Effects of Cone-Shaped Bend Inlet Cannulas of an Axial Blood Pump on Thrombus Formation: An Experiment and Simulation Study

BACKGROUND: Cannula shape and connection style influence the risk of thrombus formation in the blood pump by varying the blood flow characteristics inside the pump. Inlet cannulas should be designed based on the need for anatomical fit and reducing the risk of thrombus generation in the blood pump....

Descripción completa

Detalles Bibliográficos
Autores principales: Liu, Guangmao, Zhou, Jianye, Sun, Hansong, Zhang, Yan, Chen, Haibo, Hu, Shengshou
Formato: Online Artículo Texto
Lenguaje:English
Publicado: International Scientific Literature, Inc. 2017
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5389533/
https://www.ncbi.nlm.nih.gov/pubmed/28379938
http://dx.doi.org/10.12659/MSM.903421
_version_ 1782521286228967424
author Liu, Guangmao
Zhou, Jianye
Sun, Hansong
Zhang, Yan
Chen, Haibo
Hu, Shengshou
author_facet Liu, Guangmao
Zhou, Jianye
Sun, Hansong
Zhang, Yan
Chen, Haibo
Hu, Shengshou
author_sort Liu, Guangmao
collection PubMed
description BACKGROUND: Cannula shape and connection style influence the risk of thrombus formation in the blood pump by varying the blood flow characteristics inside the pump. Inlet cannulas should be designed based on the need for anatomical fit and reducing the risk of thrombus generation in the blood pump. The effects on thrombus formation of the cone-shaped bend inlet cannulas of axial blood pumps should be studied. MATERIAL/METHODS: The cannulas were designed as cone-shaped, with 1 bent section connecting 2 straight sections. Both the silicone tube and novel cone-shaped cannula were simulated for comparison. The flow fields of a blood pump with inlet cannula were simulated by computational fluid dynamics (CFD) at flows of 2.0, 2.5, and 3.0 liters per minute (lpm), with pump rotational speeds of 7500, 8000, and 8500 rpm, respectively. Then, 6 two-dimensional (2D) particle image velocimetry (PIV) tests were conducted and the velocity distributions were analyzed. RESULTS: A low-velocity region was located inside the pump entrance when a soft silicone tube was used. At 8500 rpm and 3.0 lpm working condition, the minimum velocity inside the pump with cone-shaped cannulas was 2.5×10(−1) m/s. The cone-shaped cannulas eliminated the low-velocity region inside the pump. Both CFD and PIV results showed that the low-velocity region did not spread to the entrance of the blood pump within the flow range from 2.0 lpm to 7.0 lpm. CONCLUSIONS: The designed cone-shaped bent cannulas can eliminate the low-velocity region inside the blood pump and reduce the risk of thrombus formation in the blood pump.
format Online
Article
Text
id pubmed-5389533
institution National Center for Biotechnology Information
language English
publishDate 2017
publisher International Scientific Literature, Inc.
record_format MEDLINE/PubMed
spelling pubmed-53895332017-04-17 Effects of Cone-Shaped Bend Inlet Cannulas of an Axial Blood Pump on Thrombus Formation: An Experiment and Simulation Study Liu, Guangmao Zhou, Jianye Sun, Hansong Zhang, Yan Chen, Haibo Hu, Shengshou Med Sci Monit Medical Technology BACKGROUND: Cannula shape and connection style influence the risk of thrombus formation in the blood pump by varying the blood flow characteristics inside the pump. Inlet cannulas should be designed based on the need for anatomical fit and reducing the risk of thrombus generation in the blood pump. The effects on thrombus formation of the cone-shaped bend inlet cannulas of axial blood pumps should be studied. MATERIAL/METHODS: The cannulas were designed as cone-shaped, with 1 bent section connecting 2 straight sections. Both the silicone tube and novel cone-shaped cannula were simulated for comparison. The flow fields of a blood pump with inlet cannula were simulated by computational fluid dynamics (CFD) at flows of 2.0, 2.5, and 3.0 liters per minute (lpm), with pump rotational speeds of 7500, 8000, and 8500 rpm, respectively. Then, 6 two-dimensional (2D) particle image velocimetry (PIV) tests were conducted and the velocity distributions were analyzed. RESULTS: A low-velocity region was located inside the pump entrance when a soft silicone tube was used. At 8500 rpm and 3.0 lpm working condition, the minimum velocity inside the pump with cone-shaped cannulas was 2.5×10(−1) m/s. The cone-shaped cannulas eliminated the low-velocity region inside the pump. Both CFD and PIV results showed that the low-velocity region did not spread to the entrance of the blood pump within the flow range from 2.0 lpm to 7.0 lpm. CONCLUSIONS: The designed cone-shaped bent cannulas can eliminate the low-velocity region inside the blood pump and reduce the risk of thrombus formation in the blood pump. International Scientific Literature, Inc. 2017-04-05 /pmc/articles/PMC5389533/ /pubmed/28379938 http://dx.doi.org/10.12659/MSM.903421 Text en © Med Sci Monit, 2017 This work is licensed under Creative Common Attribution-NonCommercial-NoDerivatives 4.0 International (CC BY-NC-ND 4.0)
spellingShingle Medical Technology
Liu, Guangmao
Zhou, Jianye
Sun, Hansong
Zhang, Yan
Chen, Haibo
Hu, Shengshou
Effects of Cone-Shaped Bend Inlet Cannulas of an Axial Blood Pump on Thrombus Formation: An Experiment and Simulation Study
title Effects of Cone-Shaped Bend Inlet Cannulas of an Axial Blood Pump on Thrombus Formation: An Experiment and Simulation Study
title_full Effects of Cone-Shaped Bend Inlet Cannulas of an Axial Blood Pump on Thrombus Formation: An Experiment and Simulation Study
title_fullStr Effects of Cone-Shaped Bend Inlet Cannulas of an Axial Blood Pump on Thrombus Formation: An Experiment and Simulation Study
title_full_unstemmed Effects of Cone-Shaped Bend Inlet Cannulas of an Axial Blood Pump on Thrombus Formation: An Experiment and Simulation Study
title_short Effects of Cone-Shaped Bend Inlet Cannulas of an Axial Blood Pump on Thrombus Formation: An Experiment and Simulation Study
title_sort effects of cone-shaped bend inlet cannulas of an axial blood pump on thrombus formation: an experiment and simulation study
topic Medical Technology
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5389533/
https://www.ncbi.nlm.nih.gov/pubmed/28379938
http://dx.doi.org/10.12659/MSM.903421
work_keys_str_mv AT liuguangmao effectsofconeshapedbendinletcannulasofanaxialbloodpumponthrombusformationanexperimentandsimulationstudy
AT zhoujianye effectsofconeshapedbendinletcannulasofanaxialbloodpumponthrombusformationanexperimentandsimulationstudy
AT sunhansong effectsofconeshapedbendinletcannulasofanaxialbloodpumponthrombusformationanexperimentandsimulationstudy
AT zhangyan effectsofconeshapedbendinletcannulasofanaxialbloodpumponthrombusformationanexperimentandsimulationstudy
AT chenhaibo effectsofconeshapedbendinletcannulasofanaxialbloodpumponthrombusformationanexperimentandsimulationstudy
AT hushengshou effectsofconeshapedbendinletcannulasofanaxialbloodpumponthrombusformationanexperimentandsimulationstudy