Effects of Cone-Shaped Bend Inlet Cannulas of an Axial Blood Pump on Thrombus Formation: An Experiment and Simulation Study
BACKGROUND: Cannula shape and connection style influence the risk of thrombus formation in the blood pump by varying the blood flow characteristics inside the pump. Inlet cannulas should be designed based on the need for anatomical fit and reducing the risk of thrombus generation in the blood pump....
Autores principales: | , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
International Scientific Literature, Inc.
2017
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5389533/ https://www.ncbi.nlm.nih.gov/pubmed/28379938 http://dx.doi.org/10.12659/MSM.903421 |
_version_ | 1782521286228967424 |
---|---|
author | Liu, Guangmao Zhou, Jianye Sun, Hansong Zhang, Yan Chen, Haibo Hu, Shengshou |
author_facet | Liu, Guangmao Zhou, Jianye Sun, Hansong Zhang, Yan Chen, Haibo Hu, Shengshou |
author_sort | Liu, Guangmao |
collection | PubMed |
description | BACKGROUND: Cannula shape and connection style influence the risk of thrombus formation in the blood pump by varying the blood flow characteristics inside the pump. Inlet cannulas should be designed based on the need for anatomical fit and reducing the risk of thrombus generation in the blood pump. The effects on thrombus formation of the cone-shaped bend inlet cannulas of axial blood pumps should be studied. MATERIAL/METHODS: The cannulas were designed as cone-shaped, with 1 bent section connecting 2 straight sections. Both the silicone tube and novel cone-shaped cannula were simulated for comparison. The flow fields of a blood pump with inlet cannula were simulated by computational fluid dynamics (CFD) at flows of 2.0, 2.5, and 3.0 liters per minute (lpm), with pump rotational speeds of 7500, 8000, and 8500 rpm, respectively. Then, 6 two-dimensional (2D) particle image velocimetry (PIV) tests were conducted and the velocity distributions were analyzed. RESULTS: A low-velocity region was located inside the pump entrance when a soft silicone tube was used. At 8500 rpm and 3.0 lpm working condition, the minimum velocity inside the pump with cone-shaped cannulas was 2.5×10(−1) m/s. The cone-shaped cannulas eliminated the low-velocity region inside the pump. Both CFD and PIV results showed that the low-velocity region did not spread to the entrance of the blood pump within the flow range from 2.0 lpm to 7.0 lpm. CONCLUSIONS: The designed cone-shaped bent cannulas can eliminate the low-velocity region inside the blood pump and reduce the risk of thrombus formation in the blood pump. |
format | Online Article Text |
id | pubmed-5389533 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2017 |
publisher | International Scientific Literature, Inc. |
record_format | MEDLINE/PubMed |
spelling | pubmed-53895332017-04-17 Effects of Cone-Shaped Bend Inlet Cannulas of an Axial Blood Pump on Thrombus Formation: An Experiment and Simulation Study Liu, Guangmao Zhou, Jianye Sun, Hansong Zhang, Yan Chen, Haibo Hu, Shengshou Med Sci Monit Medical Technology BACKGROUND: Cannula shape and connection style influence the risk of thrombus formation in the blood pump by varying the blood flow characteristics inside the pump. Inlet cannulas should be designed based on the need for anatomical fit and reducing the risk of thrombus generation in the blood pump. The effects on thrombus formation of the cone-shaped bend inlet cannulas of axial blood pumps should be studied. MATERIAL/METHODS: The cannulas were designed as cone-shaped, with 1 bent section connecting 2 straight sections. Both the silicone tube and novel cone-shaped cannula were simulated for comparison. The flow fields of a blood pump with inlet cannula were simulated by computational fluid dynamics (CFD) at flows of 2.0, 2.5, and 3.0 liters per minute (lpm), with pump rotational speeds of 7500, 8000, and 8500 rpm, respectively. Then, 6 two-dimensional (2D) particle image velocimetry (PIV) tests were conducted and the velocity distributions were analyzed. RESULTS: A low-velocity region was located inside the pump entrance when a soft silicone tube was used. At 8500 rpm and 3.0 lpm working condition, the minimum velocity inside the pump with cone-shaped cannulas was 2.5×10(−1) m/s. The cone-shaped cannulas eliminated the low-velocity region inside the pump. Both CFD and PIV results showed that the low-velocity region did not spread to the entrance of the blood pump within the flow range from 2.0 lpm to 7.0 lpm. CONCLUSIONS: The designed cone-shaped bent cannulas can eliminate the low-velocity region inside the blood pump and reduce the risk of thrombus formation in the blood pump. International Scientific Literature, Inc. 2017-04-05 /pmc/articles/PMC5389533/ /pubmed/28379938 http://dx.doi.org/10.12659/MSM.903421 Text en © Med Sci Monit, 2017 This work is licensed under Creative Common Attribution-NonCommercial-NoDerivatives 4.0 International (CC BY-NC-ND 4.0) |
spellingShingle | Medical Technology Liu, Guangmao Zhou, Jianye Sun, Hansong Zhang, Yan Chen, Haibo Hu, Shengshou Effects of Cone-Shaped Bend Inlet Cannulas of an Axial Blood Pump on Thrombus Formation: An Experiment and Simulation Study |
title | Effects of Cone-Shaped Bend Inlet Cannulas of an Axial Blood Pump on Thrombus Formation: An Experiment and Simulation Study |
title_full | Effects of Cone-Shaped Bend Inlet Cannulas of an Axial Blood Pump on Thrombus Formation: An Experiment and Simulation Study |
title_fullStr | Effects of Cone-Shaped Bend Inlet Cannulas of an Axial Blood Pump on Thrombus Formation: An Experiment and Simulation Study |
title_full_unstemmed | Effects of Cone-Shaped Bend Inlet Cannulas of an Axial Blood Pump on Thrombus Formation: An Experiment and Simulation Study |
title_short | Effects of Cone-Shaped Bend Inlet Cannulas of an Axial Blood Pump on Thrombus Formation: An Experiment and Simulation Study |
title_sort | effects of cone-shaped bend inlet cannulas of an axial blood pump on thrombus formation: an experiment and simulation study |
topic | Medical Technology |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5389533/ https://www.ncbi.nlm.nih.gov/pubmed/28379938 http://dx.doi.org/10.12659/MSM.903421 |
work_keys_str_mv | AT liuguangmao effectsofconeshapedbendinletcannulasofanaxialbloodpumponthrombusformationanexperimentandsimulationstudy AT zhoujianye effectsofconeshapedbendinletcannulasofanaxialbloodpumponthrombusformationanexperimentandsimulationstudy AT sunhansong effectsofconeshapedbendinletcannulasofanaxialbloodpumponthrombusformationanexperimentandsimulationstudy AT zhangyan effectsofconeshapedbendinletcannulasofanaxialbloodpumponthrombusformationanexperimentandsimulationstudy AT chenhaibo effectsofconeshapedbendinletcannulasofanaxialbloodpumponthrombusformationanexperimentandsimulationstudy AT hushengshou effectsofconeshapedbendinletcannulasofanaxialbloodpumponthrombusformationanexperimentandsimulationstudy |