Comparison of the biometric measurements calculated with Zeiss IOL-Master and WaveLight OB-820

BACKGROUND: The primary objective of this study was to investigate the level of agreement between IOL-Master and OB-820 ocular biometers. MATERIALS AND METHODS: In this prospective randomized case series, we measured the anterior chamber depth (ACD), the axial length (AL), the corneal radii (R1, R2)...

Descripción completa

Detalles Bibliográficos
Autores principales: Labiris, Georgios, Ntonti, Panagiota, Ntitsos, Konstantinos, Katsanos, Andreas, Sideroudi, Haris, Kozobolis, Vasilios
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Dove Medical Press 2017
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5408945/
https://www.ncbi.nlm.nih.gov/pubmed/28461737
http://dx.doi.org/10.2147/OPTH.S133207
_version_ 1783232387891593216
author Labiris, Georgios
Ntonti, Panagiota
Ntitsos, Konstantinos
Katsanos, Andreas
Sideroudi, Haris
Kozobolis, Vasilios
author_facet Labiris, Georgios
Ntonti, Panagiota
Ntitsos, Konstantinos
Katsanos, Andreas
Sideroudi, Haris
Kozobolis, Vasilios
author_sort Labiris, Georgios
collection PubMed
description BACKGROUND: The primary objective of this study was to investigate the level of agreement between IOL-Master and OB-820 ocular biometers. MATERIALS AND METHODS: In this prospective randomized case series, we measured the anterior chamber depth (ACD), the axial length (AL), the corneal radii (R1, R2), the ratio of mean corneal radius and spherical equivalent and the corneal astigmatism (cylinder [Cyl]) before and after cataract extraction surgery. RESULTS: Significant differences between pre- and postoperative data were observed for ACD for both biometric devices (P<0.01) and Cyl parameter in IOL-Master. Range and 95% limits of agreement (LoA) were clinically significantly different for AL parameter pre- and postoperatively and for R2 and radius and spherical equivalent postoperatively (P<0.001). The rest of the parameters presented sufficient 95% LoA, which imply good agreement. CONCLUSION: In clinical practice, the IOL-master and OB-820 should not be used interchangeably due to discrepancy in the important AL parameter. Both biometers may provide consistent results regarding Cyl, ACD and R1.
format Online
Article
Text
id pubmed-5408945
institution National Center for Biotechnology Information
language English
publishDate 2017
publisher Dove Medical Press
record_format MEDLINE/PubMed
spelling pubmed-54089452017-05-01 Comparison of the biometric measurements calculated with Zeiss IOL-Master and WaveLight OB-820 Labiris, Georgios Ntonti, Panagiota Ntitsos, Konstantinos Katsanos, Andreas Sideroudi, Haris Kozobolis, Vasilios Clin Ophthalmol Original Research BACKGROUND: The primary objective of this study was to investigate the level of agreement between IOL-Master and OB-820 ocular biometers. MATERIALS AND METHODS: In this prospective randomized case series, we measured the anterior chamber depth (ACD), the axial length (AL), the corneal radii (R1, R2), the ratio of mean corneal radius and spherical equivalent and the corneal astigmatism (cylinder [Cyl]) before and after cataract extraction surgery. RESULTS: Significant differences between pre- and postoperative data were observed for ACD for both biometric devices (P<0.01) and Cyl parameter in IOL-Master. Range and 95% limits of agreement (LoA) were clinically significantly different for AL parameter pre- and postoperatively and for R2 and radius and spherical equivalent postoperatively (P<0.001). The rest of the parameters presented sufficient 95% LoA, which imply good agreement. CONCLUSION: In clinical practice, the IOL-master and OB-820 should not be used interchangeably due to discrepancy in the important AL parameter. Both biometers may provide consistent results regarding Cyl, ACD and R1. Dove Medical Press 2017-04-21 /pmc/articles/PMC5408945/ /pubmed/28461737 http://dx.doi.org/10.2147/OPTH.S133207 Text en © 2017 Labiris et al. This work is published and licensed by Dove Medical Press Limited The full terms of this license are available at https://www.dovepress.com/terms.php and incorporate the Creative Commons Attribution – Non Commercial (unported, v3.0) License (http://creativecommons.org/licenses/by-nc/3.0/). By accessing the work you hereby accept the Terms. Non-commercial uses of the work are permitted without any further permission from Dove Medical Press Limited, provided the work is properly attributed.
spellingShingle Original Research
Labiris, Georgios
Ntonti, Panagiota
Ntitsos, Konstantinos
Katsanos, Andreas
Sideroudi, Haris
Kozobolis, Vasilios
Comparison of the biometric measurements calculated with Zeiss IOL-Master and WaveLight OB-820
title Comparison of the biometric measurements calculated with Zeiss IOL-Master and WaveLight OB-820
title_full Comparison of the biometric measurements calculated with Zeiss IOL-Master and WaveLight OB-820
title_fullStr Comparison of the biometric measurements calculated with Zeiss IOL-Master and WaveLight OB-820
title_full_unstemmed Comparison of the biometric measurements calculated with Zeiss IOL-Master and WaveLight OB-820
title_short Comparison of the biometric measurements calculated with Zeiss IOL-Master and WaveLight OB-820
title_sort comparison of the biometric measurements calculated with zeiss iol-master and wavelight ob-820
topic Original Research
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5408945/
https://www.ncbi.nlm.nih.gov/pubmed/28461737
http://dx.doi.org/10.2147/OPTH.S133207
work_keys_str_mv AT labirisgeorgios comparisonofthebiometricmeasurementscalculatedwithzeissiolmasterandwavelightob820
AT ntontipanagiota comparisonofthebiometricmeasurementscalculatedwithzeissiolmasterandwavelightob820
AT ntitsoskonstantinos comparisonofthebiometricmeasurementscalculatedwithzeissiolmasterandwavelightob820
AT katsanosandreas comparisonofthebiometricmeasurementscalculatedwithzeissiolmasterandwavelightob820
AT sideroudiharis comparisonofthebiometricmeasurementscalculatedwithzeissiolmasterandwavelightob820
AT kozobolisvasilios comparisonofthebiometricmeasurementscalculatedwithzeissiolmasterandwavelightob820