Continuous Non-invasive finger cuff CareTaker® comparable to invasive intra-arterial pressure in patients undergoing major intra-abdominal surgery
BACKGROUND: Despite increased interest in non-invasive arterial pressure monitoring, the majority of commercially available technologies have failed to satisfy the limits established for the validation of automatic arterial pressure monitoring by the Association for the Advancement of Medical Instru...
Autores principales: | , , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
BioMed Central
2017
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5361833/ https://www.ncbi.nlm.nih.gov/pubmed/28327093 http://dx.doi.org/10.1186/s12871-017-0337-z |
_version_ | 1782516845428867072 |
---|---|
author | Gratz, Irwin Deal, Edward Spitz, Francis Baruch, Martin Allen, I. Elaine Seaman, Julia E. Pukenas, Erin Jean, Smith |
author_facet | Gratz, Irwin Deal, Edward Spitz, Francis Baruch, Martin Allen, I. Elaine Seaman, Julia E. Pukenas, Erin Jean, Smith |
author_sort | Gratz, Irwin |
collection | PubMed |
description | BACKGROUND: Despite increased interest in non-invasive arterial pressure monitoring, the majority of commercially available technologies have failed to satisfy the limits established for the validation of automatic arterial pressure monitoring by the Association for the Advancement of Medical Instrumentation (AAMI). According to the ANSI/AAMI/ISO 81060–2:2013 standards, the group-average accuracy and precision are defined as acceptable if bias is not greater than 5 mmHg and standard deviation is not greater than 8 mmHg. In this study, these standards are used to evaluate the CareTaker® (CT) device, a device measuring continuous non-invasive blood pressure via a pulse contour algorithm called Pulse Decomposition Analysis. METHODS: A convenience sample of 24 patients scheduled for major abdominal surgery were consented to participate in this IRB approved pilot study. Each patient was monitored with a radial arterial catheter and CT using a finger cuff applied to the contralateral thumb. Hemodynamic variables were measured and analyzed from both devices for the first thirty minutes of the surgical procedure including the induction of anesthesia. The mean arterial pressure (MAP), systolic and diastolic blood pressures continuously collected from the arterial catheter and CT were compared. Pearson correlation coefficients were calculated between arterial catheter and CT blood pressure measurements, a Bland-Altman analysis, and polar and 4Q plots were created. RESULTS: The correlation of systolic, diastolic, and mean arterial pressures were 0.92, 0.86, 0.91, respectively (p < 0.0001 for all the comparisons). The Bland-Altman comparison yielded a bias (as measured by overall mean difference) of −0.57, −2.52, 1.01 mmHg for systolic, diastolic, and mean arterial pressures, respectively with a standard deviation of 7.34, 6.47, 5.33 mmHg for systolic, diastolic, and mean arterial pressures, respectively (p < 0.001 for all comparisons). The polar plot indicates little bias between the two methods (90%/95% CI at 31.5°/52°, respectively, overall bias = 1.5°) with only a small percentage of points outside these lines. The 4Q plot indicates good concordance and no bias between the methods. CONCLUSIONS: In this study, blood pressure measured using the non-invasive CT device was shown to correlate well with the arterial catheter measurements. Larger studies are needed to confirm these results in more varied settings. Most patients exhibited very good agreement between methods. Results were well within the limits established for the validation of automatic arterial pressure monitoring by the AAMI. |
format | Online Article Text |
id | pubmed-5361833 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2017 |
publisher | BioMed Central |
record_format | MEDLINE/PubMed |
spelling | pubmed-53618332017-03-24 Continuous Non-invasive finger cuff CareTaker® comparable to invasive intra-arterial pressure in patients undergoing major intra-abdominal surgery Gratz, Irwin Deal, Edward Spitz, Francis Baruch, Martin Allen, I. Elaine Seaman, Julia E. Pukenas, Erin Jean, Smith BMC Anesthesiol Research Article BACKGROUND: Despite increased interest in non-invasive arterial pressure monitoring, the majority of commercially available technologies have failed to satisfy the limits established for the validation of automatic arterial pressure monitoring by the Association for the Advancement of Medical Instrumentation (AAMI). According to the ANSI/AAMI/ISO 81060–2:2013 standards, the group-average accuracy and precision are defined as acceptable if bias is not greater than 5 mmHg and standard deviation is not greater than 8 mmHg. In this study, these standards are used to evaluate the CareTaker® (CT) device, a device measuring continuous non-invasive blood pressure via a pulse contour algorithm called Pulse Decomposition Analysis. METHODS: A convenience sample of 24 patients scheduled for major abdominal surgery were consented to participate in this IRB approved pilot study. Each patient was monitored with a radial arterial catheter and CT using a finger cuff applied to the contralateral thumb. Hemodynamic variables were measured and analyzed from both devices for the first thirty minutes of the surgical procedure including the induction of anesthesia. The mean arterial pressure (MAP), systolic and diastolic blood pressures continuously collected from the arterial catheter and CT were compared. Pearson correlation coefficients were calculated between arterial catheter and CT blood pressure measurements, a Bland-Altman analysis, and polar and 4Q plots were created. RESULTS: The correlation of systolic, diastolic, and mean arterial pressures were 0.92, 0.86, 0.91, respectively (p < 0.0001 for all the comparisons). The Bland-Altman comparison yielded a bias (as measured by overall mean difference) of −0.57, −2.52, 1.01 mmHg for systolic, diastolic, and mean arterial pressures, respectively with a standard deviation of 7.34, 6.47, 5.33 mmHg for systolic, diastolic, and mean arterial pressures, respectively (p < 0.001 for all comparisons). The polar plot indicates little bias between the two methods (90%/95% CI at 31.5°/52°, respectively, overall bias = 1.5°) with only a small percentage of points outside these lines. The 4Q plot indicates good concordance and no bias between the methods. CONCLUSIONS: In this study, blood pressure measured using the non-invasive CT device was shown to correlate well with the arterial catheter measurements. Larger studies are needed to confirm these results in more varied settings. Most patients exhibited very good agreement between methods. Results were well within the limits established for the validation of automatic arterial pressure monitoring by the AAMI. BioMed Central 2017-03-21 /pmc/articles/PMC5361833/ /pubmed/28327093 http://dx.doi.org/10.1186/s12871-017-0337-z Text en © The Author(s). 2017 Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated. |
spellingShingle | Research Article Gratz, Irwin Deal, Edward Spitz, Francis Baruch, Martin Allen, I. Elaine Seaman, Julia E. Pukenas, Erin Jean, Smith Continuous Non-invasive finger cuff CareTaker® comparable to invasive intra-arterial pressure in patients undergoing major intra-abdominal surgery |
title | Continuous Non-invasive finger cuff CareTaker® comparable to invasive intra-arterial pressure in patients undergoing major intra-abdominal surgery |
title_full | Continuous Non-invasive finger cuff CareTaker® comparable to invasive intra-arterial pressure in patients undergoing major intra-abdominal surgery |
title_fullStr | Continuous Non-invasive finger cuff CareTaker® comparable to invasive intra-arterial pressure in patients undergoing major intra-abdominal surgery |
title_full_unstemmed | Continuous Non-invasive finger cuff CareTaker® comparable to invasive intra-arterial pressure in patients undergoing major intra-abdominal surgery |
title_short | Continuous Non-invasive finger cuff CareTaker® comparable to invasive intra-arterial pressure in patients undergoing major intra-abdominal surgery |
title_sort | continuous non-invasive finger cuff caretaker® comparable to invasive intra-arterial pressure in patients undergoing major intra-abdominal surgery |
topic | Research Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5361833/ https://www.ncbi.nlm.nih.gov/pubmed/28327093 http://dx.doi.org/10.1186/s12871-017-0337-z |
work_keys_str_mv | AT gratzirwin continuousnoninvasivefingercuffcaretakercomparabletoinvasiveintraarterialpressureinpatientsundergoingmajorintraabdominalsurgery AT dealedward continuousnoninvasivefingercuffcaretakercomparabletoinvasiveintraarterialpressureinpatientsundergoingmajorintraabdominalsurgery AT spitzfrancis continuousnoninvasivefingercuffcaretakercomparabletoinvasiveintraarterialpressureinpatientsundergoingmajorintraabdominalsurgery AT baruchmartin continuousnoninvasivefingercuffcaretakercomparabletoinvasiveintraarterialpressureinpatientsundergoingmajorintraabdominalsurgery AT allenielaine continuousnoninvasivefingercuffcaretakercomparabletoinvasiveintraarterialpressureinpatientsundergoingmajorintraabdominalsurgery AT seamanjuliae continuousnoninvasivefingercuffcaretakercomparabletoinvasiveintraarterialpressureinpatientsundergoingmajorintraabdominalsurgery AT pukenaserin continuousnoninvasivefingercuffcaretakercomparabletoinvasiveintraarterialpressureinpatientsundergoingmajorintraabdominalsurgery AT jeansmith continuousnoninvasivefingercuffcaretakercomparabletoinvasiveintraarterialpressureinpatientsundergoingmajorintraabdominalsurgery |