Cargando…
Importance of exercise testing shortly after subcutaneous implantable cardioverter-defibrillator implantation in patients with Brugada syndrome – The first case of associated inappropriate shock in Japan
We report the case of a 51-year-old patient with Brugada syndrome (BrS) who experienced inappropriate shock due to T-wave oversensing (TWOS) during exercise when the optimal sensing vector was selected based on the automatic analysis by a subcutaneous implantable cardioverter-defibrillator (S-ICD)....
Autores principales: | Ishibashi, Kohei, Noda, Takashi, Kamakura, Tsukasa, Wada, Mitsuru, Inoue, Yuko, Okamura, Hideo, Kusano, Kengo |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Elsevier
2017
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5388046/ https://www.ncbi.nlm.nih.gov/pubmed/28416987 http://dx.doi.org/10.1016/j.joa.2016.07.006 |
Ejemplares similares
-
T-wave oversensing during drug challenge test after subcutaneous implantable cardioverter-defibrillator implantation in a patient with Brugada syndrome
por: Kamakura, Tsukasa, et al.
Publicado: (2016) -
Two cases of inappropriate implantable cardioverter-defibrillator therapies due to T wave oversensing induced by body twisting
por: Nishii, Nobuhiro, et al.
Publicado: (2017) -
Subcutaneous Implantable Cardioverter-Defibrillator Shock: Appropriate or Inappropriate?
por: Ip, James E.
Publicado: (2017) -
Inappropriate subcutaneous implantable cardioverter-defibrillator shocks for bradycardia
por: Chua, Kelvin C.M., et al.
Publicado: (2021) -
Intermuscular pocket for subcutaneous implantable cardioverter defibrillator: Single-center experience
por: Ferrari, Paola, et al.
Publicado: (2016)