Two cases of inappropriate implantable cardioverter-defibrillator therapies due to T wave oversensing induced by body twisting

T wave oversensing (TWOS) is a common cause of inappropriate implantable cardioverter-defibrillator (ICD) therapies. Various algorithms to avoid inappropriate ICD therapy are available; however, they are not helpful to avoid TWOS. Although the reproduction of TWOS is useful to resolve the problem of...

Descripción completa

Detalles Bibliográficos
Autores principales: Nishii, Nobuhiro, Miyoshi, Akihito, Kubo, Motoki, Sugiyama, Hiroyasu, Tachibana, Motomi, Nakagawa, Koji, Watanabe, Atsuyuki, Morita, Hiroshi, Ito, Hiroshi
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Elsevier 2017
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5388035/
https://www.ncbi.nlm.nih.gov/pubmed/28416986
http://dx.doi.org/10.1016/j.joa.2016.06.003
_version_ 1782521057340555264
author Nishii, Nobuhiro
Miyoshi, Akihito
Kubo, Motoki
Sugiyama, Hiroyasu
Tachibana, Motomi
Nakagawa, Koji
Watanabe, Atsuyuki
Morita, Hiroshi
Ito, Hiroshi
author_facet Nishii, Nobuhiro
Miyoshi, Akihito
Kubo, Motoki
Sugiyama, Hiroyasu
Tachibana, Motomi
Nakagawa, Koji
Watanabe, Atsuyuki
Morita, Hiroshi
Ito, Hiroshi
author_sort Nishii, Nobuhiro
collection PubMed
description T wave oversensing (TWOS) is a common cause of inappropriate implantable cardioverter-defibrillator (ICD) therapies. Various algorithms to avoid inappropriate ICD therapy are available; however, they are not helpful to avoid TWOS. Although the reproduction of TWOS is useful to resolve the problem of TWOS, it is sometimes difficult to reproduce TWOS. We report two cases of inappropriate ICD therapy due to TWOS, which were induced only by body twisting. We can successfully manage the device based on the evidence of reproduced TWOS.
format Online
Article
Text
id pubmed-5388035
institution National Center for Biotechnology Information
language English
publishDate 2017
publisher Elsevier
record_format MEDLINE/PubMed
spelling pubmed-53880352017-04-17 Two cases of inappropriate implantable cardioverter-defibrillator therapies due to T wave oversensing induced by body twisting Nishii, Nobuhiro Miyoshi, Akihito Kubo, Motoki Sugiyama, Hiroyasu Tachibana, Motomi Nakagawa, Koji Watanabe, Atsuyuki Morita, Hiroshi Ito, Hiroshi J Arrhythm Rapid Firing T wave oversensing (TWOS) is a common cause of inappropriate implantable cardioverter-defibrillator (ICD) therapies. Various algorithms to avoid inappropriate ICD therapy are available; however, they are not helpful to avoid TWOS. Although the reproduction of TWOS is useful to resolve the problem of TWOS, it is sometimes difficult to reproduce TWOS. We report two cases of inappropriate ICD therapy due to TWOS, which were induced only by body twisting. We can successfully manage the device based on the evidence of reproduced TWOS. Elsevier 2017-04 2016-11-01 /pmc/articles/PMC5388035/ /pubmed/28416986 http://dx.doi.org/10.1016/j.joa.2016.06.003 Text en © 2016 Japanese Heart Rhythm Society. Published by Elsevier B.V. http://creativecommons.org/licenses/by-nc-nd/4.0/ This is an open access article under the CC BY-NC-ND license (http://creativecommons.org/licenses/by-nc-nd/4.0/).
spellingShingle Rapid Firing
Nishii, Nobuhiro
Miyoshi, Akihito
Kubo, Motoki
Sugiyama, Hiroyasu
Tachibana, Motomi
Nakagawa, Koji
Watanabe, Atsuyuki
Morita, Hiroshi
Ito, Hiroshi
Two cases of inappropriate implantable cardioverter-defibrillator therapies due to T wave oversensing induced by body twisting
title Two cases of inappropriate implantable cardioverter-defibrillator therapies due to T wave oversensing induced by body twisting
title_full Two cases of inappropriate implantable cardioverter-defibrillator therapies due to T wave oversensing induced by body twisting
title_fullStr Two cases of inappropriate implantable cardioverter-defibrillator therapies due to T wave oversensing induced by body twisting
title_full_unstemmed Two cases of inappropriate implantable cardioverter-defibrillator therapies due to T wave oversensing induced by body twisting
title_short Two cases of inappropriate implantable cardioverter-defibrillator therapies due to T wave oversensing induced by body twisting
title_sort two cases of inappropriate implantable cardioverter-defibrillator therapies due to t wave oversensing induced by body twisting
topic Rapid Firing
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5388035/
https://www.ncbi.nlm.nih.gov/pubmed/28416986
http://dx.doi.org/10.1016/j.joa.2016.06.003
work_keys_str_mv AT nishiinobuhiro twocasesofinappropriateimplantablecardioverterdefibrillatortherapiesduetotwaveoversensinginducedbybodytwisting
AT miyoshiakihito twocasesofinappropriateimplantablecardioverterdefibrillatortherapiesduetotwaveoversensinginducedbybodytwisting
AT kubomotoki twocasesofinappropriateimplantablecardioverterdefibrillatortherapiesduetotwaveoversensinginducedbybodytwisting
AT sugiyamahiroyasu twocasesofinappropriateimplantablecardioverterdefibrillatortherapiesduetotwaveoversensinginducedbybodytwisting
AT tachibanamotomi twocasesofinappropriateimplantablecardioverterdefibrillatortherapiesduetotwaveoversensinginducedbybodytwisting
AT nakagawakoji twocasesofinappropriateimplantablecardioverterdefibrillatortherapiesduetotwaveoversensinginducedbybodytwisting
AT watanabeatsuyuki twocasesofinappropriateimplantablecardioverterdefibrillatortherapiesduetotwaveoversensinginducedbybodytwisting
AT moritahiroshi twocasesofinappropriateimplantablecardioverterdefibrillatortherapiesduetotwaveoversensinginducedbybodytwisting
AT itohiroshi twocasesofinappropriateimplantablecardioverterdefibrillatortherapiesduetotwaveoversensinginducedbybodytwisting