Revisiting Hartert’s 1962 Calculation of the Physical Constants of Thrombelastography
Thrombelastography (TEG)/thromboelastometry (ROTEM) devices measure viscoelastic clot strength as clot amplitude (A). Transformation of clot amplitude into clot elasticity (E with TEG; CE with ROTEM) is sometimes necessary (eg, when calculating platelet component of the clot). With TEG, clot amplitu...
Autores principales: | , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
SAGE Publications
2015
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5349316/ https://www.ncbi.nlm.nih.gov/pubmed/26400661 http://dx.doi.org/10.1177/1076029615606531 |
Sumario: | Thrombelastography (TEG)/thromboelastometry (ROTEM) devices measure viscoelastic clot strength as clot amplitude (A). Transformation of clot amplitude into clot elasticity (E with TEG; CE with ROTEM) is sometimes necessary (eg, when calculating platelet component of the clot). With TEG, clot amplitude is commonly transformed into shear modulus (G; expressed in Pa or dyn/cm(2)) as follows: G = (5000 × A)/(100 – A). Use of the constant “5000” stems from Hartert's 50-year-old calculation of G for a normal blood clot. We question the value of calculating G as follows: (1) It may be questioned whether TEG/ROTEM analysis enable measurement of elasticity because viscosity may also contribute to clot amplitude. (2) It has been suggested that absolute properties of a blood clot cannot be measured with TEG/ROTEM analysis because the strain amplitude applied by the device is uncontrolled and changes during the course of coagulation. (3) A review of the calculation of G using Hartert's methods and some updated assumptions suggests that the value of 5000 is unreliable. (4) Recalculation of G for the ROTEM device yields a different value from that with Hartert TEG, indicating a degree of inaccuracy with the calculations. (5) Shear modulus is simply a multiple of E/CE and, because of the unreliability of G in absolute terms, it provides no additional value versus E/CE. The TEG and ROTEM are valuable coagulation assessment tools that provide an evaluation of the viscoelastic properties of a clot, not through measuring absolute viscoelastic forces but through continuous reading of the clot amplitude relative to an arbitrary, preset scale. |
---|