Non-invasive Measures of Core Temperature versus Ingestible Thermistor during Exercise in the Heat
The accuracy of core temperature (T(c)) thermometry from temporal, tympanic, and oral thermometry devices has been variable during exercise in a hot, humid environment. The purpose of the present study was to cross-validate temporal, two tympanic devices, and oral devices compared to an ingestible t...
Autores principales: | , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Berkeley Electronic Press
2017
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5360367/ https://www.ncbi.nlm.nih.gov/pubmed/28344737 |
Sumario: | The accuracy of core temperature (T(c)) thermometry from temporal, tympanic, and oral thermometry devices has been variable during exercise in a hot, humid environment. The purpose of the present study was to cross-validate temporal, two tympanic devices, and oral devices compared to an ingestible thermistor during exercise in a hot, humid environment. Fourteen young, active adults (6 women) completed a graded exercise test until voluntary exhaustion in an environmental chamber (35.5 ± 0.6 °C, 53.9 ± 5.8 % RH). There was no statistical difference in mean temperature between tympanic device 1 and pill-based core temperature (PBT(c)) measurements across all time points and were positively correlated (0.357; P<0.001). Temperatures of tympanic device 2 were statistically higher than PBT(c) (37.8 ± 0.7 ºC vs. 37.6 ± 1.0 ºC; respectfully) (P=0.008). At all time points, temperatures for the second tympanic device and PBT(c) were positively correlated (0.192; P=0.043). Temporal and PBT(c) values did not differ across time points and were positively correlated (0.262; P=0.005) across all time points. Mean oral temperature was significantly less than mean PBT(c) across all time points. (37.0 ± 0.4 ºC vs. 37.6 ± 1.0 ºC, respectively) (P<0.001). Across all time points, oral and PBT(c) were positively correlated (0.262; P=0.010). Tympanic and temporal devices can reflect T(c) while exercising in a hot, humid environment. However, care should be taken when selecting the tympanic or temporal measurement device and validation is advised prior to heat illness mitigation in the field. |
---|