Instant velocity and consistency of emitted cloud change by the different levels of canister filling with Metered Dose Inhalers (MDIs), but not with Soft Mist Inhalers (SMIs): a bench study
BACKGROUND: Inhalation is the preferred route for respiratory drug delivery, but several factors contribute to the variability of the respirable dose fraction. Instant velocity and the dynamic characteristics of the droplet cloud represent crucial factors. Aim was to measure and compare the instant...
Autores principales: | , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
BioMed Central
2017
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5447303/ https://www.ncbi.nlm.nih.gov/pubmed/28560037 http://dx.doi.org/10.1186/s40248-017-0096-1 |
_version_ | 1783239307306205184 |
---|---|
author | Dal Negro, Roberto W. Longo, Pietro Ziani, Orestepaolo Villanis Bonadiman, Luca Turco, Paola |
author_facet | Dal Negro, Roberto W. Longo, Pietro Ziani, Orestepaolo Villanis Bonadiman, Luca Turco, Paola |
author_sort | Dal Negro, Roberto W. |
collection | PubMed |
description | BACKGROUND: Inhalation is the preferred route for respiratory drug delivery, but several factors contribute to the variability of the respirable dose fraction. Instant velocity and the dynamic characteristics of the droplet cloud represent crucial factors. Aim was to measure and compare the instant velocity and the consistency of emitted cloud from five different MDIs (A - Salbutamol sulphate 100mcg, GSK; B - Salbutamol sulphate 100mcg, Valeas; C - Salmeterol xinafoate/Fluticasone propionate 25/125mcg, GSK; D - Formoterol fumarate/Bechlomethasone propionate 6/100mcg, Chiesi; E - Formoterol fumarate/Fluticasone dipropionate 5/125mcg, Mundipharma) and one SMI (Tiotropium bromide 5mcg, Boehringer Ingelheim), at different distance from the nozzle and canister filling. METHODS: Measurements were made at 90, 50, and 10% of canister filling, and at 5, 10, and 20 cm from the nozzle, for a total of 972 puffs. A high speed video photography protocol was adopted and high speed cameras (1.200 frames/sec.) were used. Data were acquired by means of specialized softwares. Temperature, humidity, and vibrations occurrence were strictly controlled during measurements. Statistics: Anova and p < 0.05 were accepted as the minimum significance level. RESULTS: MDIs generated different Instant velocities: MDI B generated the highest, while MDI A the lowest. As expected, velocity decreased in proportion to the distance from the nozzle. Except with MDI C, instant velocity decreased significantly over the first 50% of canister emptying, but dropped by >33% at 90% of emptying with all other MDIs (p < 0–037; p < 0.001; p < 0.005, and p < 0.001, respectively). Instant velocity was extremely lower (p < 0.001) and constant for all levels of canister filling (p = ns) with SMI. All MDIs had a very fast jet phase, ranging 0.01–0.03 s at 10 cm, and 0.03–0.05 s at 20 cm from the nozzle, without any significant difference from each other (p = ns). MDIs generated a cloud similarly tight (p = ns) at 10 and 20 cm from the nozzle, while it was extremely wider and constant with the SMI (p = 0–001). Also the cloud turbulence was minimized during the SMI emission. DISCUSSION AND CONCLUSIONS: MDIs are characterized by a substantial variability in both their instant velocity and consistency of the emitted cloud at different levels of canister filling. SMI generates a much slower soft mist cloud which is constantly homogeneous and independent of canister emptying. These peculiarities assessed at bench are suggesting a higher dose consistency and a much more effective therapeutic performance also in real life. |
format | Online Article Text |
id | pubmed-5447303 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2017 |
publisher | BioMed Central |
record_format | MEDLINE/PubMed |
spelling | pubmed-54473032017-05-30 Instant velocity and consistency of emitted cloud change by the different levels of canister filling with Metered Dose Inhalers (MDIs), but not with Soft Mist Inhalers (SMIs): a bench study Dal Negro, Roberto W. Longo, Pietro Ziani, Orestepaolo Villanis Bonadiman, Luca Turco, Paola Multidiscip Respir Med Original Research Article BACKGROUND: Inhalation is the preferred route for respiratory drug delivery, but several factors contribute to the variability of the respirable dose fraction. Instant velocity and the dynamic characteristics of the droplet cloud represent crucial factors. Aim was to measure and compare the instant velocity and the consistency of emitted cloud from five different MDIs (A - Salbutamol sulphate 100mcg, GSK; B - Salbutamol sulphate 100mcg, Valeas; C - Salmeterol xinafoate/Fluticasone propionate 25/125mcg, GSK; D - Formoterol fumarate/Bechlomethasone propionate 6/100mcg, Chiesi; E - Formoterol fumarate/Fluticasone dipropionate 5/125mcg, Mundipharma) and one SMI (Tiotropium bromide 5mcg, Boehringer Ingelheim), at different distance from the nozzle and canister filling. METHODS: Measurements were made at 90, 50, and 10% of canister filling, and at 5, 10, and 20 cm from the nozzle, for a total of 972 puffs. A high speed video photography protocol was adopted and high speed cameras (1.200 frames/sec.) were used. Data were acquired by means of specialized softwares. Temperature, humidity, and vibrations occurrence were strictly controlled during measurements. Statistics: Anova and p < 0.05 were accepted as the minimum significance level. RESULTS: MDIs generated different Instant velocities: MDI B generated the highest, while MDI A the lowest. As expected, velocity decreased in proportion to the distance from the nozzle. Except with MDI C, instant velocity decreased significantly over the first 50% of canister emptying, but dropped by >33% at 90% of emptying with all other MDIs (p < 0–037; p < 0.001; p < 0.005, and p < 0.001, respectively). Instant velocity was extremely lower (p < 0.001) and constant for all levels of canister filling (p = ns) with SMI. All MDIs had a very fast jet phase, ranging 0.01–0.03 s at 10 cm, and 0.03–0.05 s at 20 cm from the nozzle, without any significant difference from each other (p = ns). MDIs generated a cloud similarly tight (p = ns) at 10 and 20 cm from the nozzle, while it was extremely wider and constant with the SMI (p = 0–001). Also the cloud turbulence was minimized during the SMI emission. DISCUSSION AND CONCLUSIONS: MDIs are characterized by a substantial variability in both their instant velocity and consistency of the emitted cloud at different levels of canister filling. SMI generates a much slower soft mist cloud which is constantly homogeneous and independent of canister emptying. These peculiarities assessed at bench are suggesting a higher dose consistency and a much more effective therapeutic performance also in real life. BioMed Central 2017-05-29 /pmc/articles/PMC5447303/ /pubmed/28560037 http://dx.doi.org/10.1186/s40248-017-0096-1 Text en © The Author(s). 2017 Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated. |
spellingShingle | Original Research Article Dal Negro, Roberto W. Longo, Pietro Ziani, Orestepaolo Villanis Bonadiman, Luca Turco, Paola Instant velocity and consistency of emitted cloud change by the different levels of canister filling with Metered Dose Inhalers (MDIs), but not with Soft Mist Inhalers (SMIs): a bench study |
title | Instant velocity and consistency of emitted cloud change by the different levels of canister filling with Metered Dose Inhalers (MDIs), but not with Soft Mist Inhalers (SMIs): a bench study |
title_full | Instant velocity and consistency of emitted cloud change by the different levels of canister filling with Metered Dose Inhalers (MDIs), but not with Soft Mist Inhalers (SMIs): a bench study |
title_fullStr | Instant velocity and consistency of emitted cloud change by the different levels of canister filling with Metered Dose Inhalers (MDIs), but not with Soft Mist Inhalers (SMIs): a bench study |
title_full_unstemmed | Instant velocity and consistency of emitted cloud change by the different levels of canister filling with Metered Dose Inhalers (MDIs), but not with Soft Mist Inhalers (SMIs): a bench study |
title_short | Instant velocity and consistency of emitted cloud change by the different levels of canister filling with Metered Dose Inhalers (MDIs), but not with Soft Mist Inhalers (SMIs): a bench study |
title_sort | instant velocity and consistency of emitted cloud change by the different levels of canister filling with metered dose inhalers (mdis), but not with soft mist inhalers (smis): a bench study |
topic | Original Research Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5447303/ https://www.ncbi.nlm.nih.gov/pubmed/28560037 http://dx.doi.org/10.1186/s40248-017-0096-1 |
work_keys_str_mv | AT dalnegrorobertow instantvelocityandconsistencyofemittedcloudchangebythedifferentlevelsofcanisterfillingwithmetereddoseinhalersmdisbutnotwithsoftmistinhalerssmisabenchstudy AT longopietro instantvelocityandconsistencyofemittedcloudchangebythedifferentlevelsofcanisterfillingwithmetereddoseinhalersmdisbutnotwithsoftmistinhalerssmisabenchstudy AT zianiorestepaolovillanis instantvelocityandconsistencyofemittedcloudchangebythedifferentlevelsofcanisterfillingwithmetereddoseinhalersmdisbutnotwithsoftmistinhalerssmisabenchstudy AT bonadimanluca instantvelocityandconsistencyofemittedcloudchangebythedifferentlevelsofcanisterfillingwithmetereddoseinhalersmdisbutnotwithsoftmistinhalerssmisabenchstudy AT turcopaola instantvelocityandconsistencyofemittedcloudchangebythedifferentlevelsofcanisterfillingwithmetereddoseinhalersmdisbutnotwithsoftmistinhalerssmisabenchstudy |