lee, Eun Sun
Lee, Jin Seok
Joo, Min Cheol
Kim, Ji Hee
Noh, Se Eung
Accuracy of Heart Rate Measurement Using Smartphones During Treadmill Exercise in Male Patients With Ischemic Heart Disease
title Accuracy of Heart Rate Measurement Using Smartphones During Treadmill Exercise in Male Patients With Ischemic Heart Disease
title_full Accuracy of Heart Rate Measurement Using Smartphones During Treadmill Exercise in Male Patients With Ischemic Heart Disease
title_fullStr Accuracy of Heart Rate Measurement Using Smartphones During Treadmill Exercise in Male Patients With Ischemic Heart Disease
title_full_unstemmed Accuracy of Heart Rate Measurement Using Smartphones During Treadmill Exercise in Male Patients With Ischemic Heart Disease
title_short Accuracy of Heart Rate Measurement Using Smartphones During Treadmill Exercise in Male Patients With Ischemic Heart Disease
title_sort accuracy of heart rate measurement using smartphones during treadmill exercise in male patients with ischemic heart disease
topic Original Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5344814/
https://www.ncbi.nlm.nih.gov/pubmed/28289645
http://dx.doi.org/10.5535/arm.2017.41.1.129
work_keys_str_mv AT leeeunsun accuracyofheartratemeasurementusingsmartphonesduringtreadmillexerciseinmalepatientswithischemicheartdisease
AT leejinseok accuracyofheartratemeasurementusingsmartphonesduringtreadmillexerciseinmalepatientswithischemicheartdisease
AT joomincheol accuracyofheartratemeasurementusingsmartphonesduringtreadmillexerciseinmalepatientswithischemicheartdisease
AT kimjihee accuracyofheartratemeasurementusingsmartphonesduringtreadmillexerciseinmalepatientswithischemicheartdisease
AT nohseeung accuracyofheartratemeasurementusingsmartphonesduringtreadmillexerciseinmalepatientswithischemicheartdisease