Measurement of Total Scatter Factor for Stereotactic Cones with Plastic Scintillation Detector

Advanced radiotherapy modalities such as stereotactic radiosurgery (SRS) and image-guided radiotherapy may employ very small beam apertures for accurate localized high dose to target. Accurate measurement of small radiation fields is a well-known challenge for many dosimeters. The purpose of this st...

Descripción completa

Detalles Bibliográficos
Autores principales: Chaudhari, Suresh H, Dobhal, Rishabh, Kinhikar, Rajesh A., Kadam, Sudarshan S., Deshpande, Deepak D.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Medknow Publications & Media Pvt Ltd 2017
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5370342/
https://www.ncbi.nlm.nih.gov/pubmed/28405102
http://dx.doi.org/10.4103/jmp.JMP_114_16
_version_ 1782518221265436672
author Chaudhari, Suresh H
Dobhal, Rishabh
Kinhikar, Rajesh A.
Kadam, Sudarshan S.
Deshpande, Deepak D.
author_facet Chaudhari, Suresh H
Dobhal, Rishabh
Kinhikar, Rajesh A.
Kadam, Sudarshan S.
Deshpande, Deepak D.
author_sort Chaudhari, Suresh H
collection PubMed
description Advanced radiotherapy modalities such as stereotactic radiosurgery (SRS) and image-guided radiotherapy may employ very small beam apertures for accurate localized high dose to target. Accurate measurement of small radiation fields is a well-known challenge for many dosimeters. The purpose of this study was to measure total scatter factors for stereotactic cones with plastic scintillation detector and its comparison against diode detector and theoretical estimates. Measurements were performed on Novalis Tx(™) linear accelerator for 6MV SRS beam with stereotactic cones of diameter 6 mm, 7.5 mm, 10 mm, 12.5 mm, and 15 mm. The advantage of plastic scintillator detector is in its energy dependence. The total scatter factor was measured in water at the depth of dose maximum. Total scatter factor with plastic scintillation detector was determined by normalizing the readings to field size of 10 cm × 10 cm. To overcome energy dependence of diode detector for the determination of scatter factor with diode detector, daisy chaining method was used. The plastic scintillator detector was calibrated against the ionization chamber, and the reproducibility in the measured doses was found to be within ± 1%. Total scatter factor measured with plastic scintillation detector was 0.728 ± 0.3, 0.783 ± 0.05, 0.866 ± 0.55, 0.885 ± 0.5, and 0.910 ± 0.06 for cone sizes of 6 mm, 7.5 mm, 10 mm, 12.5 mm, and 15 mm, respectively. Total scatter factor measured with diode detector was 0.733 ± 0.03, 0.782 ± 0.02, 0.834 ± 0.07, 0.854 ± 0.02, and 0.872 ± 0.02 for cone sizes of 6 mm, 7.5 mm, 10 mm, 12.5 mm, and 15 mm, respectively. The variation in the measurement of total scatter factor with published Monte Carlo data was found to be −1.3%, 1.9%, −0.4%, and 0.4% for cone sizes of 7.5 mm, 10 mm, 12.5 mm, and 15 mm, respectively. We conclude that total scatter factor measurements for stereotactic cones can be adequately carried out with a plastic scintillation detector. Our results show a high level of consistency within our data and compared well with published data.
format Online
Article
Text
id pubmed-5370342
institution National Center for Biotechnology Information
language English
publishDate 2017
publisher Medknow Publications & Media Pvt Ltd
record_format MEDLINE/PubMed
spelling pubmed-53703422017-04-12 Measurement of Total Scatter Factor for Stereotactic Cones with Plastic Scintillation Detector Chaudhari, Suresh H Dobhal, Rishabh Kinhikar, Rajesh A. Kadam, Sudarshan S. Deshpande, Deepak D. J Med Phys Original Article Advanced radiotherapy modalities such as stereotactic radiosurgery (SRS) and image-guided radiotherapy may employ very small beam apertures for accurate localized high dose to target. Accurate measurement of small radiation fields is a well-known challenge for many dosimeters. The purpose of this study was to measure total scatter factors for stereotactic cones with plastic scintillation detector and its comparison against diode detector and theoretical estimates. Measurements were performed on Novalis Tx(™) linear accelerator for 6MV SRS beam with stereotactic cones of diameter 6 mm, 7.5 mm, 10 mm, 12.5 mm, and 15 mm. The advantage of plastic scintillator detector is in its energy dependence. The total scatter factor was measured in water at the depth of dose maximum. Total scatter factor with plastic scintillation detector was determined by normalizing the readings to field size of 10 cm × 10 cm. To overcome energy dependence of diode detector for the determination of scatter factor with diode detector, daisy chaining method was used. The plastic scintillator detector was calibrated against the ionization chamber, and the reproducibility in the measured doses was found to be within ± 1%. Total scatter factor measured with plastic scintillation detector was 0.728 ± 0.3, 0.783 ± 0.05, 0.866 ± 0.55, 0.885 ± 0.5, and 0.910 ± 0.06 for cone sizes of 6 mm, 7.5 mm, 10 mm, 12.5 mm, and 15 mm, respectively. Total scatter factor measured with diode detector was 0.733 ± 0.03, 0.782 ± 0.02, 0.834 ± 0.07, 0.854 ± 0.02, and 0.872 ± 0.02 for cone sizes of 6 mm, 7.5 mm, 10 mm, 12.5 mm, and 15 mm, respectively. The variation in the measurement of total scatter factor with published Monte Carlo data was found to be −1.3%, 1.9%, −0.4%, and 0.4% for cone sizes of 7.5 mm, 10 mm, 12.5 mm, and 15 mm, respectively. We conclude that total scatter factor measurements for stereotactic cones can be adequately carried out with a plastic scintillation detector. Our results show a high level of consistency within our data and compared well with published data. Medknow Publications & Media Pvt Ltd 2017 /pmc/articles/PMC5370342/ /pubmed/28405102 http://dx.doi.org/10.4103/jmp.JMP_114_16 Text en Copyright: © 2017 Journal of Medical Physics http://creativecommons.org/licenses/by-nc-sa/3.0 This is an open access article distributed under the terms of the Creative Commons Attribution-NonCommercial-ShareAlike 3.0 License, which allows others to remix, tweak, and build upon the work non-commercially, as long as the author is credited and the new creations are licensed under the identical terms.
spellingShingle Original Article
Chaudhari, Suresh H
Dobhal, Rishabh
Kinhikar, Rajesh A.
Kadam, Sudarshan S.
Deshpande, Deepak D.
Measurement of Total Scatter Factor for Stereotactic Cones with Plastic Scintillation Detector
title Measurement of Total Scatter Factor for Stereotactic Cones with Plastic Scintillation Detector
title_full Measurement of Total Scatter Factor for Stereotactic Cones with Plastic Scintillation Detector
title_fullStr Measurement of Total Scatter Factor for Stereotactic Cones with Plastic Scintillation Detector
title_full_unstemmed Measurement of Total Scatter Factor for Stereotactic Cones with Plastic Scintillation Detector
title_short Measurement of Total Scatter Factor for Stereotactic Cones with Plastic Scintillation Detector
title_sort measurement of total scatter factor for stereotactic cones with plastic scintillation detector
topic Original Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5370342/
https://www.ncbi.nlm.nih.gov/pubmed/28405102
http://dx.doi.org/10.4103/jmp.JMP_114_16
work_keys_str_mv AT chaudharisureshh measurementoftotalscatterfactorforstereotacticconeswithplasticscintillationdetector
AT dobhalrishabh measurementoftotalscatterfactorforstereotacticconeswithplasticscintillationdetector
AT kinhikarrajesha measurementoftotalscatterfactorforstereotacticconeswithplasticscintillationdetector
AT kadamsudarshans measurementoftotalscatterfactorforstereotacticconeswithplasticscintillationdetector
AT deshpandedeepakd measurementoftotalscatterfactorforstereotacticconeswithplasticscintillationdetector