Continuous Non-invasive finger cuff CareTaker® comparable to invasive intra-arterial pressure in patients undergoing major intra-abdominal surgery

BACKGROUND: Despite increased interest in non-invasive arterial pressure monitoring, the majority of commercially available technologies have failed to satisfy the limits established for the validation of automatic arterial pressure monitoring by the Association for the Advancement of Medical Instru...

Descripción completa

Detalles Bibliográficos
Autores principales: Gratz, Irwin, Deal, Edward, Spitz, Francis, Baruch, Martin, Allen, I. Elaine, Seaman, Julia E., Pukenas, Erin, Jean, Smith
Formato: Online Artículo Texto
Lenguaje:English
Publicado: BioMed Central 2017
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5361833/
https://www.ncbi.nlm.nih.gov/pubmed/28327093
http://dx.doi.org/10.1186/s12871-017-0337-z
_version_ 1782516845428867072
author Gratz, Irwin
Deal, Edward
Spitz, Francis
Baruch, Martin
Allen, I. Elaine
Seaman, Julia E.
Pukenas, Erin
Jean, Smith
author_facet Gratz, Irwin
Deal, Edward
Spitz, Francis
Baruch, Martin
Allen, I. Elaine
Seaman, Julia E.
Pukenas, Erin
Jean, Smith
author_sort Gratz, Irwin
collection PubMed
description BACKGROUND: Despite increased interest in non-invasive arterial pressure monitoring, the majority of commercially available technologies have failed to satisfy the limits established for the validation of automatic arterial pressure monitoring by the Association for the Advancement of Medical Instrumentation (AAMI). According to the ANSI/AAMI/ISO 81060–2:2013 standards, the group-average accuracy and precision are defined as acceptable if bias is not greater than 5 mmHg and standard deviation is not greater than 8 mmHg. In this study, these standards are used to evaluate the CareTaker® (CT) device, a device measuring continuous non-invasive blood pressure via a pulse contour algorithm called Pulse Decomposition Analysis. METHODS: A convenience sample of 24 patients scheduled for major abdominal surgery were consented to participate in this IRB approved pilot study. Each patient was monitored with a radial arterial catheter and CT using a finger cuff applied to the contralateral thumb. Hemodynamic variables were measured and analyzed from both devices for the first thirty minutes of the surgical procedure including the induction of anesthesia. The mean arterial pressure (MAP), systolic and diastolic blood pressures continuously collected from the arterial catheter and CT were compared. Pearson correlation coefficients were calculated between arterial catheter and CT blood pressure measurements, a Bland-Altman analysis, and polar and 4Q plots were created. RESULTS: The correlation of systolic, diastolic, and mean arterial pressures were 0.92, 0.86, 0.91, respectively (p < 0.0001 for all the comparisons). The Bland-Altman comparison yielded a bias (as measured by overall mean difference) of −0.57, −2.52, 1.01 mmHg for systolic, diastolic, and mean arterial pressures, respectively with a standard deviation of 7.34, 6.47, 5.33 mmHg for systolic, diastolic, and mean arterial pressures, respectively (p < 0.001 for all comparisons). The polar plot indicates little bias between the two methods (90%/95% CI at 31.5°/52°, respectively, overall bias = 1.5°) with only a small percentage of points outside these lines. The 4Q plot indicates good concordance and no bias between the methods. CONCLUSIONS: In this study, blood pressure measured using the non-invasive CT device was shown to correlate well with the arterial catheter measurements. Larger studies are needed to confirm these results in more varied settings. Most patients exhibited very good agreement between methods. Results were well within the limits established for the validation of automatic arterial pressure monitoring by the AAMI.
format Online
Article
Text
id pubmed-5361833
institution National Center for Biotechnology Information
language English
publishDate 2017
publisher BioMed Central
record_format MEDLINE/PubMed
spelling pubmed-53618332017-03-24 Continuous Non-invasive finger cuff CareTaker® comparable to invasive intra-arterial pressure in patients undergoing major intra-abdominal surgery Gratz, Irwin Deal, Edward Spitz, Francis Baruch, Martin Allen, I. Elaine Seaman, Julia E. Pukenas, Erin Jean, Smith BMC Anesthesiol Research Article BACKGROUND: Despite increased interest in non-invasive arterial pressure monitoring, the majority of commercially available technologies have failed to satisfy the limits established for the validation of automatic arterial pressure monitoring by the Association for the Advancement of Medical Instrumentation (AAMI). According to the ANSI/AAMI/ISO 81060–2:2013 standards, the group-average accuracy and precision are defined as acceptable if bias is not greater than 5 mmHg and standard deviation is not greater than 8 mmHg. In this study, these standards are used to evaluate the CareTaker® (CT) device, a device measuring continuous non-invasive blood pressure via a pulse contour algorithm called Pulse Decomposition Analysis. METHODS: A convenience sample of 24 patients scheduled for major abdominal surgery were consented to participate in this IRB approved pilot study. Each patient was monitored with a radial arterial catheter and CT using a finger cuff applied to the contralateral thumb. Hemodynamic variables were measured and analyzed from both devices for the first thirty minutes of the surgical procedure including the induction of anesthesia. The mean arterial pressure (MAP), systolic and diastolic blood pressures continuously collected from the arterial catheter and CT were compared. Pearson correlation coefficients were calculated between arterial catheter and CT blood pressure measurements, a Bland-Altman analysis, and polar and 4Q plots were created. RESULTS: The correlation of systolic, diastolic, and mean arterial pressures were 0.92, 0.86, 0.91, respectively (p < 0.0001 for all the comparisons). The Bland-Altman comparison yielded a bias (as measured by overall mean difference) of −0.57, −2.52, 1.01 mmHg for systolic, diastolic, and mean arterial pressures, respectively with a standard deviation of 7.34, 6.47, 5.33 mmHg for systolic, diastolic, and mean arterial pressures, respectively (p < 0.001 for all comparisons). The polar plot indicates little bias between the two methods (90%/95% CI at 31.5°/52°, respectively, overall bias = 1.5°) with only a small percentage of points outside these lines. The 4Q plot indicates good concordance and no bias between the methods. CONCLUSIONS: In this study, blood pressure measured using the non-invasive CT device was shown to correlate well with the arterial catheter measurements. Larger studies are needed to confirm these results in more varied settings. Most patients exhibited very good agreement between methods. Results were well within the limits established for the validation of automatic arterial pressure monitoring by the AAMI. BioMed Central 2017-03-21 /pmc/articles/PMC5361833/ /pubmed/28327093 http://dx.doi.org/10.1186/s12871-017-0337-z Text en © The Author(s). 2017 Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.
spellingShingle Research Article
Gratz, Irwin
Deal, Edward
Spitz, Francis
Baruch, Martin
Allen, I. Elaine
Seaman, Julia E.
Pukenas, Erin
Jean, Smith
Continuous Non-invasive finger cuff CareTaker® comparable to invasive intra-arterial pressure in patients undergoing major intra-abdominal surgery
title Continuous Non-invasive finger cuff CareTaker® comparable to invasive intra-arterial pressure in patients undergoing major intra-abdominal surgery
title_full Continuous Non-invasive finger cuff CareTaker® comparable to invasive intra-arterial pressure in patients undergoing major intra-abdominal surgery
title_fullStr Continuous Non-invasive finger cuff CareTaker® comparable to invasive intra-arterial pressure in patients undergoing major intra-abdominal surgery
title_full_unstemmed Continuous Non-invasive finger cuff CareTaker® comparable to invasive intra-arterial pressure in patients undergoing major intra-abdominal surgery
title_short Continuous Non-invasive finger cuff CareTaker® comparable to invasive intra-arterial pressure in patients undergoing major intra-abdominal surgery
title_sort continuous non-invasive finger cuff caretaker® comparable to invasive intra-arterial pressure in patients undergoing major intra-abdominal surgery
topic Research Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5361833/
https://www.ncbi.nlm.nih.gov/pubmed/28327093
http://dx.doi.org/10.1186/s12871-017-0337-z
work_keys_str_mv AT gratzirwin continuousnoninvasivefingercuffcaretakercomparabletoinvasiveintraarterialpressureinpatientsundergoingmajorintraabdominalsurgery
AT dealedward continuousnoninvasivefingercuffcaretakercomparabletoinvasiveintraarterialpressureinpatientsundergoingmajorintraabdominalsurgery
AT spitzfrancis continuousnoninvasivefingercuffcaretakercomparabletoinvasiveintraarterialpressureinpatientsundergoingmajorintraabdominalsurgery
AT baruchmartin continuousnoninvasivefingercuffcaretakercomparabletoinvasiveintraarterialpressureinpatientsundergoingmajorintraabdominalsurgery
AT allenielaine continuousnoninvasivefingercuffcaretakercomparabletoinvasiveintraarterialpressureinpatientsundergoingmajorintraabdominalsurgery
AT seamanjuliae continuousnoninvasivefingercuffcaretakercomparabletoinvasiveintraarterialpressureinpatientsundergoingmajorintraabdominalsurgery
AT pukenaserin continuousnoninvasivefingercuffcaretakercomparabletoinvasiveintraarterialpressureinpatientsundergoingmajorintraabdominalsurgery
AT jeansmith continuousnoninvasivefingercuffcaretakercomparabletoinvasiveintraarterialpressureinpatientsundergoingmajorintraabdominalsurgery