Physical human-robot interaction of an active pelvis orthosis: toward ergonomic assessment of wearable robots

BACKGROUND: In human-centered robotics, exoskeletons are becoming relevant for addressing needs in the healthcare and industrial domains. Owing to their close interaction with the user, the safety and ergonomics of these systems are critical design features that require systematic evaluation methodo...

Descripción completa

Detalles Bibliográficos
Autores principales: d’Elia, Nicolò, Vanetti, Federica, Cempini, Marco, Pasquini, Guido, Parri, Andrea, Rabuffetti, Marco, Ferrarin, Maurizio, Molino Lova, Raffaele, Vitiello, Nicola
Formato: Online Artículo Texto
Lenguaje:English
Publicado: BioMed Central 2017
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5391543/
https://www.ncbi.nlm.nih.gov/pubmed/28410594
http://dx.doi.org/10.1186/s12984-017-0237-y
_version_ 1783229292051693568
author d’Elia, Nicolò
Vanetti, Federica
Cempini, Marco
Pasquini, Guido
Parri, Andrea
Rabuffetti, Marco
Ferrarin, Maurizio
Molino Lova, Raffaele
Vitiello, Nicola
author_facet d’Elia, Nicolò
Vanetti, Federica
Cempini, Marco
Pasquini, Guido
Parri, Andrea
Rabuffetti, Marco
Ferrarin, Maurizio
Molino Lova, Raffaele
Vitiello, Nicola
author_sort d’Elia, Nicolò
collection PubMed
description BACKGROUND: In human-centered robotics, exoskeletons are becoming relevant for addressing needs in the healthcare and industrial domains. Owing to their close interaction with the user, the safety and ergonomics of these systems are critical design features that require systematic evaluation methodologies. Proper transfer of mechanical power requires optimal tuning of the kinematic coupling between the robotic and anatomical joint rotation axes. We present the methods and results of an experimental evaluation of the physical interaction with an active pelvis orthosis (APO). This device was designed to effectively assist in hip flexion-extension during locomotion with a minimum impact on the physiological human kinematics, owing to a set of passive degrees of freedom for self-alignment of the human and robotic hip flexion-extension axes. METHODS: Five healthy volunteers walked on a treadmill at different speeds without and with the APO under different levels of assistance. The user-APO physical interaction was evaluated in terms of: (i) the deviation of human lower-limb joint kinematics when wearing the APO with respect to the physiological behavior (i.e., without the APO); (ii) relative displacements between the APO orthotic shells and the corresponding body segments; and (iii) the discrepancy between the kinematics of the APO and the wearer’s hip joints. RESULTS: The results show: (i) negligible interference of the APO in human kinematics under all the experimented conditions; (ii) small (i.e., < 1 cm) relative displacements between the APO cuffs and the corresponding body segments (called stability); and (iii) significant increment in the human-robot kinematics discrepancy at the hip flexion-extension joint associated with speed and assistance level increase. CONCLUSIONS: APO mechanics and actuation have negligible interference in human locomotion. Human kinematics was not affected by the APO under all tested conditions. In addition, under all tested conditions, there was no relevant relative displacement between the orthotic cuffs and the corresponding anatomical segments. Hence, the physical human-robot coupling is reliable. These facts prove that the adopted mechanical design of passive degrees of freedom allows an effective human-robot kinematic coupling. We believe that this analysis may be useful for the definition of evaluation metrics for the ergonomics assessment of wearable robots.
format Online
Article
Text
id pubmed-5391543
institution National Center for Biotechnology Information
language English
publishDate 2017
publisher BioMed Central
record_format MEDLINE/PubMed
spelling pubmed-53915432017-04-14 Physical human-robot interaction of an active pelvis orthosis: toward ergonomic assessment of wearable robots d’Elia, Nicolò Vanetti, Federica Cempini, Marco Pasquini, Guido Parri, Andrea Rabuffetti, Marco Ferrarin, Maurizio Molino Lova, Raffaele Vitiello, Nicola J Neuroeng Rehabil Research BACKGROUND: In human-centered robotics, exoskeletons are becoming relevant for addressing needs in the healthcare and industrial domains. Owing to their close interaction with the user, the safety and ergonomics of these systems are critical design features that require systematic evaluation methodologies. Proper transfer of mechanical power requires optimal tuning of the kinematic coupling between the robotic and anatomical joint rotation axes. We present the methods and results of an experimental evaluation of the physical interaction with an active pelvis orthosis (APO). This device was designed to effectively assist in hip flexion-extension during locomotion with a minimum impact on the physiological human kinematics, owing to a set of passive degrees of freedom for self-alignment of the human and robotic hip flexion-extension axes. METHODS: Five healthy volunteers walked on a treadmill at different speeds without and with the APO under different levels of assistance. The user-APO physical interaction was evaluated in terms of: (i) the deviation of human lower-limb joint kinematics when wearing the APO with respect to the physiological behavior (i.e., without the APO); (ii) relative displacements between the APO orthotic shells and the corresponding body segments; and (iii) the discrepancy between the kinematics of the APO and the wearer’s hip joints. RESULTS: The results show: (i) negligible interference of the APO in human kinematics under all the experimented conditions; (ii) small (i.e., < 1 cm) relative displacements between the APO cuffs and the corresponding body segments (called stability); and (iii) significant increment in the human-robot kinematics discrepancy at the hip flexion-extension joint associated with speed and assistance level increase. CONCLUSIONS: APO mechanics and actuation have negligible interference in human locomotion. Human kinematics was not affected by the APO under all tested conditions. In addition, under all tested conditions, there was no relevant relative displacement between the orthotic cuffs and the corresponding anatomical segments. Hence, the physical human-robot coupling is reliable. These facts prove that the adopted mechanical design of passive degrees of freedom allows an effective human-robot kinematic coupling. We believe that this analysis may be useful for the definition of evaluation metrics for the ergonomics assessment of wearable robots. BioMed Central 2017-04-14 /pmc/articles/PMC5391543/ /pubmed/28410594 http://dx.doi.org/10.1186/s12984-017-0237-y Text en © The Author(s). 2017 Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.
spellingShingle Research
d’Elia, Nicolò
Vanetti, Federica
Cempini, Marco
Pasquini, Guido
Parri, Andrea
Rabuffetti, Marco
Ferrarin, Maurizio
Molino Lova, Raffaele
Vitiello, Nicola
Physical human-robot interaction of an active pelvis orthosis: toward ergonomic assessment of wearable robots
title Physical human-robot interaction of an active pelvis orthosis: toward ergonomic assessment of wearable robots
title_full Physical human-robot interaction of an active pelvis orthosis: toward ergonomic assessment of wearable robots
title_fullStr Physical human-robot interaction of an active pelvis orthosis: toward ergonomic assessment of wearable robots
title_full_unstemmed Physical human-robot interaction of an active pelvis orthosis: toward ergonomic assessment of wearable robots
title_short Physical human-robot interaction of an active pelvis orthosis: toward ergonomic assessment of wearable robots
title_sort physical human-robot interaction of an active pelvis orthosis: toward ergonomic assessment of wearable robots
topic Research
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5391543/
https://www.ncbi.nlm.nih.gov/pubmed/28410594
http://dx.doi.org/10.1186/s12984-017-0237-y
work_keys_str_mv AT delianicolo physicalhumanrobotinteractionofanactivepelvisorthosistowardergonomicassessmentofwearablerobots
AT vanettifederica physicalhumanrobotinteractionofanactivepelvisorthosistowardergonomicassessmentofwearablerobots
AT cempinimarco physicalhumanrobotinteractionofanactivepelvisorthosistowardergonomicassessmentofwearablerobots
AT pasquiniguido physicalhumanrobotinteractionofanactivepelvisorthosistowardergonomicassessmentofwearablerobots
AT parriandrea physicalhumanrobotinteractionofanactivepelvisorthosistowardergonomicassessmentofwearablerobots
AT rabuffettimarco physicalhumanrobotinteractionofanactivepelvisorthosistowardergonomicassessmentofwearablerobots
AT ferrarinmaurizio physicalhumanrobotinteractionofanactivepelvisorthosistowardergonomicassessmentofwearablerobots
AT molinolovaraffaele physicalhumanrobotinteractionofanactivepelvisorthosistowardergonomicassessmentofwearablerobots
AT vitiellonicola physicalhumanrobotinteractionofanactivepelvisorthosistowardergonomicassessmentofwearablerobots