Dose-Decreasing Effect of the First Reversed Laser Beam Collimator for C-Arm Type Angiographic Equipment

This is a study on the dose-decreasing effect of the first reversed laser beam collimator (RLBC) for C-arm type angiographic equipment. A laser beam was located at the center of each plane at an oblique angle to the angiographic equipment detector. A field of view, which could be seen with the naked...

Descripción completa

Detalles Bibliográficos
Autores principales: Heo, Yeong-Cheol, Cho, Jae-Hwan, Han, Dong-Kyoon
Formato: Online Artículo Texto
Lenguaje:English
Publicado: The Korean Academy of Medical Sciences 2017
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5461310/
https://www.ncbi.nlm.nih.gov/pubmed/28581263
http://dx.doi.org/10.3346/jkms.2017.32.7.1083
_version_ 1783242313613443072
author Heo, Yeong-Cheol
Cho, Jae-Hwan
Han, Dong-Kyoon
author_facet Heo, Yeong-Cheol
Cho, Jae-Hwan
Han, Dong-Kyoon
author_sort Heo, Yeong-Cheol
collection PubMed
description This is a study on the dose-decreasing effect of the first reversed laser beam collimator (RLBC) for C-arm type angiographic equipment. A laser beam was located at the center of each plane at an oblique angle to the angiographic equipment detector. A field of view, which could be seen with the naked eye, was made by focusing the laser beam in the direction of the X-ray source. The height of the table was fixed at 75 cm and the iron balls were located within 2 mm of the top, bottom, left, and right edges of the output image. The time needed for location fixing, fluoroscopy, and measurement of dose area product (DAP) were compared by having 30 radiologists perform location fixing by looking at the fluoroscopic image while performing location fixing (no radiation) and while the RLBC was turned on. In the next test, the time needed for location fixing, fluoroscopy, and DAP were compared when varying the location of the iron balls from 2 to 10 mm from the edges of the output image. The results showed that the time needed for location fixing, the time needed for fluoroscopy, and DAP decreased, both in the first test and the second test. This study confirmed that the use of a RLBC for C-arm type angiographic equipment decreases both the time needed to perform the procedure and the radiation dose received. It is expected that continuous advancement of RLBC technology will contribute greatly to decreasing the dose of radiation needed and improving convenience during angiography.
format Online
Article
Text
id pubmed-5461310
institution National Center for Biotechnology Information
language English
publishDate 2017
publisher The Korean Academy of Medical Sciences
record_format MEDLINE/PubMed
spelling pubmed-54613102017-07-01 Dose-Decreasing Effect of the First Reversed Laser Beam Collimator for C-Arm Type Angiographic Equipment Heo, Yeong-Cheol Cho, Jae-Hwan Han, Dong-Kyoon J Korean Med Sci Original Article This is a study on the dose-decreasing effect of the first reversed laser beam collimator (RLBC) for C-arm type angiographic equipment. A laser beam was located at the center of each plane at an oblique angle to the angiographic equipment detector. A field of view, which could be seen with the naked eye, was made by focusing the laser beam in the direction of the X-ray source. The height of the table was fixed at 75 cm and the iron balls were located within 2 mm of the top, bottom, left, and right edges of the output image. The time needed for location fixing, fluoroscopy, and measurement of dose area product (DAP) were compared by having 30 radiologists perform location fixing by looking at the fluoroscopic image while performing location fixing (no radiation) and while the RLBC was turned on. In the next test, the time needed for location fixing, fluoroscopy, and DAP were compared when varying the location of the iron balls from 2 to 10 mm from the edges of the output image. The results showed that the time needed for location fixing, the time needed for fluoroscopy, and DAP decreased, both in the first test and the second test. This study confirmed that the use of a RLBC for C-arm type angiographic equipment decreases both the time needed to perform the procedure and the radiation dose received. It is expected that continuous advancement of RLBC technology will contribute greatly to decreasing the dose of radiation needed and improving convenience during angiography. The Korean Academy of Medical Sciences 2017-07 2017-05-26 /pmc/articles/PMC5461310/ /pubmed/28581263 http://dx.doi.org/10.3346/jkms.2017.32.7.1083 Text en © 2017 The Korean Academy of Medical Sciences. https://creativecommons.org/licenses/by-nc/4.0/ This is an Open Access article distributed under the terms of the Creative Commons Attribution Non-Commercial License (https://creativecommons.org/licenses/by-nc/4.0/) which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Original Article
Heo, Yeong-Cheol
Cho, Jae-Hwan
Han, Dong-Kyoon
Dose-Decreasing Effect of the First Reversed Laser Beam Collimator for C-Arm Type Angiographic Equipment
title Dose-Decreasing Effect of the First Reversed Laser Beam Collimator for C-Arm Type Angiographic Equipment
title_full Dose-Decreasing Effect of the First Reversed Laser Beam Collimator for C-Arm Type Angiographic Equipment
title_fullStr Dose-Decreasing Effect of the First Reversed Laser Beam Collimator for C-Arm Type Angiographic Equipment
title_full_unstemmed Dose-Decreasing Effect of the First Reversed Laser Beam Collimator for C-Arm Type Angiographic Equipment
title_short Dose-Decreasing Effect of the First Reversed Laser Beam Collimator for C-Arm Type Angiographic Equipment
title_sort dose-decreasing effect of the first reversed laser beam collimator for c-arm type angiographic equipment
topic Original Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5461310/
https://www.ncbi.nlm.nih.gov/pubmed/28581263
http://dx.doi.org/10.3346/jkms.2017.32.7.1083
work_keys_str_mv AT heoyeongcheol dosedecreasingeffectofthefirstreversedlaserbeamcollimatorforcarmtypeangiographicequipment
AT chojaehwan dosedecreasingeffectofthefirstreversedlaserbeamcollimatorforcarmtypeangiographicequipment
AT handongkyoon dosedecreasingeffectofthefirstreversedlaserbeamcollimatorforcarmtypeangiographicequipment