Multicenter Study Validating Accuracy of a Continuous Respiratory Rate Measurement Derived From Pulse Oximetry: A Comparison With Capnography

BACKGROUND: Intermittent measurement of respiratory rate via observation is routine in many patient care settings. This approach has several inherent limitations that diminish the clinical utility of these measurements because it is intermittent, susceptible to human error, and requires clinical res...

Descripción completa

Detalles Bibliográficos
Autores principales: Bergese, Sergio D., Mestek, Michael L., Kelley, Scott D., McIntyre, Robert, Uribe, Alberto A., Sethi, Rakesh, Watson, James N., Addison, Paul S.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Lippincott Williams & Wilkins 2017
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5367492/
https://www.ncbi.nlm.nih.gov/pubmed/28099286
http://dx.doi.org/10.1213/ANE.0000000000001852
_version_ 1782517780342374400
author Bergese, Sergio D.
Mestek, Michael L.
Kelley, Scott D.
McIntyre, Robert
Uribe, Alberto A.
Sethi, Rakesh
Watson, James N.
Addison, Paul S.
author_facet Bergese, Sergio D.
Mestek, Michael L.
Kelley, Scott D.
McIntyre, Robert
Uribe, Alberto A.
Sethi, Rakesh
Watson, James N.
Addison, Paul S.
author_sort Bergese, Sergio D.
collection PubMed
description BACKGROUND: Intermittent measurement of respiratory rate via observation is routine in many patient care settings. This approach has several inherent limitations that diminish the clinical utility of these measurements because it is intermittent, susceptible to human error, and requires clinical resources. As an alternative, a software application that derives continuous respiratory rate measurement from a standard pulse oximeter has been developed. We sought to determine the performance characteristics of this new technology by comparison with clinician-reviewed capnography waveforms in both healthy subjects and hospitalized patients in a low-acuity care setting. METHODS: Two independent observational studies were conducted to validate the performance of the Medtronic Nellcor(TM) Respiration Rate Software application. One study enrolled 26 healthy volunteer subjects in a clinical laboratory, and a second multicenter study enrolled 53 hospitalized patients. During a 30-minute study period taking place while participants were breathing spontaneously, pulse oximeter and nasal/oral capnography waveforms were collected. Pulse oximeter waveforms were processed to determine respiratory rate via the Medtronic Nellcor Respiration Rate Software. Capnography waveforms reviewed by a clinician were used to determine the reference respiratory rate. RESULTS: A total of 23,243 paired observations between the pulse oximeter-derived respiratory rate and the capnography reference method were collected and examined. The mean reference-based respiratory rate was 15.3 ± 4.3 breaths per minute with a range of 4 to 34 breaths per minute. The Pearson correlation coefficient between the Medtronic Nellcor Respiration Rate Software values and the capnography reference respiratory rate is reported as a linear correlation, R, as 0.92 ± 0.02 (P < .001), whereas Lin’s concordance correlation coefficient indicates an overall agreement of 0.85 ± 0.04 (95% confidence interval [CI] +0.76; +0.93) (healthy volunteers: 0.94 ± 0.02 [95% CI +0.91; +0.97]; hospitalized patients: 0.80 ± 0.06 [95% CI +0.68; +0.92]). The mean bias of the Medtronic Nellcor Respiration Rate Software was 0.18 breaths per minute with a precision (SD) of 1.65 breaths per minute (healthy volunteers: 0.37 ± 0.78 [95% limits of agreement: –1.16; +1.90] breaths per minute; hospitalized patients: 0.07 ± 1.99 [95% limits of agreement: –3.84; +3.97] breaths per minute). The root mean square deviation was 1.35 breaths per minute (healthy volunteers: 0.81; hospitalized patients: 1.60). CONCLUSIONS: These data demonstrate the performance of the Medtronic Nellcor Respiration Rate Software in healthy subjects and patients hospitalized in a low-acuity care setting when compared with clinician-reviewed capnography. The observed performance of this technology suggests that it may be a useful adjunct to continuous pulse oximetry monitoring by providing continuous respiratory rate measurements. The potential patient safety benefit of using combined continuous pulse oximetry and respiratory rate monitoring warrants assessment.
format Online
Article
Text
id pubmed-5367492
institution National Center for Biotechnology Information
language English
publishDate 2017
publisher Lippincott Williams & Wilkins
record_format MEDLINE/PubMed
spelling pubmed-53674922017-04-07 Multicenter Study Validating Accuracy of a Continuous Respiratory Rate Measurement Derived From Pulse Oximetry: A Comparison With Capnography Bergese, Sergio D. Mestek, Michael L. Kelley, Scott D. McIntyre, Robert Uribe, Alberto A. Sethi, Rakesh Watson, James N. Addison, Paul S. Anesth Analg Technology, Computing, and Simulation BACKGROUND: Intermittent measurement of respiratory rate via observation is routine in many patient care settings. This approach has several inherent limitations that diminish the clinical utility of these measurements because it is intermittent, susceptible to human error, and requires clinical resources. As an alternative, a software application that derives continuous respiratory rate measurement from a standard pulse oximeter has been developed. We sought to determine the performance characteristics of this new technology by comparison with clinician-reviewed capnography waveforms in both healthy subjects and hospitalized patients in a low-acuity care setting. METHODS: Two independent observational studies were conducted to validate the performance of the Medtronic Nellcor(TM) Respiration Rate Software application. One study enrolled 26 healthy volunteer subjects in a clinical laboratory, and a second multicenter study enrolled 53 hospitalized patients. During a 30-minute study period taking place while participants were breathing spontaneously, pulse oximeter and nasal/oral capnography waveforms were collected. Pulse oximeter waveforms were processed to determine respiratory rate via the Medtronic Nellcor Respiration Rate Software. Capnography waveforms reviewed by a clinician were used to determine the reference respiratory rate. RESULTS: A total of 23,243 paired observations between the pulse oximeter-derived respiratory rate and the capnography reference method were collected and examined. The mean reference-based respiratory rate was 15.3 ± 4.3 breaths per minute with a range of 4 to 34 breaths per minute. The Pearson correlation coefficient between the Medtronic Nellcor Respiration Rate Software values and the capnography reference respiratory rate is reported as a linear correlation, R, as 0.92 ± 0.02 (P < .001), whereas Lin’s concordance correlation coefficient indicates an overall agreement of 0.85 ± 0.04 (95% confidence interval [CI] +0.76; +0.93) (healthy volunteers: 0.94 ± 0.02 [95% CI +0.91; +0.97]; hospitalized patients: 0.80 ± 0.06 [95% CI +0.68; +0.92]). The mean bias of the Medtronic Nellcor Respiration Rate Software was 0.18 breaths per minute with a precision (SD) of 1.65 breaths per minute (healthy volunteers: 0.37 ± 0.78 [95% limits of agreement: –1.16; +1.90] breaths per minute; hospitalized patients: 0.07 ± 1.99 [95% limits of agreement: –3.84; +3.97] breaths per minute). The root mean square deviation was 1.35 breaths per minute (healthy volunteers: 0.81; hospitalized patients: 1.60). CONCLUSIONS: These data demonstrate the performance of the Medtronic Nellcor Respiration Rate Software in healthy subjects and patients hospitalized in a low-acuity care setting when compared with clinician-reviewed capnography. The observed performance of this technology suggests that it may be a useful adjunct to continuous pulse oximetry monitoring by providing continuous respiratory rate measurements. The potential patient safety benefit of using combined continuous pulse oximetry and respiratory rate monitoring warrants assessment. Lippincott Williams & Wilkins 2017-04 2017-01-17 /pmc/articles/PMC5367492/ /pubmed/28099286 http://dx.doi.org/10.1213/ANE.0000000000001852 Text en Copyright © 2017 The Author(s). Published by Wolters Kluwer Health, Inc. on behalf of the International Anesthesia Research Society. This is an open-access article distributed under the terms of the Creative Commons Attribution-Non Commercial-No Derivatives License 4.0 (CCBY-NC-ND) (http://creativecommons.org/licenses/by-nc-nd/4.0/) , where it is permissible to download and share the work provided it is properly cited. The work cannot be changed in any way or used commercially without permission from the journal.
spellingShingle Technology, Computing, and Simulation
Bergese, Sergio D.
Mestek, Michael L.
Kelley, Scott D.
McIntyre, Robert
Uribe, Alberto A.
Sethi, Rakesh
Watson, James N.
Addison, Paul S.
Multicenter Study Validating Accuracy of a Continuous Respiratory Rate Measurement Derived From Pulse Oximetry: A Comparison With Capnography
title Multicenter Study Validating Accuracy of a Continuous Respiratory Rate Measurement Derived From Pulse Oximetry: A Comparison With Capnography
title_full Multicenter Study Validating Accuracy of a Continuous Respiratory Rate Measurement Derived From Pulse Oximetry: A Comparison With Capnography
title_fullStr Multicenter Study Validating Accuracy of a Continuous Respiratory Rate Measurement Derived From Pulse Oximetry: A Comparison With Capnography
title_full_unstemmed Multicenter Study Validating Accuracy of a Continuous Respiratory Rate Measurement Derived From Pulse Oximetry: A Comparison With Capnography
title_short Multicenter Study Validating Accuracy of a Continuous Respiratory Rate Measurement Derived From Pulse Oximetry: A Comparison With Capnography
title_sort multicenter study validating accuracy of a continuous respiratory rate measurement derived from pulse oximetry: a comparison with capnography
topic Technology, Computing, and Simulation
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5367492/
https://www.ncbi.nlm.nih.gov/pubmed/28099286
http://dx.doi.org/10.1213/ANE.0000000000001852
work_keys_str_mv AT bergesesergiod multicenterstudyvalidatingaccuracyofacontinuousrespiratoryratemeasurementderivedfrompulseoximetryacomparisonwithcapnography
AT mestekmichaell multicenterstudyvalidatingaccuracyofacontinuousrespiratoryratemeasurementderivedfrompulseoximetryacomparisonwithcapnography
AT kelleyscottd multicenterstudyvalidatingaccuracyofacontinuousrespiratoryratemeasurementderivedfrompulseoximetryacomparisonwithcapnography
AT mcintyrerobert multicenterstudyvalidatingaccuracyofacontinuousrespiratoryratemeasurementderivedfrompulseoximetryacomparisonwithcapnography
AT uribealbertoa multicenterstudyvalidatingaccuracyofacontinuousrespiratoryratemeasurementderivedfrompulseoximetryacomparisonwithcapnography
AT sethirakesh multicenterstudyvalidatingaccuracyofacontinuousrespiratoryratemeasurementderivedfrompulseoximetryacomparisonwithcapnography
AT watsonjamesn multicenterstudyvalidatingaccuracyofacontinuousrespiratoryratemeasurementderivedfrompulseoximetryacomparisonwithcapnography
AT addisonpauls multicenterstudyvalidatingaccuracyofacontinuousrespiratoryratemeasurementderivedfrompulseoximetryacomparisonwithcapnography