Limitations and opportunities of whole blood bilirubin measurements by GEM premier 4000®

BACKGROUND: Neonatal hyperbilirubinemia has traditionally been screened by either total serum bilirubin or transcutaneous bilirubin. Whole blood bilirubin (TwB) by the GEM Premier 4000® blood gas analyzer (GEM) is a relatively new technology and it provides fast bilirubin results with a small sample...

Descripción completa

Detalles Bibliográficos
Autores principales: Wang, Li, Albert, Arianne Y. K., Jung, Benjamin, Hadad, Keyvan, Lyon, Martha E., Basso, Melanie
Formato: Online Artículo Texto
Lenguaje:English
Publicado: BioMed Central 2017
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5372304/
https://www.ncbi.nlm.nih.gov/pubmed/28356083
http://dx.doi.org/10.1186/s12887-017-0842-8
_version_ 1782518589084925952
author Wang, Li
Albert, Arianne Y. K.
Jung, Benjamin
Hadad, Keyvan
Lyon, Martha E.
Basso, Melanie
author_facet Wang, Li
Albert, Arianne Y. K.
Jung, Benjamin
Hadad, Keyvan
Lyon, Martha E.
Basso, Melanie
author_sort Wang, Li
collection PubMed
description BACKGROUND: Neonatal hyperbilirubinemia has traditionally been screened by either total serum bilirubin or transcutaneous bilirubin. Whole blood bilirubin (TwB) by the GEM Premier 4000® blood gas analyzer (GEM) is a relatively new technology and it provides fast bilirubin results with a small sample volume and can measure co-oximetry and other analytes. Our clinical study was to evaluate the reliability of TwB measured by the GEM and identify analytical and clinical factors that may contribute to possible bias. METHODS: 440 consecutive healthy newborn samples that had plasma bilirubin ordered for neonatal hyperbilirubinemia screening were included. TwB was first measured using the GEM, after which the remainder of the blood was spun and plasma neonatal bilirubin was measured using the VITROS 5600® (VITROS). RESULTS: 62 samples (14%) were excluded from analysis due to failure in obtaining GEM results. Passing-Bablok regression suggested that the GEM results were negatively biased at low concentrations of bilirubin and positively biased at higher concentrations relative to the VITROS results (y = 1.43x-61.13). Bland-Altman plots showed an overall negative bias of the GEM bilirubin with a wide range of differences compared to VITROS. Both hemoglobin concentration and hemolysis affected the accuracy of the GEM results. Clinically, male infants had higher mean bilirubin levels, and infants delivered by caesarean section had lower hemoglobin levels. When comparing the number of results below the 40th percentile and above the 95th percentile cut-offs in the Bhutani nomogram which would trigger discharge or treatment, GEM bilirubin exhibited poor sensitivity and poor specificity in contrast to VITROS bilirubin. CONCLUSIONS: An imperfect correlation was observed between whole blood bilirubin measured on the GEM4000® and plasma bilirubin on the VITROS 5600®. The contributors to the observed differences between the two instruments were specimen hemolysis and the accuracy of hemoglobin measurements, the latter of which affects the calculation of plasma-equivalent bilirubin. Additionally, the lack of standardization of total bilirubin calibration particularly in newborn specimens, may also account for some of the disagreement in results.
format Online
Article
Text
id pubmed-5372304
institution National Center for Biotechnology Information
language English
publishDate 2017
publisher BioMed Central
record_format MEDLINE/PubMed
spelling pubmed-53723042017-03-31 Limitations and opportunities of whole blood bilirubin measurements by GEM premier 4000® Wang, Li Albert, Arianne Y. K. Jung, Benjamin Hadad, Keyvan Lyon, Martha E. Basso, Melanie BMC Pediatr Research Article BACKGROUND: Neonatal hyperbilirubinemia has traditionally been screened by either total serum bilirubin or transcutaneous bilirubin. Whole blood bilirubin (TwB) by the GEM Premier 4000® blood gas analyzer (GEM) is a relatively new technology and it provides fast bilirubin results with a small sample volume and can measure co-oximetry and other analytes. Our clinical study was to evaluate the reliability of TwB measured by the GEM and identify analytical and clinical factors that may contribute to possible bias. METHODS: 440 consecutive healthy newborn samples that had plasma bilirubin ordered for neonatal hyperbilirubinemia screening were included. TwB was first measured using the GEM, after which the remainder of the blood was spun and plasma neonatal bilirubin was measured using the VITROS 5600® (VITROS). RESULTS: 62 samples (14%) were excluded from analysis due to failure in obtaining GEM results. Passing-Bablok regression suggested that the GEM results were negatively biased at low concentrations of bilirubin and positively biased at higher concentrations relative to the VITROS results (y = 1.43x-61.13). Bland-Altman plots showed an overall negative bias of the GEM bilirubin with a wide range of differences compared to VITROS. Both hemoglobin concentration and hemolysis affected the accuracy of the GEM results. Clinically, male infants had higher mean bilirubin levels, and infants delivered by caesarean section had lower hemoglobin levels. When comparing the number of results below the 40th percentile and above the 95th percentile cut-offs in the Bhutani nomogram which would trigger discharge or treatment, GEM bilirubin exhibited poor sensitivity and poor specificity in contrast to VITROS bilirubin. CONCLUSIONS: An imperfect correlation was observed between whole blood bilirubin measured on the GEM4000® and plasma bilirubin on the VITROS 5600®. The contributors to the observed differences between the two instruments were specimen hemolysis and the accuracy of hemoglobin measurements, the latter of which affects the calculation of plasma-equivalent bilirubin. Additionally, the lack of standardization of total bilirubin calibration particularly in newborn specimens, may also account for some of the disagreement in results. BioMed Central 2017-03-29 /pmc/articles/PMC5372304/ /pubmed/28356083 http://dx.doi.org/10.1186/s12887-017-0842-8 Text en © The Author(s). 2017 Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.
spellingShingle Research Article
Wang, Li
Albert, Arianne Y. K.
Jung, Benjamin
Hadad, Keyvan
Lyon, Martha E.
Basso, Melanie
Limitations and opportunities of whole blood bilirubin measurements by GEM premier 4000®
title Limitations and opportunities of whole blood bilirubin measurements by GEM premier 4000®
title_full Limitations and opportunities of whole blood bilirubin measurements by GEM premier 4000®
title_fullStr Limitations and opportunities of whole blood bilirubin measurements by GEM premier 4000®
title_full_unstemmed Limitations and opportunities of whole blood bilirubin measurements by GEM premier 4000®
title_short Limitations and opportunities of whole blood bilirubin measurements by GEM premier 4000®
title_sort limitations and opportunities of whole blood bilirubin measurements by gem premier 4000®
topic Research Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5372304/
https://www.ncbi.nlm.nih.gov/pubmed/28356083
http://dx.doi.org/10.1186/s12887-017-0842-8
work_keys_str_mv AT wangli limitationsandopportunitiesofwholebloodbilirubinmeasurementsbygempremier4000
AT albertarianneyk limitationsandopportunitiesofwholebloodbilirubinmeasurementsbygempremier4000
AT jungbenjamin limitationsandopportunitiesofwholebloodbilirubinmeasurementsbygempremier4000
AT hadadkeyvan limitationsandopportunitiesofwholebloodbilirubinmeasurementsbygempremier4000
AT lyonmarthae limitationsandopportunitiesofwholebloodbilirubinmeasurementsbygempremier4000
AT bassomelanie limitationsandopportunitiesofwholebloodbilirubinmeasurementsbygempremier4000