Functionality and acceptability of a wireless fetal heart rate monitoring device in term pregnant women in rural Southwestern Uganda

BACKGROUND: Over 3 million stillbirths occur annually in sub Saharan Africa; most occur intrapartum and are largely preventable. The standard of care for fetal heart rate (FHR) assessment in most sub-Saharan African settings is a Pinard Stethoscope, limiting observation to one person, at one point i...

Descripción completa

Detalles Bibliográficos
Autores principales: Mugyenyi, Godfrey R, Atukunda, Esther C, Ngonzi, Joseph, Boatin, Adeline, Wylie, Blair J., Haberer, Jessica E.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: BioMed Central 2017
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5465540/
https://www.ncbi.nlm.nih.gov/pubmed/28595604
http://dx.doi.org/10.1186/s12884-017-1361-1
_version_ 1783242962085347328
author Mugyenyi, Godfrey R
Atukunda, Esther C
Ngonzi, Joseph
Boatin, Adeline
Wylie, Blair J.
Haberer, Jessica E.
author_facet Mugyenyi, Godfrey R
Atukunda, Esther C
Ngonzi, Joseph
Boatin, Adeline
Wylie, Blair J.
Haberer, Jessica E.
author_sort Mugyenyi, Godfrey R
collection PubMed
description BACKGROUND: Over 3 million stillbirths occur annually in sub Saharan Africa; most occur intrapartum and are largely preventable. The standard of care for fetal heart rate (FHR) assessment in most sub-Saharan African settings is a Pinard Stethoscope, limiting observation to one person, at one point in time. We aimed to test the functionality and acceptability of a wireless FHR monitor that could allow for expanded monitoring capacity in rural Southwestern Uganda. METHODS: In a mixed method prospective study, we enrolled 1) non-laboring healthy term pregnant women to wear the device for 30 min and 2) non-study clinicians to observe its use. The battery-powered prototype uses Doppler technology to measure fetal cardiotocographs (CTG), which are displayed via an android device and wirelessly transmit to cloud storage where they are accessible via a password protected website. Prototype functionality was assessed by the ability to obtain and transmit a 30-min CTG. Three obstetricians independently rated CTGs for readability and agreement between raters was calculated. All participants completed interviews on acceptability. RESULTS: Fifty pregnant women and 7 clinicians were enrolled. 46 (92.0%) CTGs were successfully recorded and stored. Mean scores for readability were 4.71, 4.71 and 4.83 (out of 5) with high agreement (intra class correlation 0.84; 95% CI 0.74 to 0.91). All pregnant women reported liking or really liking the device, as well as high levels of comfort, flexibility and usefulness of the prototype; all would recommend it to others. Clinicians described the prototype as portable, flexible, easy-to-use and a time saver. Adequate education for clinicians and women also seemed to improve correct usage and minimise concerns on safety of the device. CONCLUSIONS: This prototype wireless FHR monitor functioned well in a low-resource setting and was found to be acceptable and useful to both pregnant women and clinicians. The device also seemed to have potential to improve the experience of the users compared with standard of care and expand monitoring capacity in settings where bulky, wired or traditional equipment are unreliable. Further research needs to investigate the potential impact and cost of such innovations to improve perinatal outcomes.
format Online
Article
Text
id pubmed-5465540
institution National Center for Biotechnology Information
language English
publishDate 2017
publisher BioMed Central
record_format MEDLINE/PubMed
spelling pubmed-54655402017-06-09 Functionality and acceptability of a wireless fetal heart rate monitoring device in term pregnant women in rural Southwestern Uganda Mugyenyi, Godfrey R Atukunda, Esther C Ngonzi, Joseph Boatin, Adeline Wylie, Blair J. Haberer, Jessica E. BMC Pregnancy Childbirth Research Article BACKGROUND: Over 3 million stillbirths occur annually in sub Saharan Africa; most occur intrapartum and are largely preventable. The standard of care for fetal heart rate (FHR) assessment in most sub-Saharan African settings is a Pinard Stethoscope, limiting observation to one person, at one point in time. We aimed to test the functionality and acceptability of a wireless FHR monitor that could allow for expanded monitoring capacity in rural Southwestern Uganda. METHODS: In a mixed method prospective study, we enrolled 1) non-laboring healthy term pregnant women to wear the device for 30 min and 2) non-study clinicians to observe its use. The battery-powered prototype uses Doppler technology to measure fetal cardiotocographs (CTG), which are displayed via an android device and wirelessly transmit to cloud storage where they are accessible via a password protected website. Prototype functionality was assessed by the ability to obtain and transmit a 30-min CTG. Three obstetricians independently rated CTGs for readability and agreement between raters was calculated. All participants completed interviews on acceptability. RESULTS: Fifty pregnant women and 7 clinicians were enrolled. 46 (92.0%) CTGs were successfully recorded and stored. Mean scores for readability were 4.71, 4.71 and 4.83 (out of 5) with high agreement (intra class correlation 0.84; 95% CI 0.74 to 0.91). All pregnant women reported liking or really liking the device, as well as high levels of comfort, flexibility and usefulness of the prototype; all would recommend it to others. Clinicians described the prototype as portable, flexible, easy-to-use and a time saver. Adequate education for clinicians and women also seemed to improve correct usage and minimise concerns on safety of the device. CONCLUSIONS: This prototype wireless FHR monitor functioned well in a low-resource setting and was found to be acceptable and useful to both pregnant women and clinicians. The device also seemed to have potential to improve the experience of the users compared with standard of care and expand monitoring capacity in settings where bulky, wired or traditional equipment are unreliable. Further research needs to investigate the potential impact and cost of such innovations to improve perinatal outcomes. BioMed Central 2017-06-08 /pmc/articles/PMC5465540/ /pubmed/28595604 http://dx.doi.org/10.1186/s12884-017-1361-1 Text en © The Author(s). 2017 Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.
spellingShingle Research Article
Mugyenyi, Godfrey R
Atukunda, Esther C
Ngonzi, Joseph
Boatin, Adeline
Wylie, Blair J.
Haberer, Jessica E.
Functionality and acceptability of a wireless fetal heart rate monitoring device in term pregnant women in rural Southwestern Uganda
title Functionality and acceptability of a wireless fetal heart rate monitoring device in term pregnant women in rural Southwestern Uganda
title_full Functionality and acceptability of a wireless fetal heart rate monitoring device in term pregnant women in rural Southwestern Uganda
title_fullStr Functionality and acceptability of a wireless fetal heart rate monitoring device in term pregnant women in rural Southwestern Uganda
title_full_unstemmed Functionality and acceptability of a wireless fetal heart rate monitoring device in term pregnant women in rural Southwestern Uganda
title_short Functionality and acceptability of a wireless fetal heart rate monitoring device in term pregnant women in rural Southwestern Uganda
title_sort functionality and acceptability of a wireless fetal heart rate monitoring device in term pregnant women in rural southwestern uganda
topic Research Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5465540/
https://www.ncbi.nlm.nih.gov/pubmed/28595604
http://dx.doi.org/10.1186/s12884-017-1361-1
work_keys_str_mv AT mugyenyigodfreyr functionalityandacceptabilityofawirelessfetalheartratemonitoringdeviceintermpregnantwomeninruralsouthwesternuganda
AT atukundaestherc functionalityandacceptabilityofawirelessfetalheartratemonitoringdeviceintermpregnantwomeninruralsouthwesternuganda
AT ngonzijoseph functionalityandacceptabilityofawirelessfetalheartratemonitoringdeviceintermpregnantwomeninruralsouthwesternuganda
AT boatinadeline functionalityandacceptabilityofawirelessfetalheartratemonitoringdeviceintermpregnantwomeninruralsouthwesternuganda
AT wylieblairj functionalityandacceptabilityofawirelessfetalheartratemonitoringdeviceintermpregnantwomeninruralsouthwesternuganda
AT habererjessicae functionalityandacceptabilityofawirelessfetalheartratemonitoringdeviceintermpregnantwomeninruralsouthwesternuganda