Cargando…

Accuracy Evaluation of a Stereolithographic Surgical Template for Dental Implant Insertion Using 3D Superimposition Protocol

THE AIM: of this study was to evaluate the accuracy of a stereolithographic template, with sleeve structure incorporated into the design, for computer-guided dental implant insertion in partially edentulous patients. MATERIALS AND METHODS: Sixty-five implants were placed in twenty-five consecutive p...

Descripción completa

Detalles Bibliográficos
Autores principales: Cristache, Corina Marilena, Gurbanescu, Silviu
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Hindawi 2017
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5438863/
https://www.ncbi.nlm.nih.gov/pubmed/28555157
http://dx.doi.org/10.1155/2017/4292081
Descripción
Sumario:THE AIM: of this study was to evaluate the accuracy of a stereolithographic template, with sleeve structure incorporated into the design, for computer-guided dental implant insertion in partially edentulous patients. MATERIALS AND METHODS: Sixty-five implants were placed in twenty-five consecutive patients with a stereolithographic surgical template. After surgery, digital impression was taken and 3D inaccuracy of implants position at entry point, apex, and angle deviation was measured using an inspection tool software. Mann–Whitney U test was used to compare accuracy between maxillary and mandibular surgical guides. A p value < .05 was considered significant. RESULTS: Mean (and standard deviation) of 3D error at the entry point was 0.798 mm (±0.52), at the implant apex it was 1.17 mm (±0.63), and mean angular deviation was 2.34 (±0.85). A statistically significant reduced 3D error was observed at entry point p = .037, at implant apex p = .008, and also in angular deviation p = .030 in mandible when comparing to maxilla. CONCLUSIONS: The surgical template used has proved high accuracy for implant insertion. Within the limitations of the present study, the protocol for comparing a digital file (treatment plan) with postinsertion digital impression may be considered a useful procedure for assessing surgical template accuracy, avoiding radiation exposure, during postoperative CBCT scanning.